Skip to content

NA 802802202Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K181171); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 24, 2025Open, Classified

    The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

    Z-1648-2025 · Root cause: Under Investigation by firm · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K181171. One submission often covers many models, so check the product and lot details in each record.

  • Z-1649-2025 · Mar 24, 2025 · Open, Classified
    Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
  • Z-1647-2025 · Mar 24, 2025 · Open, Classified
    Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, -3 Neck, Model Number 802802201; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
  • Z-1645-2025 · Mar 24, 2025 · Open, Classified
    Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncem
  • Z-1646-2025 · Mar 24, 2025 · Open, Classified
    Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncem

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.