Skip to content

ZYSTON CURVE SPACER SYS 14-533128Recalls & reported problems

Biomet Spine LLC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888480611493), received between Feb 19, 2025 and Feb 19, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1
  • Fracture1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2025Fracture; Device-Device Incompatibility; Fracture; Device-Device Incompatibility

    IT WAS REPORTED THAT A ZYSTON CURVE SPACER FRACTURED IN THE DISC SPACE INTRA-OPERATIVELY. THE DOCTOR FELT THAT THE INSERTER CONNECTING WITH THE CAGE WAS LOOSE, AND HE DECIDED TO PULL IT OUT AND REPLACE THE CAGE. DURING THE REMOVAL PROCESS, HE DISCOVERED THAT THE CAGE WAS BROKEN. THE CRACKED PEEK WAS REMOVED AS MUCH AS POSSIBLE, HOWEVER IT IS POSSIBLE THAT A SMALL AMOUNT REMAINED IN THE PATIENT BODY. THE IMPLANT WAS REMOVED AND REPLACED WITH A NEW IMPLANT OF THE SAME SIZE USING THE SAME INSERTER, SO IT IS UNLIKELY THERE WAS AN ISSUE WITH THE INSERTER. THERE WAS ONLY A FIVE MINUTE PROCEDURAL DELAY AND OTHERWISE NO PATIENT HARM.

    Patient problems: Foreign Body In Patient

    MAUDE report 21410313 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page