DJO SURGICAL 425-99-012Recalls & reported problems
Encore Medical, L.P. · Linear, Lateral, Reduced Neck, Sz 12, W/P2 Coating · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888912078986), received between Jul 20, 2018 and Mar 5, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Fracture” (1 report).
1 other recall cover devices cleared under the same FDA submission (K120241); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Appropriate Device Problem Term/Code Not Available1
- Fracture1
Patient problems
- Bone Fracture(s)1
- Joint Dislocation1
- Joint Laxity1
- Limb Fracture1
- No Code Available1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjurySep 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryOct 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION SURGERY - PATIENT HAD A PRIMARY TOTAL HIP. PRESENTED AT SURGEON OFFICE WITH A LEG LENGTH DISCREPANCY. AN XRAY REVEALED THAT THE LINEAR STEM HAD SUBSIDED 1CM. THE SURGEON SCHEDULED A REVISION TOTAL HIP FOR DATE. DURING THE REVISION, HE REMOVED THE SIZE 12 LATERAL OFFSET LINEAR STEM, 36MM FEMORAL HEAD AND -3.5MM OFFSET SLEEVE. HE BROACHED UP TO A SIZE 15 LINEAR WHICH WAS MORE STABLE IN THE FEMORAL CANAL. HE IMPLANTED A SIZE 15 LINEAR STEM, 36MM FEMORAL HEAD AND A NEUTRAL OFFSET SLEEVE. THERE WERE NO ISSUES DURING THE REVISION.
Patient problems: Joint Laxity
MAUDE report 15692415 ↗InjuryMay 11, 2022Appropriate Device Problem Term/Code Not Available
REVISION SURGERY - THE PATIENT HAS TERRIBLE BONE. THE SURGEON FRACTURED THE FEMUR AS HE WAS INSERTING THE STEM. BECAUSE THE FEMUR WAS FRACTURED DURING THE PRIMARY SURGERY AND IMPLANTED COMPONENTS HAD TO BE REMOVED, IT TURNED INTO A REVISION SURGERY.
Patient problems: Bone Fracture(s)
MAUDE report 14364632 ↗InjuryJul 20, 2018Fracture
REVISION SURGERY - DUE TO THE PATIENT HAVING A PERI PROSTHETIC FRACTURE.
Patient problems: No Code Available
MAUDE report 7707624 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K120241. One submission often covers many models, so check the product and lot details in each record.
- Z-1692-2020 · Jul 10, 2018 · Open, Classifieddjosurgical Linear Neck Trials with Spring Sides, Standard, nonsterile, REF S-200635; and djosurgical Linear Neck Trials with Spring Sides, Lateral, nonsterile, REF S-200636. The
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page