Skip to content

Arthrex® AR-9582-35Recalls & reported problems

Arthrex, Inc. · MODULAR POST, 35MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867341869), received between Jun 7, 2021 and Jan 12, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Use of Device Problem1

Patient problems

  • Bone Fracture(s)1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 12, 2026Use of Device Problem

    ON 22-DEC-2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-9582-35 MODULAR POST FOR AUGMENTED MGS BASEPLATE 35 MM EXPERIENCED AN ISSUE WITH INSERTING A SCREW INTO THE BASEPLATE. THIS OCCURRED DURING A REVERSE TOTAL SHOULDER REPLACEMENT ON (B)(6) 2025. THIS RESULTED IN HAVING TO REMOVE THE AR-9582-35 AND AN AR-9580-2420-2. THE CASE CONTINUED USING ANOTHER AR-9582-35 AND ANOTHER AR-9580-2420-2. THERE WAS NO DELAY IN THE PROCEDURE.

    Patient problems: Insufficient Information

    MAUDE report 24035272 ↗
  • InjuryJun 7, 2021Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SURGEON WAS ATTEMPTING TO DO A 10 DEGREE AUGMENTED MGS GLENOID WITH A 35MM CENTRAL POST. AFTER IMPACTING THE GLENOSPHERE ONTO THE BASEPLATE, THE PATIENT'S GLENOID FRACTURED AND THE ENTIRE CONSTRUCT CAME LOOSE (SEE ATTACHED PICTURE). THE SURGEON WAS NOT SURE IF THEY WERE TOO ANTERIOR WITH THEIR INITIAL GUIDE PIN AND SUBSEQUENT REAMING, POSSIBLY WEAKENING THE BONE ANTERIORLY. THE SURGEON ALSO WENT BI-CORTICAL IN THE VAULT WITH THE CENTRAL POST ((B)(4)) AND DID NOT KNOW IF THIS ULTIMATELY LED TO THE FAILURE. THE SURGEON WAS ABLE TO USE BONE GRAFT FROM THE PATIENT'S RESECTED HUMERAL HEAD TO ADDRESS THE GLENOID FRACTURE AND THEN IMPLANTED A NON-AUGMENTED MGS BASEPLATE AND FINISHED THE PROCEDURE WITHOUT FURTHER ISSUE.

    Patient problems: Bone Fracture(s)

    MAUDE report 11946788 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page