Arthrex® AR-9563-32Recalls & reported problems
Arthrex, Inc. · 5.5x32mm Peripheral Screw, LOCKING · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00888867296886), received between May 6, 2020 and Mar 4, 2026 — classified as 9 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem4
- Break3
- Defective Device2
- Positioning Failure2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Fracture1
- Loose or Intermittent Connection1
Patient problems
- Pain3
- Unspecified Infection3
- Failure of Implant2
- No Clinical Signs, Symptoms or Conditions2
- Discomfort1
- Fall1
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMar 4, 2026Break
ON 28-JAN-2026, THE PATIENT REPORTED VIA EMAIL THAT, FOLLOWING A REVERSE TOTAL SHOULDER SURGERY, THEY EXPERIENCED INCREASING POSTOPERATIVE PAIN. X-RAYS REPORTEDLY SHOWED A BROKEN SCREW, AND A CT SCAN DEMONSTRATED A LOOSE IMPLANT. THE PATIENT IS NOW SCHEDULED FOR A CORRECTIVE SURGICAL PROCEDURE. NO ADDITIONAL DETAILS WERE PROVIDED. ADDITIONAL INFORMATION WAS RECEIVED VIA PHONE FROM THE PATIENT'S RELATIVE ON 09-FEB-2026: THE PATIENT IS SCHEDULED TO UNDERGO A REVISION LEFT REVERSE TOTAL SHOULDER ARTHROPLASTY ON (B)(6) 2026. THE ORIGINAL PROCEDURE WAS PERFORMED ON (B)(6) 2024 BY A DIFFERENT SURGEON AT A DIFFERENT FACILITY. OVER TIME, THE PATIENT DEVELOPED SIGNIFICANT LEFT SHOULDER PAIN. THE PATIENT¿S SPOUSE REPORTED PROGRESSIVE DISCOMFORT IN THE PATIENT¿S LEFT ARM. DIAGNOSTIC IMAGING WAS PERFORMED, INCLUDING AN X-RAY ON (B)(6) 2025, WHICH IDENTIFIED A FRACTURED SCREW, FOLLOWED BY A CT SCAN ON (B)(6) 2025 THAT DEMONSTRATED LOOSENING OF THE IMPLANT. ADDITIONAL INFORMATION WAS RECEIVED VIA PHONE FROM THE PATIENT'S RELATIVE ON 10-FEB-2026: THE PATIENT'S SPOUSE REPORTED THAT THE REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, ALL COMPONENTS OF THE LEFT SHOULDER WERE R
Patient problems: Discomfort; Pain
MAUDE report 24506868 ↗InjuryJul 16, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2025, A CLINDEX NOTIFICATION WAS RECEIVED INDICATING THAT A PATIENT LISTED ON THE SHOULDER ARTHROPLASTY REGISTRY HAD EXPERIENCED A DEEP INFECTION. ON (B)(6) 2025, THE PATIENT PRESENTED WITH INCREASED SWELLING AND ACHING IN THE OPERATIVE SHOULDER, WITH SYMPTOMS REPORTED TO HAVE BEGUN ABOUT THREE TO FOUR WEEKS PRIOR TO THE CLINIC VISIT ON (B)(6) 2025. THE PATIENT WAS REFERRED FOR SYNOVASURE TESTING ON /B)(6) 2025, AND A VISIT ON (B)(6) 2025 CONFIRMED THE INFECTION. A REVISION WAS SUBSEQUENTLY PERFORMED ON (B)(6) 2025. THE EVENT WAS INDICATED AS UNRELATED TO THE UNIVERS REVERS APEX IMPLANTED ON (B)(6) 2024. ADDITIONAL INFORMATION RECEIVED ON 6/20/2025: DURING THE REVISION SURGERY ON (B)(6) 2025, AN AR-9501-11S UNIVERS REVERS APEX HUMERAL