Arthrex, Inc. · ANGEL CENTRIFUGE US (REFURBISHED) · Class 2

Quick summary

This Arthrex Angel centrifuge is a refurbished tabletop unit used to separate platelets and plasma from blood during bone graft and PRP procedures. The 70 reports mostly describe the machine not working as expected: failing to separate blood at the end of a spin, error messages about the lid not being closed, a knob that would not drain, odd smells, and one unit that started smoking mid-procedure. Five reports involve injuries and none involve deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
66
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 66 adverse event reports that list this exact device (DI 00888867249585), received between Nov 1, 2016 and Dec 29, 2025 — classified as 5 injuries, 61 malfunctions.

The most frequently reported specific device problem is “Defective Device” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 6 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 4 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device9
  • Use of Device Problem8
  • Output Problem7
  • Display or Visual Feedback Problem4
  • Separation Failure4
  • Device Emits Odor3
  • Fluid/Blood Leak3
  • Insufficient Flow or Under Infusion3

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • Insufficient Information27
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 66 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025Use of Device Problem; Use of Device Problem

    ON 05-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE DID NOT WORK PROPERLY. THE KNOB THAT MAKES THE BLOOD FLOW, WOULD NOT DRAIN. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905998 ↗
  • MalfunctionDec 8, 2025Separation Failure; Separation Failure

    ON 11/17/2025, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE MALFUNCTIONED AND WOULD NOT ADEQUATELY SEPARATE THE BLOOD AT THE END OF THE CYCLE DURING A PROCEDURE. THE ABS-10061T PRP KIT WAS UNABLE TO BE USED DUE TO THE IMPROPER FUNCTION OF THE CENTRIFUGE, BUT IT DID CONTAIN BLOOD FROM THE PATIENT WHO WAS NOT AFFECTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23742758 ↗
  • MalfunctionDec 3, 2025Use of Device Problem; Use of Device Problem

    ON 11/11/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE LID LOCK SENSOR WAS FLASHING WHEN PAUSING AND EVENTUALLY CANCELED THE SPIN. THIS WAS DISCOVERED DURING THE CASE, BUT THEY WERE ABLE TO EMPTY THE CASSETTE AND RE-SPIN IT. THE LID HAD TO BE MANUALLY HELD DOWN WITH FORCE TO COMPLETE THE CASE. THE CASE WAS COMPLETED WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707323 ↗
  • MalfunctionDec 1, 2025Use of Device Problem; Use of Device Problem

    ON 11/07/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE WAS GIVING AN ERROR MESSAGE STATING THAT THE LID IS NOT CLOSED. THE MAGNET RESPONSIBLE FOR HOLDING THE LID CLOSED WAS DETACHED AND LATER REATTACHED, BUT THE ERROR CONTINUED TO BE GIVEN AND THE CENTRIFUGE COULD NOT BE USED. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM. ADDITIONAL INFORMATION PROVIDED ON 11/13/2025: IT WAS FURTHER REPORTED THAT THIS OCCURRED DURING A LATERAL COLUMN LENGTHENING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23680472 ↗
  • MalfunctionNov 25, 2025Output Problem; Output Problem

    ON 11/05/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE EXPERIENCED INCONSISTENT PRP OUTPUT. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645476 ↗
  • MalfunctionNov 18, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    ON 11/03/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE SCREEN WAS NO LONGER WORKING. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025. THE BLOOD WAS PROCESSED SUCCESSFULLY WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583822 ↗
  • MalfunctionNov 4, 2025Use of Device Problem; Use of Device Problem

    ON 10/23/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE DISPLAYED AN ERROR MESSAGE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23463725 ↗
  • MalfunctionAug 12, 2025False Alarm; False Alarm

    ON 07/25/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE GIVES AN ERROR CODE STATING THAT THE LID IS NOT CLOSED WHEN IT IS CLOSED ADEQUATELY. UNFORTUNATELY, IT IS UNABLE TO SPIN AS A RESULT. THIS WAS DISCOVERED DURING A CASE. AFTER 15 MINUTES OF TROUBLE SHOOTING WITHOUT SUCCESS, IT WAS DECIDED THAT BMA WAS UTILIZED INSTEAD OF BMAC.

