Skip to content

DJO SURGICAL 425-97-010Recalls & reported problems

Encore Medical, L.P. · TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 10 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DJO Surgical TaperFill hip stem is an implant used in hip replacement surgery. The 36 reports mostly describe a stem that loosens in the bone, or a break in the stem or the bone around it, which in several cases led to a second operation. Reports also mention falls and joint looseness. There are no deaths, and 36 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
37
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00888912078689), received between Feb 14, 2017 and Sep 25, 2026 — classified as 37 injuries.

The most frequently reported specific device problem is “Fracture” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem17
  • Fracture7
  • Loosening of Implant Not Related to Bone-Ingrowth7
  • Appropriate Device Problem Term/Code Not Available5
  • Device Contamination with Chemical or Other Material1
  • Device Dislodged or Dislocated1
  • Loose or Intermittent Connection1
  • Unintended Movement1

Patient problems

  • Joint Laxity9
  • Bone Fracture(s)8
  • Fall6
  • No Code Available4
  • Insufficient Information2
  • Limb Fracture2
  • No Clinical Signs, Symptoms or Conditions2
  • Unspecified Infection2

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Fracture

    Patient problems: Periprosthetic fracture

    MAUDE report 50892498 ↗
  • InjuryJul 8, 2026Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 25755537 ↗
  • InjuryApr 22, 2026Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY DUE TO STEM WAS LOOSE.

    Patient problems: Joint Laxity

    MAUDE report 24961032 ↗
  • InjuryMar 20, 2026Fracture

    REVISION SURGERY - DUE TO PERIPROSTHETIC FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 24656426 ↗
  • InjuryMar 11, 2026Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 24574080 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    THE STEM WAS LOOSE. M/71.

    Patient problems: Joint Laxity

    MAUDE report 24408304 ↗
  • InjuryAug 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PERIPROSTHETIC FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 22789915 ↗
  • InjuryMay 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PERIPROSTHETIC FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 22081448 ↗
  • InjuryOct 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 20563704 ↗
  • InjuryOct 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO SUBSIDENCE.

    Patient problems: Joint Laxity

    MAUDE report 20477954 ↗
  • InjurySep 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO SUBSIDENCE.

    Patient problems: Joint Laxity

    MAUDE report 20297405 ↗
  • InjuryApr 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO FALL LOWER BODY COMPONENT AND FRACTURE.

    Patient problems: Limb Fracture

    MAUDE report 19218428 ↗
  • InjuryMar 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO FALL LOWER BODY COMPONENT AND FRACTURE.

    Patient problems: Limb Fracture

    MAUDE report 18871691 ↗
  • InjuryDec 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO FRACTURE

    Patient problems: Bone Fracture(s)

    MAUDE report 18361419 ↗
  • InjuryOct 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 17874006 ↗
  • InjurySep 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 17786471 ↗
  • InjuryJun 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 17243843 ↗
  • InjuryMay 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 16865761 ↗
  • InjuryNov 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - THE STEM SUBSIDED AND THEY CHANGED IT OUT ALONG WITH THE LINER AND HEAD.

    Patient problems: Joint Laxity

    MAUDE report 15793927 ↗
  • InjuryOct 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO STEM MOVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15670290 ↗
  • InjuryOct 13, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PRIMARY TOTAL HIP REPLACEMENT WAS PERFORMED. SURGERY PROCEEDED ROUTINELY UNTIL FEMORAL STEM WAS IMPLANTED AND SEATED. WITH LAST IMPACT OF THE STEM A FRACTURE OCCURRED AT THE TOP OF THE STEM AND THE FRACTURE SPIRALED PROXIMALLY UP THE FEMUR. LONG STEM FEMORAL IMPLANTS WERE NOT AVAILABLE AND IT WAS DECIDED TO CABLE THE FRACTURE, PLACE A FEMORAL HEAD ON THE STEM AND COME BACK IN 3 DAY TO REVISE THE FEMORAL COMPONENT TO AN EXPERT STEM AND HEAD.

    Patient problems: Bone Fracture(s)

    MAUDE report 15596944 ↗
  • InjuryAug 4, 2022Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - PATIENT FELL AND CAUSED STEM TO LOOSE. REPLACED THE STEM WITH EXPERT HIP AND DUAL MOBILITY LINEAR.

    Patient problems: Fall

    MAUDE report 15165407 ↗
  • InjuryNov 22, 2021Device Contamination with Chemical or Other Material

    REPORTED INCIDENT - WHEN THE SURGEON PULLED THE STEM OUT OF THE STERILE PACKAGE, A VERY SMALL BLUE, UNIDENTIFIED OBJECT WAS EMBEDDED IN THE POROUS COATING OF THE STEM. AFTER SEVERAL TRIES, HE WAS FINALLY ABLE TO REMOVE THE OBJECT WITH A SMALL NEEDLE NOSED KOCHER. HE INSTRUCTED THE SCRUB TECH TO REMOVE THE OBJECT OFF OF THE FIELD IMMEDIATELY. SCRUB TECH WAS ABLE TO RECOVER THE OBJECT AND HAVE IT SAVED IN A SPECIMEN CONTAINER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12853473 ↗
  • InjuryOct 26, 2021Fracture; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO PERI-PROSTHETIC FRACTURE.

    Patient problems: Insufficient Information

    MAUDE report 12699295 ↗
  • InjuryAug 6, 2021Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - DUE TO STEM LOOSENING, HIP PAIN.

    Patient problems: Pain

    MAUDE report 12287733 ↗

Other Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer devices

All Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page