HVAD® Pump Outflow Graft MCS1725OGRecalls & reported problems
Heartware Inc. · Class 3 · Implant
All models of HVAD Pump Outflow Graft (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888707007634), received between Apr 14, 2021 and Apr 14, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
59 other recalls cover devices cleared under the same FDA submission (P100047); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 3, 2021Class IOpen, Classified
Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.
Z-2196-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 14, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING THE VENTRICULAR ASSIST DEVICE (VAD) IMPLANT PROCEDURE, THE OUTFLOW GRAFT (OFG) EXHIBITED LEAKING ASSOCIATED WITH A HOLE OR TINY SLIT IN THE OFG. A SUTURE WAS PLACED, BUT THE LEAK WAS STILL PROFOUND. THE OFG WAS EXCHANGED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11668607 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100047. One submission often covers many models, so check the product and lot details in each record.
- Z-0475-2024 · Oct 31, 2023 · Open, ClassifiedHVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 1100
- Z-1993-2023 · May 16, 2023 · Open, ClassifiedHeartWare Ventricular Assist Device System. a) Monitor 1521US Explore Tech, b) Monitor 1521GB Explore Tech, c) Monitor 1521IL Explore Tech, d) Monitor 1521DE Explore Tech
- Z-1512-2022 · Jun 28, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Batteries, Product Number 1650DE
- Z-1266-2022 · May 5, 2022 · Open, ClassifiedHeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.
- Z-1207-2022 · Apr 14, 2022 · Open, ClassifiedHeartWare HVAD Pump Implant Kit, a) Model Number 1104JP, b) Model Number MCS1705PU
- Z-1103-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1407DE Kit
- Z-1107-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System, individual pamphlets: HVAD System Emergency Responder Guide (FI-Finnish) HVAD System Emergency Responder Guide (FI-Finnish) HVAD
- Z-1102-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System, Pump Implant Kit, Model Number1104
- Z-1106-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System, Model Number 1521DE, MONITOR 1521DE EXPLORE TECH
- Z-1105-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1420-Controller 2.0
- Z-1110-2022 · Mar 30, 2022 · Open, ClassifiedInstructions for Use, Patient Manuals, and Emergency Responder Guides for HeartWare Ventricular Assist Device (HVAD) System. HVAD System. IFUs are distributed as part of the Pump
- Z-1104-2022 · Mar 30, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1420
- Z-0726-2022 · Jan 27, 2022 · Open, ClassifiedHeartWare Ventricular Assist Device (HVAD) System
- Z-2478-2021 · Aug 6, 2021 · Open, ClassifiedMedtronic HVAD Pump Implant Kits
- Z-2194-2021 · Jun 3, 2021 · Open, ClassifiedHeartWare HVAD Implant Kit, REF 1104JP
…and 44 more.
Other Ventricular (Assist) Bypass devices
- Thoratec® HeartMate 3™ 106524USThoratec Corporation · Ventricular (Assisst) Bypass7 recalls5,096 deaths16,478 reports
- THORATEC® HEARTMATE®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 106015Thoratec Corporation · Ventricular (Assisst) Bypass2 recalls2,141 deaths8,000 reports
- Thoratec® HeartMate 3™ 106531USThoratec Corporation · Ventricular (Assisst) Bypass1 recall11 deaths4,646 reports
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HEARTMATE®, MOBILE POWER UNIT, NA 107754Thoratec Corporation · Ventricular (Assisst) Bypass1 death2,098 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page