    Patient problems: Insufficient Information

    MAUDE report 22783038 ↗
  • MalfunctionJul 17, 2025Device Emits Odor; Device Emits Odor

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE SCREEN WAS COMPLETELY BLACK AND, WHEN TURNED ON, SMELLED OF SMOKE AND BURNT RUBBER. THE ISSUE WAS DISCOVERED AFTER THE SURGERY. NO ADVERSE EFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22530171 ↗
  • MalfunctionJul 9, 2025Device Alarm System; Device Alarm System

    ON 06/13/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT WHEN AN ABS-10060R ANGEL CENTRIFUGE WAS TURNED ON IT GAVE AN UNDER-SPEED ALARM SOUND. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22460970 ↗
  • MalfunctionJun 23, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    ON 06/03/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE SCREEN STAYED WHITE WHEN POWERED ON. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22309996 ↗
  • MalfunctionJun 20, 2025No Display/Image; No Display/Image

    ON 06/02/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE DID NOT ALLOW THE MAIN SCREEN TO BE ACCESSED AND ONLY ONE OF THE BLUE LIGHTS WAS WORKING OUT OF THE THREE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22298822 ↗
  • MalfunctionMay 28, 2025No Flow; No Flow

    ON 05/06/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE PART OF ABS-10061T ANGEL PRP KIT MALFUNCTIONED. INTERMITTENTLY IT WOULD NOT PUT OUT ANY PRP WHICH LED TO THE DISC SET JAMMING IN THE SEPARATION CHAMBER. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22096419 ↗
  • MalfunctionMay 1, 2025Separation Failure; No Flow; Separation Failure; No Flow

    ON 04/14/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE NEVER SEPARATED. THIS OCCURRED DURING USE, THE PRP NEVER SEPARATED THE BLOOD AND PLASMA, IT ALL WENT INTO THE BAG. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21945012 ↗
  • MalfunctionApr 22, 2025Separation Failure; Separation Failure

    ON 04/01/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE HAD BEEN EXPERIENCING A FEW ISSUES DURING PROCEDURES. ALL OF THE SPIN CYCLES WOULD FINISH FINE BUT THE SEPARATION OF THE BLOOD WASN'T WORKING. A FEW TIMES NOTHING CAME OUT OF THE DISC, DESPITE THERE BEING NO ERROR CODE AND NORMAL READINGS. AND ANOTHER TIME ALL OF THE BLOOD DISCHARGED INTO THE RBC BAG WITH NO SEPARATION.

    Patient problems: Insufficient Information

    MAUDE report 21883006 ↗
  • MalfunctionApr 22, 2025Separation Failure; Separation Failure

    ON 04/01/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(6) THAT AN ABS-10060R ANGEL CENTRIFUGE HAD BEEN EXPERIENCING A FEW ISSUES DURING PROCEDURES. ALL OF THE SPIN CYCLES WOULD FINISH FINE, BUT THE SEPARATION OF THE BLOOD WASN'T WORKING. A FEW TIMES NOTHING CAME OUT OF THE DISC, DESPITE THERE BEING NO ERROR CODE AND NORMAL READINGS. AND ANOTHER TIME ALL OF THE BLOOD DISCHARGED INTO THE RBC BAG WITH NO SEPARATION.

    Patient problems: Insufficient Information

    MAUDE report 21882582 ↗
  • MalfunctionMar 20, 2025Use of Device Problem; Use of Device Problem

    ON 03/05/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(6) THAT AN ABS-10060R ANGEL CENTRIFUGE DIDN'T SPIN DURING A CASE. THIS OCCURRED DURING USE IN A MIS BUNION CASE WITH NO PATIENT EFFECT. CASE WAS COMPLETED USING THE BMAC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652425 ↗
  • MalfunctionMar 17, 2025Infusion or Flow Problem; Infusion or Flow Problem

    ADDITIONAL INFORMATION WAS PROVIDED ON 03/06/2025 WHERE THE REPRESENTATIVE STATED THEY WERE SCHEDULED TO USE THE NEW SYSTEM FOR THE INTRAPERITONEAL CATHETERS WHICH IS CALLED THE ANGEL SYSTEM. IT IS WHAT PREPARES THE PRP. THE ANESTHESIOLOGIST ATTEMPTED TO WITHDRAW OVER 50CC OF THE PATIENT¿S BLOOD AND WAS NOT SUCCESSFUL. ONLY 50 WAS COLLECTED, AFTER MULTIPLE ATTEMPTS BY THE ANESTHESIOLOGIST. THE BLOOD COLLECTION WAS CONNECTED TO THE DEVICE AND NO PRP WAS PRODUCED AFTER THE CYCLE. DR. MARSH WAS MADE AWARE AND WAS PRESENT DURING THE COLLECTION AND PROCESSING OF THE BLOOD. THEY CONCLUDED THE PROCEDURE WITHOUT THE USE OF THE PRP.

    Patient problems: Insufficient Information

    MAUDE report 21619325 ↗
  • MalfunctionFeb 13, 2025Calibration Problem; Calibration Problem

    ON 01/23/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(6) THAT AN ABS-10060R ANGEL CENTRIFUGE WAS PRODUCING A "LIGHT SENSOR CALIBRATION ERROR." THIS OCCURRED UPON ITS FIRST USE ON (B)(6) 2025. THIS SPIN PROTOCOL WAS CORRECTLY PERFORMED. THE PHYSICIAN WAS FORCED TO ALTER THE PROCEDURE IN QUESTION, AND UNCENTRIFUGED BONE MARROW ASPIRATE WAS HARVESTED FROM THE PATIENT AND THEN ADDED TO BIOCARTILAGE TO COMPLETE CARTILAGE RESTORATION OF A TALUS DEFECT. THE PROCEDURE WAS COMPLETED SAFELY BUT NOT WITHOUT DIFFICULTY.

    Patient problems: Insufficient Information

    MAUDE report 21373455 ↗
  • MalfunctionFeb 11, 2025Use of Device Problem; Use of Device Problem

    ON 01/17/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE EYE/LIGHT SENSOR ERROR WON¿T GO OFF. THIS OCCURRED DURING A BMAC PROCEDURE WHERE THE TUBING WAS NOT LOADED. THEY CLEANED THE SENSOR.

    Patient problems: Insufficient Information

    MAUDE report 21353969 ↗
  • MalfunctionFeb 10, 2025Output Problem; Output Problem

    ON 01/16/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE HAS CONTINUED TO GIVE TROUBLE AND MALFUNCTION WITH MULTIPLE ERRORS. THE MOST RECENT ERROR IS AIR BEING DETECTED IN THE TUBING, AS WELL AS THE LIGHT SENSOR COMING ON. THE ISSUE WAS DISCOVERED DURING THE CASE, WHICH ASKED THEM TO TRY REFILLING THE MACHINE MULTIPLE TIMES AND THEN EMPTYING THE BLOOD. WE TRIED TWO DIFFERENT ANGEL KITS, AND BOTH HAD THE SAME PROBLEM OF FILLING WITH AIR. THERE WERE NO ADVERSE EFFECTS FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21348424 ↗
  • MalfunctionFeb 5, 2025Failure to Power Up; Failure to Power Up

    ON 01/13/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGES SCREEN WOULD NOT POWER ON. THIS OCCURRED DURING A CASE, AND THE CASE WAS NOT COMPLETED AS THE CENTRIFUGE WOULDN'T WORK.

    Patient problems: Insufficient Information

    MAUDE report 21314017 ↗
  • MalfunctionJan 6, 2025Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    ON 12/09/2024 IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA SEMS-06730461 THAT AN ABS-10060R ANGEL CENTRIFUGE DISPLAYED AN INSUFFICIENT FILL ALARM. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21079464 ↗
  • MalfunctionNov 11, 2024Defective Device; Defective Device

    ON 10/29/2024 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE GAVE A CONSTANT ERROR MALFUNCTION. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20653352 ↗
  • MalfunctionNov 7, 2024Failure to Power Up; Failure to Power Up

    ON (B)(6) 2024, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN ABS-10060R ANGEL CENTRIFUGE US (REFURBISHED) HAD A "DISK BOOT FAILURE." NO PATIENT WAS HARMED, AND THEY WERE ABLE TO RUN ANOTHER ANGEL MACHINE QUICKLY TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20633185 ↗

Other Platelet and plasma separator for bone graft handling devices