Skip to content

Oxford® Partial Knee System 166846Recalls & reported problems

Biomet Uk Ltd · Class 3 · Implant

All models of Oxford Partial Knee System (92)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887868354823), received between Oct 3, 2023 and Apr 18, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Migration” (1 report).

22 other recalls cover devices cleared under the same FDA submission (P010014); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Migration1
  • Use of Device Problem1

Patient problems

  • Bone Fracture(s)1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 18, 2024Migration; Migration

    IT WAS REPORTED THAT A PATIENT HAD INITIAL BILATERAL UNICOMPARTMENTAL MEDIAL KNEE ARTHROPLASTIES PERFORMED APPROXIMATELY FIVE MONTHS AGO. SUBSEQUENTLY, TWO MONTHS LATER, THE PATIENT¿S RIGHT KNEE WAS CONVERTED TO A TOTAL KNEE DUE TO A DEPRESSED TIBIA. THE PATIENT HAD INADEQUATE BONE STOCK TO SUPPORT THE PARTIAL KNEE IMPLANTS. NO MEDICAL RECORDS OR FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19132612 ↗
  • InjuryOct 3, 2023Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WHO UNDERWENT UNICOMPARTMENTAL KNEE ARTHROPLASTY EXPERIENCED A TIBIA BONE FRACTURE POSTOPERATIVELY. THE IMPLANTS REMAIN IN THE PATIENT'S BODY. THE SURGEON COMMENTED THAT THE TIBIAL CEMENTLESS BASE PLATE IS LARGER THAN THE ACTUAL DIMENSIONS INDICATED ON THE LABEL, AND THE CEMENTLESS BASE PLATE WITH A LARGER DIMENSION THAN INDICATED MAY HAVE BEEN USED, RESULTING IN A TIBIAL FRACTURE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Bone Fracture(s)

    MAUDE report 17859354 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010014. One submission often covers many models, so check the product and lot details in each record.

  • Z-1152-2017 · Jan 5, 2016 · Terminated
    Oxford Knee System Tibial Resector Body Tube & Guides Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1169-2017 · Jan 5, 2016 · Terminated
    Oxford Unicompartmental Knee Phase 3 Shim Size 2 Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1154-2017 · Jan 5, 2016 · Terminated
    Oxford Unicompartmental Knee Phase 3 Tibial Impactor Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1171-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size D Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1163-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Left Medial Tibial Trial Tray Size D Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1157-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Left Medial Tibial Trial Tray Size F Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1164-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1173-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1168-2017 · Jan 5, 2016 · Terminated
    Oxford Unicompartmental Knee Phase 3 Shim Size 3 Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1172-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Phase 3 Tibial Template Left Medial Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1155-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size A Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1161-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Left Medial Tibial Trial Tray Size E Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1158-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Right Medial Tibial Trial Tray Size A Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1170-2017 · Jan 5, 2016 · Terminated
    Oxford Unicompartmental Knee Phase 3 Shim Size 1 Product Usage: Instruments for use with the Oxford Uni Partial Knee System
  • Z-1153-2017 · Jan 5, 2016 · Terminated
    Oxford Partial Knee System Femoral Slap Hammer Product Usage: Instruments for use with the Oxford Uni Partial Knee System

…and 7 more.

Other Prosthesis, knee, femorotibial, unicompartmental, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, femorotibial, unicompartmental, semi-constrained, metal/polymer, mobile bearing devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page