Arthrex® AR-4151DSRecalls & reported problems

Arthrex, Inc. · TRIM-IT DRILL PIN, 1.5MM X 100MM · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867032781), received between Mar 31, 2017 and Jun 4, 2024 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem3
  • Patient-Device Incompatibility3
  • Break1
  • Material Fragmentation1
  • Material Protrusion/Extrusion1
  • Migration or Expulsion of Device1
  • Use of Device Problem1

Patient problems

  • Pain4
  • Swelling3
  • Discomfort2
  • Unspecified Infection2
  • Erythema1
  • Foreign Body In Patient1
  • Hypersensitivity/Allergic reaction1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2024Break; Break

    ON 12/20/2022, IT WAS REPORTED BY A DISTRIBUTOR VIA SEMS THAT AN AR-4151DS TRIM-IT DRILL PIN DISPOSABLES KIT DID NOT PROVIDE THE PROPER ROTATIONAL STABILITY AS IT SNAPPED UPON MANIPULATION. CASE WAS COMPLETED USING A 2MM PIN INSTEAD. THIS WAS DISCOVERED DURING A FUSION 3RD PHALANGE PROCEDURE ON (B)(6) 2022. ADDITIONAL INFORMATION RECEIVED ON 12/22/2022: THE DEVICE DID NOT COME IN CONTACT WITH THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19457391 ↗
  • InjuryMay 23, 2019Patient-Device Incompatibility; Patient Device Interaction Problem; Patient-Device Incompatibility; Patient Device Interaction Problem

    A PATIENT HAS REPORTED THROUGH THE ARTHREX WEBSITE THAT SHE HAD UNDERGONE SURGERY IN (B)(6) 2018 DURING WHICH A TRIM-IT PIN WAS IMPLANTED. PATIENT STATES SHE HAS AN ALLERGY TO ACRYLIC POLYMERS AND WAS INQUIRING AS TO THE MATERIAL COMPOSITION OF THE IMPLANT. SPECIFIC ARTHREX PART NUMBER WAS NOT PROVIDED. PATIENT STATES THE DEVICE MAY BE CAUSING A DELAYED FOREIGN BODY REACTION IN THE TOES THAT THE SURGERY WAS PERFORMED ON. ADDITIONAL INFORMATION PROVIDED 4/29/2019: DATE OF SURGERY WAS (B)(6) 2018. SURGERY WAS A HAMMER TOE PROCEDURE OF MIDDLE TOE/RIGHT FOOT. PATIENT ISSUES FIRST BEGAN IN (B)(6) 2018. DURING THE 9/19/18 PROCEDURE AN ARTHREX AR-4151DS TRIM IT PIN WAS IMPLANTED. SYMPTOMS INCLUDE SWELLING, REDNESS, PAIN WHICH GENERATES FROM PATIENT TOE TO HER CALF. TO DATE SHE HAS BEEN TREATED WITH CORTISONE SHOTS. PATIENT IS BEING PROVIDED WITH THE MATERIAL COMPOSITION OF THE DEVICE AS REQUESTED.

    Patient problems: Erythema; Hypersensitivity/Allergic reaction; Pain; Swelling

    MAUDE report 8636768 ↗
  • InjuryNov 1, 2018Patient-Device Incompatibility; Patient Device Interaction Problem; Patient-Device Incompatibility; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT IS EXPERIENCING PAIN, SWELLING AND A REPORTED INFECTION SINCE A TRIM IT PIN WAS IMPLANTED IN (B)(6) 2016. THE REP STATED THAT THE TRIM IT PIN WAS IMPLANTED FOR A HAMMERTOE CORRECTION. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED ON 10/10/2018: THE REP CONFIRMED THAT THE ORIGINAL PROCEDURE TOOK PLACE ON (B)(6) 2016. THE ORIGINAL PROCEDURE WAS A HAMMERTOE CORRECTION PROCEDURE ON TOES, 2, 3, AND 4. THE FOLLOWING PART AND LOT NUMBERS WERE IMPLANTED DURING THE ORIGINAL (B)(6) 2016 PROCEDURE: AR-4151DS: LOT: 10048351 / QTY. 2. AR-4151DS: LOT: 10029027 / QTY. 1. THE REP STATED THEY ARE NOT AWARE OF THE SPECIFIC DATE OF WHEN THE PATIENT STARTED EXPERIENCING PAIN, SWELLING, DISCOMFORT, AND REPORTED INFECTION. THE REP REPORTED THAT THEY WERE NOT GIVEN ANY SPECIFICS ON TYPE OF INFECTION, OR IF THERE HAS BEEN TESTING DONE. THERE ARE NOT ANY DETAILS ON CURRENT TREATMENT FOR THE PATIENT'S REPORTED INFECTION AND SYMPTOMS. THE REP STATED THAT THE SAME SURGEON WHO PERFORMED THE ORIGINAL PROCEDURE DOES PLAN TO REMOVE THE PIN AT THE SAME LOCATION OF THE ORIGINAL SURGERY. THE REP CONFIRMED THAT AS OF NOW THERE HAS NOT BEEN A REMOVAL SURGERY

