Skip to content

Comprehensive® Prolong® 110031423Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of Comprehensive Prolong (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887868231544), received between Jun 23, 2023 and Oct 23, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Naturally Worn1

Patient problems

  • Joint Dislocation2
  • Failure of Implant1
  • Insufficient Information1
  • Joint Impingement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SHOULDER ARTHROPLASTY APPROXIMATELY TWO (2) MONTHS AGO DUE TO INSTABILITY. SUBSEQUENTLY, THE PATIENT IS BEING CONSIDERED FOR A REVISION TO INCREASE CONSTRAINT ON AN UNKNOWN DATE. A CUSTOM IMPLANT REQUEST WAS SUBMITTED TO A COMPETITOR TO REPLACE ALL CURRENT IMPLANTS EXCEPT THE STEM AND GLENOID BASEPLATE. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20515474 ↗
  • InjuryApr 16, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED BY THE PMI GROUP THAT A PATIENT UNDERWENT THEIR SECOND REVISION APPROXIMATELY FIVE (5) MONTHS POST-IMPLANTATION DUE TO CHRONIC DISLOCATION CAUSED BY THE HUMERAL COMPONENT COMING IN CONTACT WITH THE VRS. DURING THIS PROCEDURE, THE POLY AND GLENOSPHERE WERE EXCHANGED AND THE VRS COMPONENT REMAINED IMPLANTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation; Joint Impingement

    MAUDE report 19116185 ↗
  • InjuryJun 23, 2023Detachment of Device or Device Component; Device Dislodged or Dislocated; Naturally Worn; Detachment of Device or Device Component; Device Dislodged or Dislocated; Naturally Worn

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 17189675 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page