Skip to content

DJO SURGICAL 508-00-008Recalls & reported problems

Encore Medical, L.P. · SOCKET, SHELL +8MM RSP HUMERAL · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888912024631), received between Jun 3, 2016 and Feb 17, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K041066); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Component Falling1
  • Detachment Of Device Component1
  • Device Slipped1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Unspecified Infection2
  • Joint Laxity1
  • No Code Available1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 17, 2026Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 24373528 ↗
  • InjuryJan 10, 2018Insufficient Device Problem Information

    SECOND REVISION SURGERY - DUE TO THE PATIENT HAVING AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 7180527 ↗
  • InjurySep 29, 2016Detachment Of Device Component; Component Falling; Device Slipped

    REVISION SURGERY - DUE TO A +8 REVERSE SHOULDER PROSTHESIS (RSP) SHELL COMING OFF THE TAPER OF THE DJO 6 LONG STEM; THE LOCKING SCREW WAS LOOSE IN HEAD OF A 36 MM COMPONENT.

    Patient problems: No Code Available

    MAUDE report 5988793 ↗
  • InjuryJun 3, 2016Insufficient Device Problem Information

    SECOND REVISION SURGERY - DUE TO AN INFECTION, THE COMPONENTS WERE EXCHANGED.

    Patient problems: Unspecified Infection

    MAUDE report 5699399 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041066. One submission often covers many models, so check the product and lot details in each record.

  • Z-0098-2024 · Aug 22, 2023 · Open, Classified
    REVERSE PROSTHESIS SHOULDER SYSTEM, SCREW, LOCKING BONE, RSP, 5mmX30mm long, REF: 506-03-130
  • Z-0184-2008 · Aug 8, 2007 · Terminated
    Encore Reverse Shoulder Prosthesis (RSP) System, Standard Size 36 RSP Humeral Socket Insert; Catalog #: 508-00-036; Sterile; Manufactured by Encore, Medical, L.P., Austin, TX 7875
  • Z-0185-2008 · Aug 8, 2007 · Terminated
    Encore Reverse Shoulder Prosthesis (RSP) System, Standard Size 40 RSP Humeral Socket Insert; Catalog #: 508-00-040; Sterile; Manufactured by Encore, Medical, L.P., Austin, TX 7875
  • Z-1566-05 · Jul 28, 2005 · Terminated
    Encore Reverse Shoulder Prosthesis (RSP), manufactured and distributed by Encore Medical, L.P. 9800 Metric Blvd., Austin, Texas 78758

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page