Encore Medical, L.P. · SCREW, RETAINING GLENOID HEAD RSP · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888912024617), received between Sep 20, 2016 and Sep 20, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Break” (1 report).

5 other recalls cover devices cleared under the same FDA submission (K051075); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • No Apparent Adverse Event2
  • Break1
  • Detachment Of Device Component1

Patient problems

  • Foreign Body In Patient2
  • No Known Impact Or Consequence To Patient1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 20, 2024Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event

    REVISION SURGERY - DUE TO WHEN INSERTING THE SET SCREW INTO THE GLENOSPHERE WITH THE RED HANDLE TORQUE SCREWDRIVER THE SCREW BROKE OFF HALFWAY. BROKEN/FRAGMENTED PIECES LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20274343 ↗
  • InjuryJan 17, 2024No Apparent Adverse Event; No Apparent Adverse Event

    INSTRUMENT FAILURE - PIECES LEFT IN PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 18531774 ↗
  • InjuryMay 16, 2017Break; Detachment Of Device Component

    PRIMARY SURGERY - WHILE THE SURGEON WAS TURNING THE SCREW, IT OVER TORQUED AND THE HEAD OF THE SCREW SNAPPED OFF. THE SCREW IS NOT IN THE PATIENT BUT THE REMAINING SHAFT OF THE SCREW STAYED IN THE BASEPLATE/GLENOSPHERE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6569407 ↗
  • InjurySep 20, 2016Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO THE PATIENT PRESENTING AFTER A REVISION IN PAIN AND INFECTED. THE SURGEON DECIDED TO GO IN AND REMOVE ALL THE GLENOID COMPONENTS, THE 44 POLY AS WELL AS THE +8 SPACER. HE THEN DECIDED TO PUT IN A HEMI SPACER AND A HEMI HEAD.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 5963635 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051075. One submission often covers many models, so check the product and lot details in each record.

  • Z-1460-2026 · Jan 15, 2026 · Open, Classified
    Brand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP STANDARD HUMERAL SOCKET INSERT, 36MM, HXe-plus Model/Catalog Number: 509-00-036
  • Z-0625-2024 · Oct 5, 2023 · Completed
    Reverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797
  • Z-0282-2024 · Sep 25, 2023 · Open, Classified
    RSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.
  • Z-2606-2023 · Aug 4, 2023 · Open, Classified
    DJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder Syst
  • Z-2739-2011 · Apr 18, 2011 · Terminated
    Reverse Shoulder Prosthesis System RSP Humeral Socket Shell DJO 9800 Metric Blvd. Austin, TX 78758 STERILE R Model numbers: 508-00-000 and 508-00-008. The Reverse Shoulder Prost

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates