Encore Medical, L.P. · FMP Hemispherical Shells w/Screw Holes, 50mm, w/P2 Coating · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
21
Deaths reported
0
Injuries reported
21
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00888912084017), received between Apr 5, 2016 and Feb 5, 2026 — classified as 21 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem9
  • Device Dislodged or Dislocated4
  • Appropriate Device Problem Term/Code Not Available3
  • Unstable3
  • Insufficient Device Problem Information2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Malposition of Device2
  • Loose or Intermittent Connection1

Patient problems

  • Implant Pain4
  • No Code Available4
  • Failure of Implant2
  • Insufficient Information2
  • Joint Dislocation2
  • Joint Laxity2
  • Pain2
  • Fall1

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 5, 2026Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 24274725 ↗
  • InjuryApr 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 21940280 ↗
  • InjuryMar 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATIENT WAS EXPERIENCING PAIN. SURGEON SUSPECTED INFECTION.

    Patient problems: Implant Pain; Unspecified Infection

    MAUDE report 21735600 ↗
  • InjuryJan 16, 2025Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 21175485 ↗
  • InjuryAug 22, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 20034888 ↗
  • InjuryNov 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 18188594 ↗
  • InjuryOct 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 18000402 ↗
  • InjuryApr 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 16710683 ↗
  • InjurySep 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - REPRESENTATIVE WAS CONTACTED FOR DJO FEMORAL HEADS. NO INFORMATION WAS PROVIDED REGARDING SURGICAL PLAN, OTHER THAN EXISTING CUP WILL BE REMOVED AND REPLACED WITH A CUP FROM A DIFFERENT VENDOR. REPRESENTATIVE WAS NOT PRESENT DURING REVISION SURGERY, AND RECEIVED INFORMATION AFTER SURGERY INCLUDING PATIENT STICKER, AND IMPLANT STICKERS (DELTA OPTION CERAMIC HEAD AND TI SLEEVE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15341450 ↗
  • InjuryJul 13, 2022Malposition of Device

    REVISION SURGERY - CUP IMPROPERLY PLACED DUE TO MALFORMATION OF THE PELVIS. CUP WAS PLACED TOO HIGH BY PREVIOUS SURGEON.

    Patient problems: Insufficient Information

    MAUDE report 15001895 ↗
  • InjuryFeb 23, 2022Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - CUP BECAME LOOSE AND SPUN OUT, BREAKING THE SCREWS. THE HEAD ENDED UP RUBBING A HOLE IN THE LINER AND CUP.

    Patient problems: Failure of Implant

    MAUDE report 13584450 ↗
  • InjuryOct 12, 2021Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Insufficient Information

    MAUDE report 12618984 ↗
  • InjuryNov 8, 2019Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - THE SHELL CAME LOOSE.

    Patient problems: No Code Available

    MAUDE report 9295963 ↗
  • InjuryApr 4, 2019Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO A DISLOCATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8483871 ↗
  • InjuryJan 14, 2019Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - THE PATIENT HAD IMPINGEMENT SUPERIORITY AND PAIN. THE X-RAYS LOOKED FINE, ALL COMPONENTS WERE WELL SEATED AND WERE NOT LOOSE AT THE TIME OF THE REVISION. THE SURGEON CHOSE TO EXPLANT THE ACETABULAR CUP, POLY, AND SCREW, AND REVISE TO A BIOMET G7 DUAL MOBILITY.

    Patient problems: Pain

    MAUDE report 8245631 ↗
  • InjuryNov 1, 2018Unstable

    REVISION SURGERY - DUE TO THE CUP GOING THROUGH THE MEDIAL WALL POST OPERATIVELY AND BECOMING UNSTABLE. THE SURGEON REMOVED AND REPLACED THE CUP AND LINER WITH COMPETITOR PRODUCTS. THE SURGEON ALSO REMOVED AND REPLACED THE HEAD AND ADDED A SLEEVE WITH A DJO PRODUCT. THE ORIGINAL STEM REMAINED IN THE PATIENT.

    Patient problems: No Code Available

    MAUDE report 8030517 ↗
  • InjuryMar 6, 2018Appropriate Device Problem Term/Code Not Available

    SECOND REVISION SURGERY - DUE TO THE PATIENT HAVING INSTABILITY. THE SURGEON PUT IN A DIFFERENT VENDOR SHELL AND LINER.

    Patient problems: Failure of Implant

    MAUDE report 7317741 ↗
  • InjuryJun 16, 2017Loose or Intermittent Connection; Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO THE PATIENT HAVING A LOOSE CUP; THE HIP WAS DISLOCATING. A DJO 50MM CUP WITH 1 SCREW WAS REMOVED, A 56MM DJO P2 CUP WAS IMPLANTED WITH 3 SCREWS, 40 MM LINER AND HEAD WERE USED FOR STABILITY. NOT A PRODUCT PROBLEM.

    Patient problems: No Code Available

    MAUDE report 6648719 ↗
  • InjuryFeb 21, 2017Insufficient Device Problem Information

    REVISION SURGERY - THE COMPONENTS HAD TO BE REVISED BECAUSE THE PATIENT COMPLAINED OF FEMORAL NERVE PAIN.

    Patient problems: Pain

    MAUDE report 6346798 ↗
  • InjuryOct 4, 2016Insufficient Device Problem Information

    REVISION SURGERY - DUE TO THE PATIENT FALLING AND PUSHING THE SHELL THROUGH THE ACETABULUM.

    Patient problems: Fall

    MAUDE report 5998012 ↗
  • InjuryApr 5, 2016Malposition of Device

    REVISION SURGERY: DUE TO THE FMP CUP MOVING FROM THE ORIGINAL POSITION; THE SURGEON NEEDED TO REPLACE THE IMPLANT AND REPOSITION IT IN A BETTER LOCATION.

    Patient problems: No Code Available

    MAUDE report 5547520 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →