Skip to content

Arthrex® AR-9504-06Recalls & reported problems

Arthrex, Inc. · Humeral Spacer 33+6mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867406902), received between Jul 20, 2023 and Mar 19, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure2
  • Patient Device Interaction Problem1

Patient problems

  • Insufficient Information2
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2025Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION RECEIVED ON 3/24/025: AR-9564-2433-LAT ARTHREX UNIVERS REVERS GLENOSPHERE WERE EXPLANTED AS WELL. AR-9503-3336-3 UNIVERS REVERS COMBO HUMERAL INSERT WAS ATTEMPTED TO BE USED WITH THE SPACER BUT IT WOULD NOT FIT, SO THE SURGEON TURNED TO AN AR-9503-3336-6C UNIVERS REVERS MODULAR GLENOID SYSTEM, CONSTRAINED COMBO HUMERAL INSERT . IT WAS ALSO REPORTED THAT THE SCREW FROM THE INITIAL AR-9504-06 UNIVERS REVERS SPACER WAS NOT WORKING AND AN AR-9502RSC UNIVERS REVERS REPLACEMENT CUP SCREW HAD TO BE USED.

    Patient problems: Insufficient Information

    MAUDE report 21641445 ↗
  • MalfunctionMar 18, 2025Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION RECEIVED ON 3/24/025: AR-9564-2433-LAT ARTHREX UNIVERS REVERS GLENOSPHERE WERE EXPLANTED AS WELL. AR-9503-3336-3 UNIVERS REVERS COMBO HUMERAL INSERT WAS ATTEMPTED TO BE USED WITH THE SPACER, BUT IT WOULD NOT FIT, SO THE SURGEON TURNED TO AN AR-9503-3336-6C UNIVERS REVERS MODULAR GLENOID SYSTEM, CONSTRAINED COMBO HUMERAL INSERT. IT WAS ALSO REPORTED THAT THE SCREW FROM THE INITIAL AR-9504-06 UNIVERS REVERS SPACER WAS NOT WORKING AND AN AR-9502RSC UNIVERS REVERS REPLACEMENT CUP SCREW HAD TO BE USED.

    Patient problems: Insufficient Information

    MAUDE report 21633534 ↗
  • InjuryJul 20, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ON 6/26/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDERWENT A REVISION SHOULDER ARTHROPLASTY PROCEDURE ON (B)(6) 2023 DUE TO AN INFECTION AND LOOSENED BASEPLATE. ON (B)(6) 2023, THE PATIENT COMPLAINED OF PAIN AFTER THE ORIGINAL PROCEDURE ON (B)(6) 20203. DURING THE POST-OPERATIVE VISIT, RADIOGRAPHS INDICATED A LOOSENED BASEPLATE, BUT THE INFECTION TYPE WAS NOT YET CONFIRMED. DURING THE REVISION SURGERY, AN AR-9564-2433-LAT GLENOSPHERE, AN AR-9580-2420 BASEPLATE, AN AR-9582-20 MODULAR POST, AN AR-9501-10S HUMERAL STEM, AN AR-9502F-33CPC SUTURE CUP, AN AR-9503XS-03 HUMERAL INSERT, AN AR-9504-06 HUMERAL SPACER, AN AR-9562-24NL PERIPHERAL NON-LOCKING SCREW, AN AR-9563-16 PERIPHERAL LOCKING SCREW, AN AR-9563-20 PERIPHERAL LOCKING SCREW, AND AR-9563-24 PERIPHERAL LOCKING SCREW WERE EXPLANTED. THE CASE WAS COMPLETED BY TAKING CULTURES FOR FURTHER TESTING AND PLACING ANTIBIOTIC SPACERS. NO ADDITIONAL ARTHREX PRODUCTS WERE IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 17363965 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page