Virage® 07.01764.002Recalls & reported problems

Zimmer Biomet Inc · Class 1

All models of Virage (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00887868289095), received between Aug 11, 2022 and May 28, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT TWO VIRAGE DRIVER TIPS FRACTURED WHEN ATTEMPTING TO REMOVE A SCREW AT T-2 INTRA-OPERATIVELY. THE SURGERY WAS COMPLETED WITH A THIRD DRIVER, RESULTING IN A 20 MINUTE DELAY BUT NO PATIENT HARM. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22100068 ↗
  • MalfunctionMay 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT TWO VIRAGE DRIVER TIPS FRACTURED WHEN ATTEMPTING TO REMOVE A SCREW AT T-2 INTRA-OPERATIVELY. THE SURGERY WAS COMPLETED WITH A THIRD DRIVER, RESULTING IN A 20 MINUTE DELAY BUT NO PATIENT HARM. THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22100058 ↗
  • MalfunctionOct 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY, A VIRAGE SCREWDRIVER FRACTURED AT THE TIP AND A 5.5MM DIAMETER VIRAGE SCREW BROKE APART AT THE TULIP HEAD. THE PROCEDURE WAS COMPLETED WITH ADDITIONAL SCREW DRIVERS AND SCREWS IN THEIR SET. THERE WAS NO PATIENT IMPACT; HOWEVER, THERE WAS A 10-15 MINUTE DELAY TO GRAB NEW SCREW DRIVERS AND SCREW.THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20452810 ↗
  • MalfunctionMay 22, 2023Fracture; Fracture

    IT WAS REPORTED THE TIP OF THE VIRAGE POLY-AXIAL SCREWDRIVER BROKE WHILE POSITIONING DURING A CERVICAL FIXATION AT C4. THE BROKEN PIECE WAS RETRIEVED WITH A PAIR OF SURGICAL TWEEZERS AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DRIVER. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16973857 ↗
  • MalfunctionAug 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY, THE TIP OF A SCREW DRIVER FRACTURED. ALL PARTS WERE RETRIEVED AND THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15214668 ↗