Arthrex® AR-9580-2410Recalls & reported problems

Arthrex, Inc. · 24MM BASEPLATE, 10° FULL AUG, OB · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867341630), received between Mar 15, 2021 and Jan 13, 2025 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Connection Problem1
  • Defective Device1
  • Fitting Problem1
  • Loose or Intermittent Connection1
  • Material Separation1
  • Positioning Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Bone Fracture(s)1
  • Insufficient Information1
  • Pulmonary Embolism1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2025Fitting Problem; Fitting Problem

    ON (B)(6) 2024 , IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-9580-2410 UNIVERS REVERS MODULAR GLENOID SYSTEM AUGMENTED BASEPLATE AND AN AR-9582-20 MODULAR POST FOR AUGMENTED MGS BASEPLATE WOULD NOT ENGAGE WITH THE TAPER. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2024 .

    Patient problems: Insufficient Information

    MAUDE report 21138062 ↗
  • MalfunctionMay 8, 2024Connection Problem; Connection Problem

    ON 10/27/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A AR-9580-2410 AUGMENTED BASE PLATE, AND A AR-9582-25 MODULAR POST FOR AUGMENTED MGS BASEPLATE ARE NOT LOCKING. THIS OCCURRED ON (B)(6) 2022 DURING A REVERS TOTAL SHOULDER ARTHROPLASTY WHEN THE (B)(4) DEVICES WOULD NOT CONNECT TOGETHER. ANOTHER DEVICE OF EACH WERE OPENED, AND WORKED ACCORDINGLY. THE CASE WAS COMPLETED, WITH NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19267499 ↗
  • MalfunctionNov 2, 2023Defective Device; Defective Device

    ON 10/9/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-9582-20 MODULAR POST, 20MM, AND AN AR-9580-2410 24MM BASEPLATE WERE NOT LOCKING TOGETHER DURING A PROCEDURE. THE CASE WAS COMPLETED USING ANOTHER OF THE SAME DEVICES WITH NO REPORTED DELAY IN THE PROCEDURE. THIS WAS DISCOVERED DURING A REVERSE SHOULDER ARTHROPLASTY PROCEDURE ON 10/9/2023 WITH NO ADVERSE EVENT OR PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18054452 ↗
  • MalfunctionOct 11, 2022Material Separation; Positioning Problem

    IT WAS REPORTED ON (B)(6) 2022 BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-9580-2410 MGS BASEPLATE AND AR-9582-25 MODULAR POST WOULD NOT SEAT AND WERE UNABLE TO BE RE-COMPRESSED. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM. PER FACILITY: "AFTER COMPRESSING THE MGS AUGMENTED BASEPLATE AND THE MODULAR POST TOGETHER, THE IMPLANT WAS INSERTED INTO THE GLENOID. IT WAS NOT SEATING, AND WHEN REMOVED, THE TWO PIECES CAME APART. WE WERE THEN UNABLE TO RE-COMPRESS THEM AND FORCED TO OPEN NEW IMPLANTS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15577555 ↗
  • InjuryJun 7, 2021Loose or Intermittent Connection; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE SURGEON WAS ATTEMPTING TO DO A 10 DEGREE AUGMENTED MGS GLENOID WITH A 35MM CENTRAL POST. AFTER IMPACTING THE GLENOSPHERE ONTO THE BASEPLATE, THE PATIENT'S GLENOID FRACTURED AND THE ENTIRE CONSTRUCT CAME LOOSE (SEE ATTACHED PICTURE). THE SURGEON WAS NOT SURE IF THEY WERE TOO ANTERIOR WITH THEIR INITIAL GUIDE PIN AND SUBSEQUENT REAMING, POSSIBLY WEAKENING THE BONE ANTERIORLY. THE SURGEON ALSO WENT BI-CORTICAL IN THE VAULT WITH THE CENTRAL POST (B)(4)) AND DID NOT KNOW IF THIS ULTIMATELY LED TO THE FAILURE. THE SURGEON WAS ABLE TO USE BONE GRAFT FROM THE PATIENT'S RESECTED HUMERAL HEAD TO ADDRESS THE GLENOID FRACTURE AND THEN IMPLANTED A NON-AUGMENTED MGS BASEPLATE AND FINISHED THE PROCEDURE WITHOUT FURTHER ISSUE.

    Patient problems: Bone Fracture(s)

    MAUDE report 11946802 ↗
  • InjuryMar 15, 2021Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE SURGICAL OUTCOME SYSTEM THAT A PULMONARY EMBOLISM RESULTING IN INPATIENT HOSPITALIZATION OCCURRED DURING A SHOULDER ARTHROPLASTY PROCEDURE IN WHICH ARTHREX PRODUCT WAS BEING USED. THE PATIENT WAS TREATED WITH CONCOMITANT MEDICATION. THE FACILITY REPORTED THAT THE PATIENT HAS BEEN DISCHARGED FROM THE HOSPITAL AND IS RECOVERING.

    Patient problems: Pulmonary Embolism

    MAUDE report 11484215 ↗

Other shoulder prosthesis, reverse configuration devices