Arthrex® AR-8981-06SRecalls & reported problems
Arthrex, Inc. · SMALL JOINT OATS SET, 6 MM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867056817), received between May 11, 2023 and May 11, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Nov 2, 2022Open, Classified
The devices may be packaged with the wrong reamer size.
Z-0465-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 11, 2023Break; Break
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT THE DONOR T-HANDLE FROM AN AR-1981-06S SINGLE USE OSTEOCHONDRAL AUTOGRAFT TRANSFER SYSTEM OATS SET AUTOGRAFT DIAMETER 6 MM METAL BEGAN TO WARP AFTER IT WAS INSERTED, AND AT THE DESIRED DEPTH. THE SURGEON WAS ABLE TO REMOVE THE DONOR T-HANDLE WITHOUT BREAKING IT. AN AR-8981-06S SMALL JOINT OATS, DONOR T-HANDLE WAS OPENED AND USED. THE SAME THING HAPPENED; IT WARPED, BUT DURING REMOVAL, IT BROKE IN THE CONDYLE. THEN, THE SURGEON ATTEMPTED TO USE AN AR-1981-08S, OATS SIZE 8MM, AND AN AR-1981-10S OATS, SIZE 10 MM, BUT IT WOULD NOT HARVEST A PLUG. AFTER AN EXTENDED PERIOD OF TIME, THE SURGEON WAS ABLE TO REMOVE THE BROKEN METAL WITH A CLAMP. THE SURGEON WAS ABLE TO USE THE PLUG OUT OF THE BROKEN PIECE AND COMPLETED THE CASE SUCCESSFULLY. THIS WAS DISCOVERED DURING AN AUTOGRAFT OATS TRANSFER PROCEDURE. ADDITIONAL INFORMATION RECEIVED ON 4/18/2023: THE CASE WAS DELAYED ABOUT FORTY-FIVE MINUTES TO AN HOUR, AND ADDITIONAL ANESTHESIA WAS ADMINISTERED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16914428 ↗
Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices
- Arthrex® AR-13995NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE430 reports
- Arthrex® AR-12990NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE197 reports
- Arthrex® AR-13991NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE128 reports
- Arthrex® AR-8981-08SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall3 reports
- Centurion EB16Centurion Medical Products Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- Medline Industries, Inc. DYNJ901391DMedline Industries, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- HarmonyAIR LAS0042Steris Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- NA Lighthandle CoverSteris Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page