STEM, AN AR-9502F-36CPC ARTHREX UNIVERS REVERS SUTURE CUP, AN AR-9503-3639-3 UNIVERS REVERS MODULAR GLENOID SYSTEM COMBO HUMERAL INSERT, TWO AR-9563-16 UNIVERS REVERS MODULAR GLENOID SYSTEM PERIPHERAL LOCKING SCREW, AN AR-9563-24 UNIVERS REVERS MODULAR GLENOID SYSTEM PERIPHERAL LOCKING SCREW, AN AR-9563-32 UNIVERS REVERS MODULAR GLENOID SYSTEM PERIPHERAL LOCKING SCREW, AN AR-9564-2439-LAT ARTHREX UNIVERS REVERS MODULAR GLENOID SYSTEM GLENOSP
Patient problems: Unspecified Infection
MAUDE report 22521370 ↗MalfunctionMay 13, 2025Positioning Failure; Defective Device; Positioning Failure; Defective Device
ON 04/16/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (QTY. 2) OF AN AR-9563-32 PERIPHERAL SCREW FAILED TO THREAD INTO THE BASEPLATE. TWO DIFFERENT LOCKING SCREWS FROM SEPARATE LOTS WERE ATTEMPTED, BUT NEITHER FIT. AS A RESULT, THE SURGEON OPTED FOR A NON-LOCKING SCREW TO COMPLETE THE PROCEDURE. NO COMPONENTS BROKE IN THE PATIENT. THIS WAS DISCOVERED DURING A REVERSE SHOULDER ARTHROPLASTY PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22009328 ↗MalfunctionMay 13, 2025Positioning Failure; Defective Device; Positioning Failure; Defective Device
ON 04/16/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (QTY. 2) OF AN AR-9563-32 PERIPHERAL SCREW FAILED TO THREAD INTO THE BASEPLATE. TWO DIFFERENT LOCKING SCREWS FROM SEPARATE LOTS WERE ATTEMPTED, BUT NEITHER FIT. AS A RESULT, THE SURGEON OPTED FOR A NON-LOCKING SCREW TO COMPLETE THE PROCEDURE. NO COMPONENTS BROKE IN THE PATIENT. THIS WAS DISCOVERED DURING A REVERSE SHOULDER ARTHROPLASTY PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22009322 ↗InjuryFeb 18, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON 11/13/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA EMAIL THAT ARTHREX DEVICES WERE EXPLANTED FROM A PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED ON 11/25/2024: THE PATIENT UNDERWENT A TOTAL SHOULDER ARTHROPLASTY, ROTATOR CUFF REPAIR, CAPSULAR RELEASE PROCEDURE ON (B)(6) 2024. A 24MM GLENOID UNIVERS REVERS BASEPLATE, A 35MM CENTRAL UNIVERS REVERS SCREW, A 5.5 X 28MM PERIPHERAL LOCK UNIVERS REVERS SCREW, A 5.5 X 32MM PERIPHERAL LOCK UNIVERS REVERS SCREW, A 5.5 X 16MM PERIPHERAL LOCK UNIVERS REVERS SCREW, A 5.5 X 16MM PERIPHERAL LOCK UNIVERS REVERS SCREW, A 24 X 39MM 4 LAT TAPER UNIVERS REVERS GLENOSPHERE, A 39 2MM REVERS SHOULDER CUP, A 14 APEX UNIVERS REVERS STEM, A 39 3 REVERS SHOULDER LINER, AND A 39 6MM REVERS SHOULDER HUMERAL SPACER WERE IMPLANT. DUE TO AN INJURY THE PATIENT UNDERWENT REVISION SURGERY AND HAD THE 35MM CENTRAL UNIVERS REVERS SCREW, THE 5.5 X 28MM PERIPHERAL LOCK UNIVERS REVERS SCREW, AND THE 5.5 X 32MM PERIPHERAL LOCK UNIVERS REVERS SCREW REMOVE.
