Skip to content

Identity® Shoulder System 110047346Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887868580765), received between Jul 25, 2025 and Jul 25, 2025 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SHOULDER REVISION SURGERY APPROXIMATELY THREE (3) MONTHS AGO. SUBSEQUENTLY, THE PATIENT IS BEING CONSIDERED FOR A REVISION SURGERY DUE TO AN UNKNOWN REASON USING A CUSTOM GLENOSPHERE AND A CUSTOM BEARING. ATTEMPT HAS BEEN MADE TO OBTAIN ADDITIONAL INFORMATION. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 22608029 ↗

FDA clearance and classification

Cleared through 510(k) K240876 on Jun 12, 2024 (decision: Substantially Equivalent). Submitted by Zimmer, Inc. as “Identity Shoulder System”. The FDA review took 75 days. FDA record ↗

The FDA regulates product code KWT (prosthesis, shoulder, non-constrained, metal/polymer cemented) under 21 CFR 888.3650 as a Class II device. 21 CFR 888.3650 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Identity® Shoulder System 110047346

Has the Identity® Shoulder System 110047346 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887868580765), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Identity® Shoulder System 110047346?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jul 25, 2025. Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Identity® Shoulder System 110047346?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Identity® Shoulder System 110047346 FDA approved?

It was cleared through 510(k) K240876 on Jun 12, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Identity® Shoulder System 110047346 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Identity® Shoulder System 110047346, and is it still sold?

The FDA's device database lists Zimmer, Inc. as the company responsible for it (brand Identity® Shoulder System, model 110047346). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page