DJO SURGICAL 497-36-000Recalls & reported problems

Encore Medical, L.P. · HEAD, FEMORAL, COCR 36MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DJO Surgical hip implant is a metal femoral head used in total hip replacement surgery. The 37 adverse event reports mostly describe hip dislocation or the joint coming apart or feeling unstable after implantation, along with pain, falls and infection, with 37 injuries and no deaths. Most reports led to revision surgery to replace or adjust the hip components. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
37
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00888912024051), received between Sep 2, 2015 and Mar 17, 2026 — classified as 37 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated12
  • Adverse Event Without Identified Device or Use Problem10
  • Appropriate Device Problem Term/Code Not Available8
  • Unstable4
  • Fracture2
  • Disassembly1
  • Excessive Cooling1
  • Loose or Intermittent Connection1

Patient problems

  • No Code Available14
  • Joint Dislocation5
  • Unspecified Infection4
  • Fall3
  • Pain3
  • Insufficient Information2
  • Joint Laxity2
  • No Information2

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem

    THIS PATIENT HAD A TOTAL HIP ON 6/26/2014. THE PATIENT SUSTAINED A FALL YESTERDAY AND DISLOCATED THE HIP. DR. YACOUBIAN WAS NOT ABLE TO DO A CLOSED REDUCTION AND NEEDED TO DO AN OPEN REDUCTION. HE ELECTED TO REMOVED THE 36ID MP8 10 DEG LINER, 36 METAL HEAD AND -3.5 SLEEVE AND REPLACE THEM WITH A MP8 40ID 10 DEGREE LINER, 40 METAL HEAD AND -3.5 SLEEVE.

    Patient problems: Joint Dislocation

    MAUDE report 24622300 ↗
  • InjurySep 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY: DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 23059685 ↗
  • InjurySep 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 20177225 ↗
  • InjurySep 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 17770195 ↗
  • InjuryApr 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY ,METAL ON METAL WEAR.

    Patient problems: Joint Laxity

    MAUDE report 16773678 ↗
  • InjuryNov 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Medical device site infection

    MAUDE report 15865055 ↗
  • InjuryAug 4, 2022Mechanical Problem

    REVISION SURGERY - THE PATIENT WAS COMPLAINING ABOUT DISCOMFORT AND LEG LENGTH. SURGEON SHORTENED HER WITH A NEUTRAL HEAD.

    Patient problems: Discomfort

    MAUDE report 15165351 ↗
  • InjuryJan 16, 2022Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Insufficient Information

    MAUDE report 13262911 ↗
  • InjuryDec 6, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - PATIENT HAD CHRONIC DISLOCATION. REMOVED AND REPLACED THE HEAD, LINER AND SLEEVE.

    Patient problems: Joint Dislocation

    MAUDE report 12936968 ↗
  • InjurySep 17, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - PATIENT HAD INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 12491973 ↗
  • InjuryJul 19, 2021Device Dislodged or Dislocated

    REVISION SURGERY - PATIENT DISLOCATED HIP AFTER FALLING. REPLACED ORIGINAL HEAD, LINER AND SLEEVE.

    Patient problems: Fall

    MAUDE report 12189310 ↗
  • InjuryFeb 25, 2021Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Insufficient Information

    MAUDE report 11379860 ↗
  • InjuryJun 19, 2020Fracture; Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - THE PATIENT DISLOCATED POST-OPERATION AND FRACTURED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10173736 ↗
  • InjuryMar 25, 2020Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - PATIENT HAD INFECTION. SURGEON WASHED OUT AND EXCHANGED OUT THE HEAD.

    Patient problems: Unspecified Infection

    MAUDE report 9878910 ↗
  • InjuryDec 3, 2019Unstable; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - PATIENT WAS UNSTABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9401894 ↗
  • InjuryNov 14, 2019Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: No Code Available

    MAUDE report 9324293 ↗
  • InjuryAug 1, 2019Device Dislodged or Dislocated

    REVISION SURGERY, HIP DISLOCATED. REMOVED AND REPLACED THE HEAD, SLEEVE AND LINER.

    Patient problems: Failure of Implant

    MAUDE report 8851259 ↗
  • InjuryJun 5, 2019Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO THE PATIENT HAVING INFECTION IN HIP; THE SURGEON REMOVED AND REPLACED ALL IMPLANTS.

    Patient problems: Unspecified Infection

    MAUDE report 8671942 ↗
  • InjuryJan 10, 2019Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: No Information

    MAUDE report 8236846 ↗
  • InjuryAug 24, 2018Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO THE PATIENT COMPLAINING OF GENERAL LATERAL THIGH PAIN NEAR GREATER TROCHANTER. ACCORDING TO THE REVISION SURGERY IT APPEARED THE ORIGINAL FEMORAL STEM WAS IMPLANTED IN "VARUS" WHICH MAY HAVE BEEN A CONTRIBUTING FACTOR TO PAIN. THE ORIGINAL STEM, HEAD AND LINER WERE REMOVED AND REPLACED.

    Patient problems: No Code Available; Pain

    MAUDE report 7814546 ↗
  • InjuryFeb 23, 2018Excessive Cooling

    REVISION SURGERY - DUE TO CHRONIC DISLOCATION.

    Patient problems: No Code Available

    MAUDE report 7292321 ↗
  • InjuryOct 27, 2017Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: No Code Available

    MAUDE report 6985216 ↗
  • InjuryOct 3, 2017Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: No Code Available

    MAUDE report 6911226 ↗
  • InjurySep 22, 2017Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO THE PATIENT PRESENTING AT THE DOCTOR 'S OFFICE WITH PAIN IN HIP. THE REVISION SURGERY WAS SCHEDULED TO PERFORM A HEAD AND LINER EXCHANGE WITH WASH OUT FOR POSSIBLE INFECTION. THE STEM AND HEAD WERE DJO PRODUCTS THE CUP AND LINER WERE MAKO. ONLY A REPLACEMENT HEAD WAS REMOVED AND REPLACED WITH DJO SURGICAL.

    Patient problems: Pain

    MAUDE report 6888538 ↗
  • InjurySep 13, 2017Device Dislodged or Dislocated

    SECOND REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 6861927 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices →