Skip to content

DJO SURGICAL 506-00-006Recalls & reported problems

Encore Medical, L.P. · STEM, HUMERAL SIZE6 RSP · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00888912024372), received between Sep 4, 2015 and May 27, 2026 — classified as 15 injuries.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (7 reports).

4 other recalls cover devices cleared under the same FDA submission (K041066); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loose or Intermittent Connection7
  • Fracture3
  • Device Dislodged or Dislocated2
  • Unstable2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Device Slipped1

Patient problems

  • No Code Available6
  • Failure of Implant3
  • Joint Laxity2
  • No Known Impact Or Consequence To Patient2
  • Bone Fracture(s)1
  • Fall1
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 27, 2026Unstable

    REVISION SURGERY DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 25311003 ↗
  • InjuryOct 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO IMPLANTS BECAME LOOSE OVER THE YEARS. REMOVED ALL ORIGINAL IMPLANTS.

    Patient problems: Joint Laxity

    MAUDE report 23444155 ↗
  • InjuryNov 16, 2021Fracture; Device Dislodged or Dislocated

    REVISION SURGERY - PATIENT HAD PERI-PROSTHETIC FRACTURE AND WAS DISLOCATED.

    Patient problems: Insufficient Information

    MAUDE report 12823872 ↗
  • InjuryOct 29, 2019Device Slipped

    REVISION SURGERY - DUE TO POSSIBLE INFECTION AND LOOSE STEM.

    Patient problems: Unspecified Infection

    MAUDE report 9249626 ↗
  • InjuryJul 19, 2019Fracture; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - PATIENT HAD A PERI-PROSTHETIC FRACTURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8809404 ↗
  • InjuryJan 4, 2019Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8219772 ↗
  • InjuryNov 30, 2018Loose or Intermittent Connection

    REVISION SURGERY - DUE TO THE STEM BECOMING LOOSE.

    Patient problems: Failure of Implant

    MAUDE report 8120483 ↗
  • InjuryJun 13, 2018Loose or Intermittent Connection

    REVISION SURGERY - DUE TO A LOOSE STEM CAUSED FROM CEMENT FAILURE TO FIXATE TO BONE.

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 7598262 ↗
  • InjuryMay 4, 2018Loose or Intermittent Connection

    REVISION SURGERY - DUE TO THE STEM LOOSENING.

    Patient problems: No Code Available

    MAUDE report 7488817 ↗
  • InjuryFeb 13, 2018Fracture

    REVISION SURGERY - DUE TO THE PATIENT NEEDING A LONG REVERSE STEM. STEM BROKE DUE TO A FRACTURE.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 7267866 ↗
  • InjuryJan 8, 2018Loose or Intermittent Connection

    REVISION SURGERY - DUE TO THE STEM LOOSENING OVER TIME.

    Patient problems: No Information

    MAUDE report 7173284 ↗
  • InjuryOct 3, 2017Loose or Intermittent Connection

    REVISION SURGERY - DUE TO THE STEM BEING CEMENTED IN AND IT BEING LOOSE.

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 6911245 ↗
  • InjurySep 11, 2017Break

    REVISION SURGERY - THE SURGEON REMOVED THE STEM AND INSERT. THE PATIENT FELL AND BROKE BONE, FRACTURING THE STEM.

    Patient problems: Fall

    MAUDE report 6855004 ↗
  • InjuryJan 6, 2017Loose or Intermittent Connection

    REVISION SURGERY - DUE TO THE NEUTRAL CUP FORM THE ORIGINAL SURGERY COMING LOOSE. THE SURGEON REMOVED ALL OF THE COMPONENTS AFTER ATTEMPTING TO REVISE. HE PUT IN A MONOBLOCK.

    Patient problems: No Code Available

    MAUDE report 6231058 ↗
  • InjurySep 4, 2015Loose or Intermittent Connection; Unstable

    REVISION SURGERY - DUE TO THE PATIENT'S HUMERAL COMPONENT LOOSENING. IT WAS CEMENTED INTO PLACE IN THE ORIGINAL SURGERY, BUT HAD SINCE BECOME UNSTABLE AND LOOSE. THE SURGEON REMOVED THE HUMERAL STEM, HUMERAL SOCKET SHELL AND HUMERAL SOCKET INSERT AND REVISED WITH NEW COMPONENTS.

    Patient problems: No Code Available

    MAUDE report 5058312 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041066. One submission often covers many models, so check the product and lot details in each record.

  • Z-0098-2024 · Aug 22, 2023 · Open, Classified
    REVERSE PROSTHESIS SHOULDER SYSTEM, SCREW, LOCKING BONE, RSP, 5mmX30mm long, REF: 506-03-130
  • Z-0185-2008 · Aug 8, 2007 · Terminated
    Encore Reverse Shoulder Prosthesis (RSP) System, Standard Size 40 RSP Humeral Socket Insert; Catalog #: 508-00-040; Sterile; Manufactured by Encore, Medical, L.P., Austin, TX 7875
  • Z-0184-2008 · Aug 8, 2007 · Terminated
    Encore Reverse Shoulder Prosthesis (RSP) System, Standard Size 36 RSP Humeral Socket Insert; Catalog #: 508-00-036; Sterile; Manufactured by Encore, Medical, L.P., Austin, TX 7875
  • Z-1566-05 · Jul 28, 2005 · Terminated
    Encore Reverse Shoulder Prosthesis (RSP), manufactured and distributed by Encore Medical, L.P. 9800 Metric Blvd., Austin, Texas 78758

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →