Skip to content

Arthrex® AR-523-A714Recalls & reported problems

Arthrex, Inc. · CR Plus Tib Bearing, Vit E, Sz 7, 14mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867231849), received between Jun 19, 2024 and Jun 19, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2024Break; Break

    ON 10/11/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-523-A714 CR PLUS TIB BEARING IMPLANT HAD AN ISSUE. DURING A TOTAL KNEE ARTHROPLASTY PROCEDURE ON 10/6/2022, WHEN THE SURGEON WAS TRYING TO USE THE DEVICE, THE DEVICE FAILED TO ENGAGE IN THE TIBIAL TRAY, IT WOULD NOT LOCK, AND THE PERIPHERAL LOCKING MECHANISM WOULD NOT ENGAGE. THE SURGEON REMOVED THE COMPLAINT DEVICE FROM THE PATIENT, IT WAS REMOVED IN ONE PIECE AND NO PIECE OF THE INSTRUMENT BROKE INSIDE THE PATIENT. UPON INVESTIGATION, IT WAS STATED THAT THE PERIPHERAL POLY WAS BENT/DEFORMED. THE SURGEON USED ANOTHER AR-523-A714, LOT NUMBER UNKNOWN TO COMPLETE THE CASE SUCCESSFULLY WITH NO IMPACT TO THE PATIENT. THE COMPLAINT DEVICE WAS DISCARDED BY THE FACILITY. IT WAS STATED THAT THIS MIGHT HAVE HAPPENED AS AN OUT-OF-BOX EVENT, BUT IT COULD NOT BE VERIFIED AS NO PICTURES WERE TAKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19567431 ↗

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page