Skip to content

DJO SURGICAL 508-01-032Recalls & reported problems

Encore Medical, L.P. · SOCKET, INSERT 32MM SEMI RSP HUMERAL · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a DJO Surgical reverse-configuration shoulder prosthesis socket insert used in shoulder replacement surgery. The 39 reports mostly describe instability of the implant, dislocation, loosening and infection, with several patients needing revision surgery; no deaths were reported, and all 39 involved injuries. One open recall from August 2023 states the components were not sterilized before shipment, which can lead to infection and possible revision surgery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
40
Recalls naming it
1

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00888912024693), received between Dec 23, 2015 and Sep 17, 2026 — classified as 40 injuries.

The most frequently reported specific device problem is “Unstable” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K051075); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 4, 2023Class IIOpen, Classified

    Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery

    Z-2606-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem20
  • Appropriate Device Problem Term/Code Not Available7
  • Unstable7
  • Device Dislodged or Dislocated5
  • Insufficient Device Problem Information2
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Noise, Audible1
  • Patient-Device Incompatibility1

Patient problems

  • Unspecified Infection21
  • Joint Laxity8
  • Joint Dislocation3
  • No Code Available3
  • Fall2
  • Implant Pain2
  • Pain2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Revision surgery - Due to infection.

    Patient problems: Unspecified Infection

    MAUDE report 50812636 ↗
  • InjuryJun 10, 2026Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY DUE TO LAXITY.

    Patient problems: Joint Laxity

    MAUDE report 25453225 ↗
  • InjuryMay 12, 2026Unstable

    REVISION SURGERY DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 25156547 ↗
  • InjuryDec 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD A SUPERFICIAL INFECTION 5 MONTHS AFTER SURGERY. THE SURGEON WANTED TO ENSURE THAT IT DID NOT PROPAGATE TO THE JOINT AND WASHED IT OUT AND REPLACED THE POLY.

    Patient problems: Unspecified Infection

    MAUDE report 23743200 ↗
  • InjurySep 14, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY

    Patient problems: Joint Laxity

    MAUDE report 20219332 ↗
  • InjuryMay 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 19347914 ↗
  • InjuryMar 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 18874142 ↗
  • InjuryNov 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 18088702 ↗
  • InjuryOct 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSENING AND INFECTION.

    Patient problems: Joint Laxity; Unspecified Infection

    MAUDE report 17981435 ↗
  • InjuryAug 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION AND PAIN.

    Patient problems: Implant Pain; Unspecified Infection

    MAUDE report 17495073 ↗
  • InjuryJul 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 17283446 ↗
  • InjuryJun 20, 2023Unstable; Unstable

    REVISION SURGERY - DUE TO INSTABILITY AND INFECTION.

    Patient problems: Joint Laxity; Unspecified Infection

    MAUDE report 17167579 ↗
  • InjuryJun 15, 2023Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 17140229 ↗
  • InjuryMay 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 16900125 ↗
  • InjuryApr 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 16812032 ↗
  • InjuryMar 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - POLY WEAR.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 16491990 ↗
  • InjuryFeb 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 16440506 ↗
  • InjuryOct 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO AN INFECTION.

    Patient problems: Medical device site infection

    MAUDE report 15544641 ↗
  • InjuryMar 9, 2022Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY: DUE TO PATIENT WAS INFECTED.

    Patient problems: Unspecified Infection

    MAUDE report 13713717 ↗
  • InjuryDec 14, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 13002222 ↗
  • InjuryNov 29, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY : DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 12882833 ↗
  • InjurySep 22, 2021Unstable; Appropriate Device Problem Term/Code Not Available; Noise, Audible

    REVISION SURGERY - PATIENT WAS UNSTABLE & HAD A "CLICKING SOUND". NO PRODUCT FAILURE.

    Patient problems: Failure of Implant

    MAUDE report 12514939 ↗
  • InjuryJan 13, 2021Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 11163279 ↗
  • InjurySep 16, 2020Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO THE PATIENT PRESENTING AT EMERGENCY ROOM WITH A DISLOCATED SHOULDER AFTER A FALL.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 10542701 ↗
  • InjurySep 16, 2020Adverse Event Without Identified Device or Use Problem

    SECOND REVISION SURGERY - DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 10541063 ↗

FDA clearance and classification

Cleared through Special 510(k) K051075 on May 27, 2005 (decision: Substantially Equivalent). Submitted by Encore Medical L.P. as “Encore Reverse Shoulder Prosthesis”. The FDA review took 30 days. FDA record ↗

The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.

Recalls under the same FDA submission

These recalls involve devices cleared under K051075. One submission often covers many models, so check the product and lot details in each record.

  • Z-1460-2026 · Jan 15, 2026 · Open, Classified
    Brand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP STANDARD HUMERAL SOCKET INSERT, 36MM, HXe-plus Model/Catalog Number: 509-00-036
  • Z-0625-2024 · Oct 5, 2023 · Completed
    Reverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797
  • Z-0282-2024 · Sep 25, 2023 · Open, Classified
    RSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.
  • Z-2739-2011 · Apr 18, 2011 · Terminated
    Reverse Shoulder Prosthesis System RSP Humeral Socket Shell DJO 9800 Metric Blvd. Austin, TX 78758 STERILE R Model numbers: 508-00-000 and 508-00-008. The Reverse Shoulder Prost

Common questions about the DJO SURGICAL 508-01-032

Has the DJO SURGICAL 508-01-032 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2606-2023 (Class II), began on Aug 4, 2023: Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the DJO SURGICAL 508-01-032?

The FDA's MAUDE database lists 40 adverse event reports for this device, received between Dec 23, 2015 and Sep 17, 2026. In those reports, the most common specific device problems are “Unstable” (7), “Device Dislodged or Dislocated” (5) and “Loosening of Implant Not Related to Bone-Ingrowth” (1); the most common patient problems are “Unspecified Infection” (21), “Joint Laxity” (8) and “Joint Dislocation” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DJO SURGICAL 508-01-032?

None of the reports are classified as deaths, 40 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DJO SURGICAL 508-01-032 FDA approved?

It was cleared through Special 510(k) K051075 on May 27, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the DJO SURGICAL 508-01-032 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the DJO SURGICAL 508-01-032, and is it still sold?

The FDA's device database lists Encore Medical, L.P. as the company responsible for it (brand DJO SURGICAL, model 508-01-032). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page