Arthrex® AR-7237-17LNRecalls & reported problems
Arthrex, Inc. · FIBERTAPE, 17" W/ STR NDL · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867190917), received between Sep 14, 2021 and Sep 14, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Positioning Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Positioning Problem1
- Unstable1
Patient problems
- Bone Fracture(s)1
- Failure of Implant1
- Fall1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 14, 2021Unstable; Positioning Problem
IT WAS REPORTED ON (B)(6) 2021 BY A SALES REP VIA EMAIL THAT AN AR-1370A-S PATELLA SUTUREPLATE¿ II WAS BEING USED IN A CASE. AFTER PLACING TWO 4.0 CANNULATED SCREWS AND FIBERTAPE TO REVISE A TRANSVERSE PATELLA FRACTURE, THE SURGEON WANTED TO ADD A PLATE TO INCREASE STABILITY DUE TO THE PATIENT'S PREVIOUS LACK OF COMPLIANCE. AFTER CHOOSING THE ARROW SUTURE PLATE, THE SURGEON WAS UNABLE TO LOCK THE SCREWS SUCCESSFULLY IN THE PLATE. THE SCREWS WERE NOT SITTING FLUSH, WHICH CAUSED THE SURGEON TO ABANDON USE OF THE PLATE AND RELY ON THE CANNULATED SCREWS AND FIBERTAPE HE UTILIZED AT THE BEGINNING OF THE CASE. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 8/30/2021: DOS IS (B)(6) 2021, DATE OF REVISION IS (B)(6) 2021. THE PATIENT FELL ON THE OPERATIVE LEG SHORTLY AFTER SURGERY, THUS REFRACTURING HER PATELLA WHICH IS THE REASON FOR THE REVISION. DURING THE REVISION, THE SURGERY EXPLANTED THE PREVIOUSLY PLACED 4.0 CANNULATED SCREWS AND FIBERTAPE. (AR-5051-30, AR-5051-34, AR-7237-17LN).
Patient problems: Bone Fracture(s); Failure of Implant; Fall
MAUDE report 12467770 ↗
Other PIN, FIXATION, SMOOTH devices
- RIGIDFIX 210133Medos International Sàrl · PIN, FIXATION, SMOOTH91 reports
- SMART TOE ST0A-16PStryker GmbH · PIN, FIXATION, SMOOTH1 recall5 reports
- SMART TOE ST0-19PStryker GmbH · PIN, FIXATION, SMOOTH1 recall2 reports
- Flower E-Kit, Advanced EWK 201Flower Orthopedics Corporation · Pin, Fixation, Smooth1 recall0 reports
- Brasseler USA 5007972M0Peter Brasseler Holdings, LLC · PIN, FIXATION, SMOOTH1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page