DJO SURGICAL 506-00-010Recalls & reported problems
Encore Medical, L.P. · STEM, HUMERAL 10MM RSP · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00888912024396), received between Jan 15, 2016 and Nov 4, 2025 — classified as 15 injuries.
The most frequently reported specific device problem is “Device Slipped” (2 reports).
5 other recalls cover devices cleared under the same FDA submission (K051075); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Device Slipped2
- Loose or Intermittent Connection2
- Unstable2
- Appropriate Device Problem Term/Code Not Available1
- Detachment of Device or Device Component1
- Device Operates Differently Than Expected1
- Fracture1
Patient problems
- No Code Available7
- Joint Laxity2
- Unspecified Infection2
- Bone Fracture(s)1
- Fall1
- Insufficient Information1
- Pain1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryNov 4, 2025Unstable; Unstable
InjuryJul 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryJun 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryFeb 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION SURGERY - REQUIRED INTERVENTION TO PREVENT IMPAIRMENT/DAMAGE.
Patient problems: Fall
MAUDE report 16342825 ↗InjuryNov 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION SURGERY - PATIENT HAD A DISTAL FRACTURE ON THE HUMERUS. TO REVISE SURGEON REMOVED THE 36 POLY AND CONVERSION SHELL AS WELL AS THE STEM.
Patient problems: Bone Fracture(s)
MAUDE report 15873578 ↗InjurySep 8, 2021Loose or Intermittent Connection
REVISION SURGERY - DUE TO LOOSE STEM.
Patient problems: Insufficient Information
MAUDE report 12439739 ↗InjuryJul 30, 2018Unstable; Appropriate Device Problem Term/Code Not Available
REVISION SURGERY - DUE TO INSTABILITY. THE PATIENT PREVIOUSLY HAD A REVERSE DONE AND CAME TO SEE REVISION SURGEON TO HELP STABILIZE. THE PREVIOUS SURGEON DID NOT HAVE IMPLANTS IN CORRECT POSITION. THE SURGEON WAS UNABLE TO DO 2ND REVERSE BECAUSE OF SUBSCAPULARIS FAILURE, INSTABILITY.
Patient problems: No Code Available
MAUDE report 7732666 ↗InjuryJan 22, 2018Device Slipped
REVISION SURGERY - DUE TO THE TAPER DISENGAGING AND THE STEM BEING LOOSE AS WELL.
Patient problems: No Code Available
MAUDE report 7209807 ↗InjuryOct 24, 2017Loose or Intermittent Connection
REVISION SURGERY - DUE TO THE PATIENT EXPERIENCING PAIN. THE STEM WAS LOOSE; THE SURGEON REMOVED EVERYTHING EXCEPT FOR THE BASEPLATE AND ONE SIZE 18MM SCREW.
Patient problems: Pain
MAUDE report 6974564 ↗InjuryOct 4, 2017Insufficient Device Problem Information
REVISION SURGERY - DUE TO THE SURGEON HAVING INDICATED THAT THE STEM WAS LOOSE AND NEEDED TO BE REMOVED. OUR REPRESENTATIVE WAS TOLD THE ORIGINAL SURGERY WAS DONE ABOUT 5 YEARS AGO AND NO OTHER PATIENT INFORMATION WAS GIVEN REGARDING IT.
Patient problems: No Code Available
MAUDE report 6914300 ↗InjuryJul 20, 2017Detachment of Device or Device Component
REVISION SURGERY - DUE TO THE SHELL DISASSOCIATING FROM THE STEM ON THE HUMERAL SIDE.
Patient problems: No Code Available
MAUDE report 6730528 ↗InjuryMar 16, 2017Adverse Event Without Identified Device or Use Problem
REVISION SURGERY - DUE TO THE PATIENT DEVELOPING AN INFECTION THE REVERSE SHOULDER PROSTHESIS (RSP) PRIMARY COMPONENTS WERE REMOVED AND A TEMPORARY ANTIBIOTIC SPACER WAS PLACED IN. SUBSEQUENTLY, A REVERSE SHOULDER WAS PLACED UPON THE HEALING OF THE INFECTION.
Patient problems: Unspecified Infection
MAUDE report 6411299 ↗InjuryApr 5, 2016Material Separation; Device Operates Differently Than Expected
REVISION SURGERY - DUE TO PRODUCT FAILURE; THE MORSE TAPER BETWEEN SHELL AND STEM DISASSOCIATED.
Patient problems: No Code Available
MAUDE report 5548480 ↗InjuryMar 1, 2016Fracture
REVISION SURGERY - DUE TO THE PATIENT SUFFERING A FRACTURE AND A SUSPECTED INFECTION.
Patient problems: No Code Available
MAUDE report 5472053 ↗InjuryJan 15, 2016Device Slipped
REVISION SURGERY - DUE TO THE HUMERAL STEM BECOMING LOOSE POST OPERATIVE. THE SURGEON REMOVED ALL OF THE COMPONENTS AND REPLACED THEM WITH A HEMI ARTHROPLASTY.
Patient problems: No Code Available
MAUDE report 5370530 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051075. One submission often covers many models, so check the product and lot details in each record.
- Z-1460-2026 · Jan 15, 2026 · Open, ClassifiedBrand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP STANDARD HUMERAL SOCKET INSERT, 36MM, HXe-plus Model/Catalog Number: 509-00-036
- Z-0625-2024 · Oct 5, 2023 · CompletedReverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797
- Z-0282-2024 · Sep 25, 2023 · Open, ClassifiedRSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.
- Z-2606-2023 · Aug 4, 2023 · Open, ClassifiedDJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder Syst
- Z-2739-2011 · Apr 18, 2011 · TerminatedReverse Shoulder Prosthesis System RSP Humeral Socket Shell DJO 9800 Metric Blvd. Austin, TX 78758 STERILE R Model numbers: 508-00-000 and 508-00-008. The Reverse Shoulder Prost
Other shoulder prosthesis, reverse configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page