    Patient problems: Discomfort; Pain; Swelling; Unspecified Infection

    MAUDE report 8030510 ↗
  • InjuryNov 1, 2018Patient-Device Incompatibility; Patient Device Interaction Problem; Patient-Device Incompatibility; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT IS EXPERIENCING PAIN, SWELLING AND A REPORTED INFECTION SINCE A TRIM IT PIN WAS IMPLANTED IN (B)(6) 2016. THE REP STATED THAT THE TRIM IT PIN WAS IMPLANTED FOR A HAMMERTOE CORRECTION. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED ON 10/10/2018: THE REP CONFIRMED THAT THE ORIGINAL PROCEDURE TOOK PLACE ON (B)(6) 2016. THE ORIGINAL PROCEDURE WAS A HAMMERTOE CORRECTION PROCEDURE ON TOES, 2, 3, AND 4. THE FOLLOWING PART AND LOT NUMBERS WERE IMPLANTED DURING THE ORIGINAL (B)(6) 2016 PROCEDURE: AR-4151DS: LOT: 10048351 / QTY. 2; AR-4151DS: LOT: 10029027 / QTY. 1. THE REP STATED THEY ARE NOT AWARE OF THE SPECIFIC DATE OF WHEN THE PATIENT STARTED EXPERIENCING PAIN, SWELLING, DISCOMFORT, AND REPORTED INFECTION. THE REP REPORTED THAT THEY WERE NOT GIVEN ANY SPECIFICS ON TYPE OF INFECTION, OR IF THERE HAS BEEN TESTING DONE. THERE ARE NOT ANY DETAILS ON CURRENT TREATMENT FOR THE PATIENT'S REPORTED INFECTION AND SYMPTOMS. THE REP STATED THAT THE SAME SURGEON WHO PERFORMED THE ORIGINAL PROCEDURE DOES PLAN TO REMOVE THE PIN AT THE SAME LOCATION OF THE ORIGINAL SURGERY. THE REP CONFIRMED THAT AS OF NOW THERE HAS NOT BEEN A REMOVAL SURGERY

    Patient problems: Discomfort; Pain; Swelling; Unspecified Infection

    MAUDE report 8030496 ↗
  • InjuryDec 11, 2017Material Fragmentation; Use of Device Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HAMMERTOE REPAIR PROCEDURE ON HER RIGHT FOURTH TOE ON (B)(6) 2017, USING TRIM IT PINS. THE PIN DID NOT PERFORM AS IT SHOULD, AND SHE HAD TO DO A SECOND SURGERY ON (B)(6) 2017 TO REMOVE THE PIN. NO FURTHER DETAILS PROVIDED AT TIME OF INITIAL REPORT. FOLLOW-UP INVESTIGATION FROM MEDICAL RECORDS: OP REPORT (B)(6) 2017 PROCEDURE: THE PATIENT IMPLANT WAS AN AR-4151DS, IMPLANT DELIVERY SYSTEM, TRIM-IT DRILL PIN 1.5X 100 MM. PATIENT DIAGNOSIS WHICH LED TO THE (B)(6) 2017 PROCEDURE WAS RIGHT FOURTH DIGIT HAMMERTOE. PATIENT (B)(6) AT TIME OF REVISION. (B)(6) LBS. HEIGHT (B)(6) INCHES, (B)(6). OP REPORT (B)(6) 2017 PROCEDURE: ON (B)(6) 2017 PATIENT UNDERWENT A REVISION SURGERY DUE TO PAINFUL RETAINED INTERNAL FIXATIVE, DEEP RIGHT 4TH DIGIT. EXTENSIVE DISSECTION WAS NEEDED TO REMOVE THE TRIM-IT PIN. IT WAS NECESSARY TO CUT THE TRIM-IT PIN AT THE LEVEL OF THE DISTAL INTERPHALANGEAL JOINT AND THEN IT WAS REMOVED PROXIMALLY AND DISTALLY. THE PROXIMAL PORTION WAS FOUND TO BE IN AN ODD SHAPE AND ONE PORTION WAS FOUND TO BE IN A U SHAPE. THE SECOND SMALL FRAGMENT WAS REMOVED DISTALLY AND THE ENTIRE PIECE MEASURED 3 CM WHICH WAS FELT TO BE THE ENTIRE LENG

    Patient problems: Foreign Body In Patient

    MAUDE report 7105364 ↗
  • InjuryMar 31, 2017Migration or Expulsion of Device; Material Protrusion/Extrusion

    IT WAS REPORTED BY PATIENT THAT SHE HAD SURGERY ON (B)(6) 2015 FOR THE FOLLOWING PROCEDURES: RIGHT AUSTIN BUNIONECTOMY, SECOND TOE ARTHRODESIS, SECOND METATARSOPHALANGEAL JOINT RELEASE, AND TAILOR'S BUNION EXOSTECTOMY. DURING THE PROCEDURE, AN ARTHREX AR-4151DS, IMPLANT DELIVERY SYSTEM, TRIM-IT DRILL PIN, 1.5MM X 100MM WAS IMPLANTED. PATIENT STATES THAT AFTER 14 MONTHS, THE PIN HAD NOT DEGRADED, IT HAD MIGRATED AND STARTED PROTRUDING AND CAUSE HER A LOT OF PAIN. ON (B)(6) 2016, PATIENT UNDERWENT A SECOND SURGERY TO REMOVE THE IMPLANT. WHEN PIN WAS REMOVED IT WAS FULLY INTACT WITH NO SIGNS OF HAVING BEGUN TO DEGRADE.

    Patient problems: Pain

    MAUDE report 6449705 ↗

Other PIN, FIXATION, SMOOTH devices

All PIN, FIXATION, SMOOTH devices →