Skip to content

Arthrex® AR-9563-36Recalls & reported problems

Arthrex, Inc. · 5.5x36mm Peripheral Screw, LOCKING · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867296909), received between Sep 16, 2020 and Apr 16, 2026 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Mechanics Altered1
  • Patient Device Interaction Problem1

Patient problems

  • Failure of Implant2
  • Impaired Healing1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 16, 2026Patient Device Interaction Problem

    IT WAS REPORTED THAT AFTER AN INVERSE SHOULDER PROSTHESIS SURGERY ON (B)(6) 2026 THE PATIENT HAS DEVELOPED AN INFECTION. ON (B)(6) 2026 ALL IMPLANTS WERE REMOVED AND THE AFFECTED AREAS WERE CLEANSED AND TREATED. A NEW PROSTHESIS WILL BE IMPLANTED IN 6-8 WEEKS.

    Patient problems: Unspecified Infection

    MAUDE report 24903225 ↗
  • InjuryAug 13, 2025Break; Break

    ON 07/25/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE AR-9580-2420-2S UNIVERS REVERS MODULAR GLENOID SYSTEM AUGMENTED BASEPLATE (24 MM) HAD FAILED, NECESSITATING A REVISION PROCEDURE. THE ISSUE ORIGINATED DURING A REVERSE SHOULDER ARTHROPLASTY PERFORMED ON (B)(6) 2025. TO ADDRESS THE FAILURE, THE SURGICAL TEAM SUCCESSFULLY REMOVED THE AR-9580-2420-2S BASEPLATE (24 MM), AR-9582-30 MODULAR POST FOR AUGMENTED MGS BASEPLATE 30 MM, AR-9563-24 PERIPHERAL LOCKING SCREW, AND AR-9563-36 PERIPHERAL LOCKING SCREW. THEY PROCEEDED WITH THE IMPLANTATION OF A NON-ARTHREX SMALL SCREW BASEPLATE, WHICH WAS SUPPLEMENTED BY BONE GRAFTING. A NEW ARTHREX CUP AND POLYETHYLENE LINER WERE ALSO PLACED. THE REVISION SURGERY WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 07/31/2025.

    Patient problems: Failure of Implant

    MAUDE report 22792977 ↗
  • MalfunctionOct 17, 2022Break

    ON 9/28/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN A PATIENT UNDERWENT A REVERSE MODULAR SHOULDER PROCEDURE ON (B)(6) 2020. DURING THIS PROCEDURE AN AR-9501-13S APEX HUMERAL STEM, AN AR-9502F-39LCPC SUTURE CUP, AN AR-9503M-03 HUMERAL INSERT, AN AR-9560-28-4 MODULAR BASEPLATE, AN AR-9561-35S CENTRAL SCREW, AN AR-9563-36 PERIPHERAL LOCKING SCREW, AN AR-9563-40 PERIPHERAL LOCKING SCREW, AND AN AR-9564-2839-LAT GLENOSPHERE, WERE SUCCESSFULLY IMPLANTED. TWO YEARS LATER, ON (B)(6) 2022, PATIENT UNDERWENT A REVISION REVERSE TOTAL SHOULDER PROCEDURE BECAUSE THE CENTRAL AND PERIPHERAL SCREWS BROKE OFF THE BASEPLATE. ADDITIONAL INFORMATION RECEIVED ON 9/30/2022: DURING REVISION SURGERY ON (B)(6) 2022, AN AR-9560-28-4 MODULAR BASEPLATE, AN AR-9561-35S CENTRAL SCREW, AN AR-9563-36 PERIPHERAL LOCKING SCREW, AND AR-9563-40 PERIPHERAL LOCKING SCREW, AND AN AR-9564-2839-LAT GLENOSPHERE WERE REMOVED. PROCEDURE WAS COMPLETED BY IMPLANTING AN AR-9502-39ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER FROM LOT NUMBER 18.00756 AND AN AR-9550-19RCA UNIVERS REVERS CA HUMERAL HEAD FROM LOT NUMBER 10248799 WERE IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15615280 ↗
  • InjurySep 16, 2020Mechanics Altered

    IT WAS REPORTED THAT (B)(6) 2019 A PATIENT HAD A REVERSE SHOULDER PROCEDURE. ON (B)(6) 2020 THE PATIENT UNDERWENT A REVISION SURGERY. THE PATIENT HAD AN AUTOGRAFT TO THE GLENOID ON HER INITIAL SURGERY AND IT DID NOT APPEAR TO HAVE HEALED. THE SALES REPRESENTATIVE WHO WAS PRESENT DURING THE REVISION REPORTS THAT THE SURGEON HAD MENTIONED IN SURGERY THAT HE BELIEVED THAT THE LOCKING MECHANISM ON THE BASEPLATE MAY NOT HAVE BEEN FUNCTIONING APPROPRIATELY (THE SCREW WAS NOT LOCKED IN). DURING THE ORIGINAL PROCEDURE THE FOLLOWING ARTHREX DEVICES HAD BEEN IMPLANTED: AR-9564-2436-LAT, GLENOSPHERE, LOT 18.00689. AR-9503S-03, HUMERAL INSERT, LOT 18.00870 . AR-9502F-36CPC, UNIVERS REVERS SUTURE CUP, LOT 170141708 . AR-9561-25P, MODULAR CENTRAL POST, LOT 5368 . AR-9560-24-2, MODULAR BASEPLATE, LOT 5408. AR-9562-36NL, PERIPHERAL SCREW NON-LOCKING, LOT 2018001207. AR-9563-36, PERIPHERAL SCREW LOCKING, LOT 2018001490. AR-9563-40, PERIPHERAL SCREW LOCKING, LOT 2018001491 . AR-9501-12P, UNIVERS REVERS HUMERAL STEM SIZE 12, LOT 170125705 . DURING THE REVISION SURGERY ALL OFF THE ABOVE LISTED ORIGINAL IMPLANTS WERE EXPLANTED, EXCEPT FOR THE AR-9501-12P, UNIVERS REVERS HUMERAL STEM SIZE 12, WHICH WAS

    Patient problems: Impaired Healing

    MAUDE report 10540490 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page