Skip to content

CHATTANOOGA 2502Recalls & reported problems

Djo, LLC · HYDROCOLLATOR, M-4, C/W 24 STD. PACKS · Class 1

All models of CHATTANOOGA (39)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888912021463), received between Aug 21, 2019 and Dec 30, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Overheating of Device2

Patient problems

  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 30, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT "PHYSICAL INSPECTION REVEALED NEUTRAL PRONG ON CORD WAS MELTED.. [R]RECEPTACLE WAS MELTED AS WELL AND NEUTRAL PRONG FELL OFF. [I]INTERNAL WIRING HAS EVIDENCE OF HEAT DAMAGE". THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9529950 ↗
  • MalfunctionAug 21, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT "WHILE THE STAFF WAS ALREADY OUT FOR THE DAY, THE UNIT OVERHEATED AND THE OVER-TEMP THERMOSTAT DID NOT SHUT DOWN THE UNIT; NO ONE WAS INJURED; THE POWER CORD MELTED". [SIC] NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 8914860 ↗

Other UNIT, HEATING, POWERED devices

All UNIT, HEATING, POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page