Arthrex® AR-9236-02Recalls & reported problems
Arthrex, Inc. · TRIAL, MEDIUM, PEGGED GLENOID · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867059481), received between Aug 14, 2019 and Feb 4, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Appropriate Device Problem Term/Code Not Available1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Discomfort1
- Fall1
- Injury1
- Insufficient Information1
- Joint Disorder1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2025Break; Break
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9236-02 PEGGED GLENOID BROKE. THIS OCCURRED DURING USE UPON IMPACTION. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21303619 ↗MalfunctionMay 3, 2024Break; Break
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-9236-02 GLENOID TRIAL CENTRAL PEG BROKE OFF IN THE PATIENT. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. ALL FRAGMENTS WERE REMOVED. DELAY IN CASE 20 SECONDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19240198 ↗MalfunctionJul 3, 2023Break; Break
ON 06/07/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-9236-02 TRIAL BROKE. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT EFFECT. ADDITIONAL INFORMATION REQUESTED ON 6/14/2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17248124 ↗InjuryAug 14, 2019Positioning Failure; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE TRUNNION OF THE AR-9100-06P UNIVERS II STEM (SIZE 6) WOULD NOT REMAIN ENGAGED WITH THE PROXIMAL STEM BODY TO FACILITATE STEM EXTRACTION DURING A TSA REVISION TO RSA PROCEDURE. AFTER SEVERAL ATTEMPTS TO RE-ENGAGE, THE TRUNNION WOULD STILL NOT ENGAGE. A SEPARATE STEM AR-9100-13P (LOT:16.00283) WAS OPENED AND THAT TRUNNION WAS USED TO SUCCESSFULLY EXTRACT THE STEM FROM THE PATIENT WITHOUT FURTHER PROBLEMS. THE COMPLAINT DEVICE WILL BE RETURNING FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED ON 07/29/2019: THE REP STATED THE REVISION PROCEDURE WAS DUE TO THE PATIENT FALLING AND TEARING THEIR SUBSCAPULARIS. THE SURGEON DETERMINED A REVISION WAS THE BEST PLAN OF ACTION. PRIOR TO THE REVISION PROCEDURE TAKING PLACE THE PATIENT WAS EXPERIENCING DISCOMFORT DUE TO CHRONICALLY DISLOCATING THEIR SHOULDER AS A RESULT OF THE TORN ROTATOR CUFF. THE REP STATED THE DATE OF THE FALL IS UNKNOWN. THE PRIMARY ANATOMIC TSA PROCEDURE TOOK PLACE IN (B)(6) 2018, AND WAS PERFORMED BY THE SAME SURGEON WHO PERFORMED THE REVISION AND TOOK PLACE AT THE SAME FACILITY. THE BONE QUALITY WAS MEDIUM. THE FOLLOWING ARTHREX PARTS WERE IMPLANTED DURING THE PRIMARY PROCEDURE, AND EXPLANTED DURI
Patient problems: Discomfort; Fall; Injury; Joint Disorder
MAUDE report 8893676 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page