Skip to content

Arthrex® AR-9502-42LCPCRecalls & reported problems

Arthrex, Inc. · UNIVERS REVS CUP,42/+2MM LFT OFFSET CTD · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867061194), received between Nov 23, 2020 and Nov 23, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 23, 2020Break

    IT WAS REPORTED THAT DURING A (B)(6) 2020 REVISION REVERSE TOTAL SHOULDER ARTHROPLASTY PROCEDURE THE FOLLOWING ISSUES OCCURRED WITH TWO ARTHREX INSTRUMENTS: INSTRUMENT 1: THE MEDIUM T-15 DRIVER (AR-9545-T15-02 #8001928) HAS BEEN COMPROMISED. WHEN TRYING TO EXTRACT THE CENTRAL SCREW FROM A UNIVERSAL GLENOID, THE TIP OF THE DRIVER BECAME TWISTED AND A SMALL PIECE BROKE OFF INTO THE HEAD OF THE CENTRAL SCREW. THERE WAS A 5 MINUTE DELAY IN SURGERY AS THE SCREW HEAD COULD NO LONGER BE ACCESSED. THE SURGEON USED OSTEOTOMES AROUND THE PERIPHERY OF THE BASEPLATE TO EVENTUALLY EXTRACT THE BASEPLATE AND CENTRAL SCREW AS ONE PIECE. THE INSTRUMENT BROKE INTO TWO PIECES. BOTH PIECES WERE RECOVERED, BUT THE SMALLER PIECE WAS LODGED INTO THE CENTRAL SCREW IMPLANT WHICH WAS NOT AVAILABLE FOR RETURN. THERE WAS NO UNPLANNED INCISION. A SECOND SURGERY IS NOT NECESSARY. DEVICE WILL BE RETURNING FOR EVALUATION. THE EVENT DID NOT OCCUR DURING THE INITIAL USE OF THE DEVICE. INSTRUMENT 2: THE TRINITY BASEPLATE IMPACTOR (AR-9165CDG #(10)051917) HAS BEEN COMPROMISED. THE SURGEON ENGAGED THE ALREADY IMPLANTED BASEPLATE WITH THE IMPACTOR SO THAT IMPACTION MIGHT DISRUPT THE BONY BOND WITH THE BASEPLATE. AFTER

    Patient problems: Foreign Body In Patient

    MAUDE report 10884330 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page