Patient problems: Failure of Implant
MAUDE report 21395238 ↗InjuryJan 15, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON 12/18/2024, A CLINDEX NOTIFICATION WAS RECEIVED INDICATING THAT A PATIENT LISTED ON THE SHOULDER ARTHROPLASTY REGISTRY, FIVE MONTHS POSTOPERATIVE, PRESENTED WITH FULLNESS IN THE ANTERIOR OF THE DELTOPECTORAL INCISION, WHICH HAS TO BE FLUCTUANCE. THE INCISION WAS DRAINED, AND A CULTURE WAS SENT TO THE LAB, WHICH INDICATED AN INFECTION. THE PHYSICIAN DECIDED TO PLACE A PICC LINE FOR IV ANTIBIOTICS. THE PATIENT ALSO UNDERWENT AN EXPLANT PROCEDURE AND HAD AN ANTIBIOTIC CEMENT SPACER IMPLANTED ON (B)(6) 2024. THE PATIENT WAS ORIGINALLY IMPLANTED ON (B)(6) 2024. THIS EVENT WAS INDICATED AS UNRELATED TO THE UNIVERS REVERS APEX IMPLANT SYSTEM. ADDITIONAL INFORMATION RECEIVED ON 1/3/2024: DURING THE REVISION SURGERY, AN AR-9503-3639-3 HUMERAL INSERT, AN AR-9563-16 PERIPHERAL SCREW, AN AR-9563-24 PERIPHERAL SCREW, AN AR-9563-32 PERIPHERAL SCREW, AN AR-9563-44 PERIPHERAL SCREW, AN AR-9564-2439-LAT GLENOSPHERE, AN AR-9580-2420-2S BASEPLATE, AND AR-9582-25 MODULAR POST WERE EXPLANTED. THE REVISION SURGERY WAS COMPLETED USING PRODUCTS FROM ANOTHER MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 21160469 ↗InjuryApr 8, 2024Break; Break
ON 3/19/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-9502-36ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 36 CAME LOOSE FROM THE SUTURECUP AND COULD NOT BE SEPARATED. THE TRIAL CA ADAPTER WAS PLACED IN THE SUTURECUP AND THE TRIAL CA HEAD WAS PLACED. AN X-RAY WAS TAKEN, AND IT WAS DETERMINED THAT THE CA HEAD WAS THE APPROPRIATE SIZE. THE TRIALS WERE REMOVED. THE STEM/CUP IMPACTOR HANDLE WAS THREADED INTO THE CA ADAPTER. THE TAB OF THE CA ADAPTER WAS ALIGNED WITH THE LATERAL SLOT OF THE SUTURECUP. THE IMPLANT WAS IMPACTED INTO THE SUTURECUP. THE CA HEAD WAS IMPACTED WITH THE GLENOSPHERE IMPACTOR MAKING SURE THE HOOD OF THE HEAD WAS LATERAL. THE CA ADAPTER CAME LOOSE FROM SUTURECUP. THE SURGEON ASKED FOR A NEW CA ADAPTER AND A CA HEAD AS THEY COULD NOT BE SEPARATED. A SECOND AR-9502-36ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER WAS PLACED ON THE IMPACTOR HANDLE AND WAS IMPACTED INTO THE SUTURECUP. THE GLENOSPHERE IMPACTOR WAS USED AS WELL TO SEAT THE ADAPTER INTO THE SUTURECUP BUT WITH LITTLE EFFORT, THE CA ADAPTER COULD BE PULLED FROM THE SUTURECUP. THE SURGEON WAS ABLE TO SEPARATE THE FIRST SET OF IMPLANTS AND TRIED AGAIN TO IMPACT THE CA ADAPTER INTO THE SUTUR
Patient problems: Failure of Implant
MAUDE report 19063505 ↗InjuryFeb 1, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
