CHATTANOOGA 2502Recalls & reported problems
Djo, LLC · HYDROCOLLATOR, M-4, C/W 24 STD. PACKS · Class 1
All models of CHATTANOOGA (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888912021463), received between Aug 21, 2019 and Dec 30, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Overheating of Device2
Patient problems
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT "PHYSICAL INSPECTION REVEALED NEUTRAL PRONG ON CORD WAS MELTED.. [R]RECEPTACLE WAS MELTED AS WELL AND NEUTRAL PRONG FELL OFF. [I]INTERNAL WIRING HAS EVIDENCE OF HEAT DAMAGE". THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9529950 ↗MalfunctionAug 21, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT "WHILE THE STAFF WAS ALREADY OUT FOR THE DAY, THE UNIT OVERHEATED AND THE OVER-TEMP THERMOSTAT DID NOT SHUT DOWN THE UNIT; NO ONE WAS INJURED; THE POWER CORD MELTED". [SIC] NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 8914860 ↗
Other UNIT, HEATING, POWERED devices
- CHATTANOOGA 2402Djo, LLC · UNIT, HEATING, POWERED2 reports
- CHATTANOOGA 2802Djo, LLC · UNIT, HEATING, POWERED1 report
- CHATTANOOGA 2302Djo, LLC · UNIT, HEATING, POWERED1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page