Skip to content

Arthrex® AR-4158DS-14BRecalls & reported problems

Arthrex, Inc. · DynaNite PIP - Bent 14mm - w/instr · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867310001), received between Apr 26, 2024 and Oct 15, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Component Missing1
  • Defective Device1
  • Patient Device Interaction Problem1

Patient problems

  • Failure of Implant2
  • Arthralgia1
  • Insufficient Information1
  • Malunion of Bone1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 15, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    ON 9/19/2025, A FACILITY REPRESENTATIVE REPORTED THAT A PATIENT ENROLLED IN THE CLINICAL STUDY 'PROSPECTIVE STUDY OF HAMMER TOE FIXATION USING AN INTRAMEDULLARY NITINOL IMPLANT' DEVELOPED MALLET TOES IN ALL OPERATIVE TOES; HOWEVER, THE IMPLANTS ARE INTACT. ADDITIONAL INFORMATION WAS RECEIVED ON 10/3/2025. THE PATIENT UNDERWENT INITIAL SURGERY ON (B)(6) 2023, DURING WHICH AN AR-4158DS-16B DYNANITE PIP, AN AR-4158P-16B DYNANITE PIP, AN AR-4158DS-14B DYNANITE PIP, AND AN AR-4158P-12B DYNANITE PIP WERE PLACED. ON (B)(6) 2023, THE HEALTHCARE PROVIDER EXPRESSED CONCERN ABOUT A POSSIBLE RECURRENCE OF DEFORMITY AND RECOMMENDED EXTENSION EXERCISES. AT THE (B)(6) 2023 VISIT, THE PATIENT WAS NOTED TO HAVE PASSIVELY CORRECTABLE CLAW TOES. BY (B)(6) 2024, MALLET TOE DEFORMITIES WERE OBSERVED IN TOES 2, 3, 4, AND 5, ALONG WITH PAIN IN THE PLANTAR PLATE AREA OF THOSE TOES. AN MRI WAS RECOMMENDED TO ASSESS FOR POSSIBLE PLANTAR PLATE TEARS, BUT THE PATIENT DECLINED. THROUGHOUT THESE VISITS, THE PATIENT EXPERIENCED MILD PAIN AND STIFFNESS IN THE TOES. NO TREATMENT WAS PURSUED FOR THE MALLET TOES, AND THE HEALTHCARE PROVIDER ADVISED THE PATIENT TO RETURN IF PAIN WORSENED. THE PATIENT HAS NOT RETURNED

    Patient problems: Arthralgia; Failure of Implant; Malunion of Bone

    MAUDE report 23277873 ↗
  • MalfunctionJun 4, 2024Defective Device; Defective Device

    ON 1/4/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT THE 3.0MM PROXIMAL DRILL BIT FROM AN AR-4158DS-14B DYNANITE PIP KIT, DID NOT FIT OVER THE K-WIRE. ANOTHER KIT WAS OPENED AND THE NEW 3.0MM DRILL BIT WAS USED WITH THE K-WIRE FROM THE PREVIOUS KIT. THIS WAS DISCOVERED DURING A HAMMERTOE PROCEDURE ON (B)(6) 2023. CASE WAS COMPLETED SUCCESSFULLY, AND NO FURTHER ISSUES HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19457345 ↗
  • MalfunctionMay 6, 2024Component Missing; Component Missing

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-4158DS-14B DYNANITE PIP KIT CONTAINED (2) 3.0 CANNULATED DRILL BITS RATHER THAN (1)2.5 AND (1)3.0. THIS WAS DISCOVERED DURING A HAMMER TOE PROCEDURE ON (B)(6) 2024, WHERE THE IMPLANT WAS STILL USED ALONG WITH A DRILL BIT FROM THE AR-4158DS-12B.

    Patient problems: Insufficient Information

    MAUDE report 19247518 ↗
  • MalfunctionApr 26, 2024Break; Break

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-4158DS-14B DYNANITE PIP HAMMERTOE IMPLANT HAS BROKEN AFTER NINE WEEKS OF POST OPERATION. THE IMPLANT APPEARS TO BE FUSED AND THE SURGEON IS NOT PLANNING ON REMOVING IT AS OF NOW. THIS WAS DISCOVERED DURING A POST OPERATIVE VISIT ON (B)(6) 2022, VIA X-RAYS. ADDITIONAL INFORMATION HAS BEEN RECEIVED ON 6/14/2022: THE PROCEDURE TOOK PLACE ON (B)(6) 2022 AND IT WAS COMPLETED SUCCESSFULLY.

    Patient problems: Failure of Implant

    MAUDE report 19190746 ↗

Other PIN, FIXATION, SMOOTH devices

All PIN, FIXATION, SMOOTH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page