ON 1/9/2024, IT WAS REPORTED ON CASE NUMBER (B)(4) BY A SALES REPRESENTATIVE VIA SEMS-(B)(4)THAT A REVISION REVERSE TSA PROCEDURE OCCURRED ON (B)(6) 2024 FOR UNKNOWN REASONS. DURING THE PROCEDURE, IT WAS REPORTED THAT AN AR-9545-T15-01 DRIVER SHAFT WAS STRIPPED WHEN TRYING TO REMOVE SCREWS FROM THE MGS BASEPLATE. THE CASE WAS COMPLETED SUCCESSFULLY, WITH NO REPORTED PATIENT HARM AND NO FURTHER INFORMATION PROVIDED. NO PIECES BROKE INSIDE THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON WHAT CAUSED THE NECESSARY REVISION SURGERY ON (B)(6) 2024 ON SEMS-(B)(4) FOR CASE NUMBER (B)(4). ADDITIONAL INFORMATION WAS RECEIVED ON 1/31/2024: ON 1/9/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-(B)(4) THAT A PATIENT HAD UNDERGONE A REVISION REVERSE TSA PROCEDURE ON 1/08/2024 DUE TO AN INFECTION. ALL COMPONENTS HAD TO BE REMOVED, AND THE SUTURE CUP HAD TO BE SEPARATED FROM THE STEM FOR EXPLANTATION. THE ORIGINAL PROCEDURE WAS PERFORMED ON (B)(6) 2023 AT THE SAME FACILITY. NO NEW PRODUCTS WERE IMPLANTED AFTER EXPLANTATION. THE CASE WAS COMPLETED USING ANOTHER AR-9545-T15-01 DRIVER SHAFT TO REMOVE THE SUTURE CUP. THE SURGEON STATED THE PATIENT'S CURRENT HEALTH STATUS HAD BEEN R
Patient problems: Unspecified Infection
MAUDE report 18623784 ↗InjuryOct 10, 2022Fracture; Loose or Intermittent Connection
ON (B)(6) 2022, IT WAS REPORTED BY A PATIENT VIA EMAIL THAT SHE UNDERWENT A REVERSE SHOULDER REPLACEMENT ON (B)(6) 2021, AND IS NOW EXPERIENCING SUDDEN PAIN TO HER RIGHT SHOULDER WHEN RAISING HER ARM OVERHEAD. X-RAYS SUGGEST A STRESS FATIGUE FRACTURE THROUGH THE INFERIOR SCREW. THIS DEMONSTRATES LOOSENING OF THE LEFT GLENOSHPERE/BASEPLATE. NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED ON (B)(6) 2021: IMPLANTED DEVICES ARE AS FOLLOWS: RIGHT REVERSE TSR, ARTHREX UNIVERS 6MM STEM WITH 36/+2 RIGHT SUTURE CUP, SMALL 33/+3 HUMERAL INSERT, 24MM MONO-BLOCK BASE PLATE WITH 3 LOCKING SCREWS,+4/33MM GLENOPHERE. A CT SCAN OF THE RIGHT SHOULDER OBTAINED AT LEHIGH VALLEY HOSPITAL IS AVAILABLE FOR REVIEW. THIS DEMONSTRATES APPROPRIATE POSITIONING OF REVERSE TSR COMPONENTS. A MINIMAL AMOUNT OF FLUID IS SEEN. THERE IS DEGENERATIVE CHANGE TO THE AC JOINT. NO LUCENCY AROUND THE IMPLANTS. DUE TO THE ARTIFACT THE SCREW FRACTURE IS NOT VISIBLE. AT THIS TIME A REVISION SURGERY WILL NOT TAKE PLACE. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED (B)(6) 2022: IMPLANTED DEVICES ARE AS FOLLOWS: AR-9560-24P, BATCH 8164; AR-9502F-36RCPC,
Patient problems: Pain
MAUDE report 15570988 ↗InjuryMay 31, 2022Break
ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDER WENT A PRIMARY REVERSE PROCEDURE AND FELL 14 MONTHS LATER, POST OPERATIVE. IMMEDIATELY AFTER THE FALL, PATIENT EXPERIENCE SHOULDER PAIN RADIOGRAPHS SHOWED EVIDENCE OF A BROKEN AR-9563-32 PERIPHERAL LOCKING SCREW, A SHIFTER BASEPLATE AND GLENOSPHERE. A REVISION SURGERY TOOK PLACE ON (B)(6) 2022 FOR REMOVAL OF AR-9563-32 PERIPHERAL LOCKING SCREW, WHICH WAS ONLY PARTIALLY REMOVED, SURGEON MADE THE DECISION TO LEAVE A PORTION OF THE BROKEN SCREW, AS ATTEMPTING TO RETRIEVE IT MAY HAVE RESULTED IN ADDITIONAL BONE LOSS. AN AR-9563-28 PERIPHERAL LOCKING SCREW, AN AR-9561-25S CENTRAL SCREW, AN AR-9560-24-2 BASEPLATE, AN AR-9564-2436-LAT GLENO SPHERE, AND AN AR-9503S-06 HUMERAL INSERT WERE ALL EXPLANTED DURING REVISION REVERSE TSA PROCEDURE ON (B)(6) 2022. SURGEON COMPLETED CASE IMPLANTING NEW GLENOID IMPLANTS.
Patient problems: Fall; Foreign Body In Patient; Pain
MAUDE report 14547369 ↗MalfunctionJan 12, 2021Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED TO RETRIEVE THE PROSTHESIS. IT WAS FURTHER REPORTED THAT THE DEFECTIVE PROSTHESIS WAS CAUSED DUE TO THE POST OPERATIVE ACTIONS BY THE PATIENT. THE PATIENT DID NOT FOLLOW THE POST OP GUIDELINES AND WAS OVERLOADING THE PROSTHESIS TOO EARLY IN THE GYM.
Patient problems: Insufficient Information
MAUDE report 11153804 ↗InjuryMay 6, 2020Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED DURING A REVISION REVERSE SHOULDER TO A CLOSED REDUCTION PROCEDURE ON (B)(6) 2020, THE SURGEON WAS CHECKING IF THE PERIPHERAL SCREWS OF THE 24MM +2 MGS BASEPLATE WERE LOOSE. WHEN THE SURGEON TIGHTENED THE SCREWS, THE BLUE HANDLE OF THE AR-1999HH, RATCHETING HANDLE WAS CLICKING AND DID NOT HAVE A POSITIVE STOP. THE SURGEON DETERMINED THAT THE MGS BASEPLATE WAS LOOSE, SO THEY PROCEEDED TO EXPLANT THE SCREWS AND BASEPLATE WITHOUT ANY ISSUES. THE MGS BASEPLATE WAS IMPLANTED PRIOR TO THE REVISION SURGERY. THE IMPLANTS WERE CONVERTED TO TORNIER. AT THE TIME OF INITIAL REPORT TO ARTHREX NO FURTHER DETAILS WERE AVAILABLE. ADDITIONAL INFORMATION OBTAINED 04/13/2020 AND 04/15/2020: THE INITIAL REPORT PROVIDED WAS NOT A COMPLETE DESCRIPTION AND FURTHER, MORE DETAILED, INFORMATION HAS BEEN OBTAINED TO CLARIFY THE ENTIRE PATIENT RELATED EVENTS. THE FEMALE PATIENT UNDERWENT A PRIMARY REVERSE TOTAL SHOULDER PROCEDURE ON (B)(6) 2019. IT WAS REPORTED THAT THE PATIENT HAD SUFFERED A FALL (DATE UNKNOWN) WHICH RESULTED IN A PERIPROSTHETIC HUMERAL FRACTURE. THE PATIENT UNDERWENT A FIRST REVISION ON (B)(6) 2020 WHERE THE SURGEON EXPLANTED THE FOLLOWING DEVICES AND REPLACED THEM WITH ANOT
Patient problems: No Code Available
MAUDE report 10028355 ↗
Other shoulder prosthesis, reverse configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page