DJO SURGICAL 508-00-001Recalls & reported problems
Encore Medical, L.P. · SCREW, RETAINING GLENOID HEAD RSP · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888912024617), received between Sep 20, 2016 and Sep 20, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K051075); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- No Apparent Adverse Event2
- Break1
- Detachment Of Device Component1
Patient problems
- Foreign Body In Patient2
- No Known Impact Or Consequence To Patient1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 20, 2024Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
REVISION SURGERY - DUE TO WHEN INSERTING THE SET SCREW INTO THE GLENOSPHERE WITH THE RED HANDLE TORQUE SCREWDRIVER THE SCREW BROKE OFF HALFWAY. BROKEN/FRAGMENTED PIECES LEFT IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20274343 ↗InjuryJan 17, 2024No Apparent Adverse Event; No Apparent Adverse Event
INSTRUMENT FAILURE - PIECES LEFT IN PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 18531774 ↗InjuryMay 16, 2017Break; Detachment Of Device Component
PRIMARY SURGERY - WHILE THE SURGEON WAS TURNING THE SCREW, IT OVER TORQUED AND THE HEAD OF THE SCREW SNAPPED OFF. THE SCREW IS NOT IN THE PATIENT BUT THE REMAINING SHAFT OF THE SCREW STAYED IN THE BASEPLATE/GLENOSPHERE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6569407 ↗InjurySep 20, 2016Adverse Event Without Identified Device or Use Problem
REVISION SURGERY - DUE TO THE PATIENT PRESENTING AFTER A REVISION IN PAIN AND INFECTED. THE SURGEON DECIDED TO GO IN AND REMOVE ALL THE GLENOID COMPONENTS, THE 44 POLY AS WELL AS THE +8 SPACER. HE THEN DECIDED TO PUT IN A HEMI SPACER AND A HEMI HEAD.
Patient problems: Pain; Unspecified Infection
MAUDE report 5963635 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051075. One submission often covers many models, so check the product and lot details in each record.
- Z-1460-2026 · Jan 15, 2026 · Open, ClassifiedBrand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP STANDARD HUMERAL SOCKET INSERT, 36MM, HXe-plus Model/Catalog Number: 509-00-036
- Z-0625-2024 · Oct 5, 2023 · CompletedReverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797
- Z-0282-2024 · Sep 25, 2023 · Open, ClassifiedRSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.
- Z-2606-2023 · Aug 4, 2023 · Open, ClassifiedDJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder Syst
- Z-2739-2011 · Apr 18, 2011 · TerminatedReverse Shoulder Prosthesis System RSP Humeral Socket Shell DJO 9800 Metric Blvd. Austin, TX 78758 STERILE R Model numbers: 508-00-000 and 508-00-008. The Reverse Shoulder Prost
Other shoulder prosthesis, reverse configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page