Arthrex® AR-8719DS-1515Recalls & reported problems

Arthrex, Inc. · Dynanite NiTi Staple w/Instrs 15Wx15L · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867288126), received between May 24, 2024 and Feb 3, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fitting Problem1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Break

    ON 13-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-8719DS-1515 DYNANITE NITINOL STAPLE IMPLANT SYSTEM EXPERIENCED AN ISSUE DURING IMPLANTATION. THE STAPLE WAS PARTIALLY INSERTED A FEW MILLIMETERS INTO THE BONE BEFORE IT FELL OFF THE INSERTER. WHEN ATTEMPTING TO RELOAD THE STAPLE, THE TINES OF THE INSERTER REPEATEDLY POPPED OFF THE STAPLE AS THEY ATTEMPTED TO SPREAD THE LEGS. THE STAPLE COULD NOT BE RELOADED AND WAS UNABLE TO BE IMPLANTED. THIS ISSUE WAS DISCOVERED DURING A COTTON OSTEOTOMY PROCEDURE PERFORMED ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 16-JAN-2026: ALL FRAGMENTS WERE RETRIEVED. AN ADDITIONAL AR-8719DS-1515 DYNANITE NITINOL STAPLE IMPLANT SYSTEM OF THE SAME SIZE WAS OPENED AND IMPLANTED WITHOUT ISSUE. THE CASE WAS DELAYED APPROXIMATELY 4 MINUTES, WITH NO ADDITIONAL ANESTHESIA REPORTED FOR THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24248684 ↗
  • MalfunctionMay 24, 2024Fitting Problem; Fitting Problem

    ON 10/27/2021, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-8719DS-1515 STAPLE IMPLANT SYSTEM CANNOT BE FITTED. THE STAPLE AND THE DEVICE CAN NOT BE FITTED. A NEW DEVICE WAS USED AND THE CASE WAS COMPLETED. THERE WAS NO PATIENT EFFECT REPORTED. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED ON 10/28/2021: THE HOOK PART OF THE DEVICE DID NOT FIT THE STAPLE AND COULD NOT BE ATTACHED. PROCEDURE BEING PERFORMED WAS HALLUX VALGUS DISTAL OSTEOTOMY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19392458 ↗

Other Staple, fixation, bone devices

All Staple, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page