Zyston® Curve Interbody Fusion System 14-533041Recalls & reported problems
Biomet Spine LLC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888480470182), received between Nov 16, 2020 and Aug 26, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Material Fragmentation2
- Naturally Worn2
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50459784 ↗MalfunctionAug 12, 2025Material Fragmentation; Naturally Worn; Material Fragmentation; Naturally Worn
IT WAS REPORTED THAT EIGHT ZYSTON INNER SHAFTS HAD BECOME WORN AND ARE BEGINNING TO SEND SHARDS OF METAL UPON STRIKING WITH MALLET BY THE SURGEON. THERE WAS NO REPORTED PATIENT OR SURGICAL IMPACT. THIS IS REPORT THREE OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784681 ↗MalfunctionAug 12, 2025Material Fragmentation; Naturally Worn; Material Fragmentation; Naturally Worn
IT WAS REPORTED THAT EIGHT ZYSTON INNER SHAFTS HAD BECOME WORN AND ARE BEGINNING TO SEND SHARDS OF METAL UPON STRIKING WITH MALLET BY THE SURGEON. THERE WAS NO REPORTED PATIENT OR SURGICAL IMPACT. THIS IS REPORT ONE OF EIGHT FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784205 ↗MalfunctionMar 11, 2022Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF A THREADED VARIABLE INSERTER BROKE OFF INTO ITS MATING IMPLANT UNDER IMPACTION. THE IMPLANT WITH THE TIP WAS REMOVED AND REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13736988 ↗InjuryNov 16, 2020Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF THE ZYSTON CURVE INSERTER'S DRAW ROD BROKE OFF INSIDE A CAGE DURING SURGERY AT THE TIME OF THE FINAL IMPACTION. THE CAGE WITH THE TIP WAS LEFT IMPLANTED IN THE PATIENT AND BONE WAS PACKED BEHIND IT. THERE WERE NO IMMEDIATE PATIENT IMPACTS REPORTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10848551 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page