DJO SURGICAL 430-98-052Recalls & reported problems

Encore Medical, L.P. · FMP Hemispherical Shells w/Screw Holes, 52mm, w/P2 Coating · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DJO Surgical hip shell is an uncemented cup used in hip replacement. Reports mostly describe dislocations and pain with repeat surgery, with some mentioning loose cups or screws not locking in place. There are 23 reports, with no deaths and 23 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
23
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00888912084024), received between Jan 20, 2017 and Aug 6, 2025 — classified as 23 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Appropriate Device Problem Term/Code Not Available6
  • Device Dislodged or Dislocated4
  • Loose or Intermittent Connection2
  • Malposition of Device2
  • Fracture1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Insufficient Device Problem Information1

Patient problems

  • Joint Dislocation5
  • No Code Available4
  • No Known Impact Or Consequence To Patient4
  • Insufficient Information3
  • Pain3
  • Fall2
  • Discomfort1
  • Foreign Body In Patient1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 22734501 ↗
  • InjuryJan 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 21270254 ↗
  • InjuryAug 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 20020531 ↗
  • InjuryNov 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT PRESENTED WITH PAIN DUE TO PSOAS TENDON IMPINGEMENT.

    Patient problems: Pain

    MAUDE report 18204850 ↗
  • InjuryOct 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - MDR - DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 17936088 ↗
  • InjuryOct 26, 2022No Apparent Adverse Event; No Apparent Adverse Event

    PRIMARY SURGERY - AFTER DRILLING THROUGH CUP HOLE TO PLACE SCREW, THE 5.0 SCREW DIDN'T LOCK INTO CUP AND WENT PAST CUP.

    Patient problems: Foreign Body In Patient

    MAUDE report 15674896 ↗
  • InjuryMay 25, 2022Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - SURGEON REVISED CUP AND LINER ON THE CASE. PATIENT COMPLAINED OF PAIN.

    Patient problems: Pain

    MAUDE report 14498094 ↗
  • InjuryMay 16, 2022Loose or Intermittent Connection

    REVISION SURGERY - LOOSE CUP FLEXION, ONE SCREW. CUP PIVOT ON SCREW TO VERTICAL POSITION WITH FIBROUS IN GROWTH.

    Patient problems: Insufficient Information

    MAUDE report 14408430 ↗
  • InjuryMar 4, 2022Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY : DUE TO CUP WAS TOO ANTEVERTED, HIP WAS DISLOCATING. REMOVED AND EXCHANGED OUT THE CUP, LINER, HEAD, AND SCREWS.

    Patient problems: Insufficient Information

    MAUDE report 13667772 ↗
  • InjuryDec 2, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - PATIENT'S CUP HAD MOVED AFTER A FALL AND NEEDED TO BE REPOSITIONED.

    Patient problems: Fall

    MAUDE report 12917770 ↗
  • InjurySep 9, 2021Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT FELL AND HAD A DISLOCATION, SURGEON REPOSITIONED CUP WITH A DIFFERENT PRODUCT.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 12448870 ↗
  • InjuryJul 21, 2021Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT DISCOMFORT.

    Patient problems: Discomfort

    MAUDE report 12204374 ↗
  • InjuryJul 14, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - CONSTRAINED LINER FAILED DUE TO RETROVERTED CUP. SURGEON PULLED CUP, AND A STYKER CUP AND LINER WERE IMPLANTED. A NEW DJO HEAD WAS USED AS WELL.

    Patient problems: Insufficient Information

    MAUDE report 12170146 ↗
  • InjuryJun 23, 2020Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - ACETABULAR CUP BECAME LOOSE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10184967 ↗
  • InjuryMar 27, 2019Migration or Expulsion of Device; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO THE CUP MIGRATING SUPERIORLY. SPECIMEN WAS SENT TO PATHOLOGY CAGE WAS NEEDED FOR REVISION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8458225 ↗
  • InjuryMar 15, 2019Fracture

    REVISION SURGERY - DUE TO AN ACETABULAR FRACTURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8424784 ↗
  • InjuryJan 15, 2019Unstable

    REVISION SURGERY - REVISED FOR INSTABILITY. THE SURGEON BELIEVED THE CUP COMPONENT SPUN AFTER THE INITIAL SURGERY.

    Patient problems: No Code Available

    MAUDE report 8250307 ↗
  • InjuryDec 12, 2018Malposition of Device; Material Protrusion/Extrusion

    REVISION SURGERY - DUE TO THE CUP HAVING SHIFTED, THE SURGEON NEEDED TO REPLACE CAGE DUE TO PROTRUSIO (PROTRUSION).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8158511 ↗
  • InjuryJul 27, 2018Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO THE PATIENT HAVING DISLOCATED POST-OPERATIVE.

    Patient problems: No Code Available

    MAUDE report 7728775 ↗
  • InjuryJun 13, 2018Incorrect, Inadequate or Imprecise Result or Readings; Insufficient Device Problem Information

    REVISION SURGERY - DUE TO THE PATIENT COMPLAINING OF PAIN.

    Patient problems: Pain

    MAUDE report 7597415 ↗
  • InjuryMar 22, 2018Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO A DISLOCATION. THE SURGEON ONLY REPLACED A DJO HEAD AND THE OTHER IMPLANTS WERE REPLACED WITH A COMPETITOR PART.

    Patient problems: Joint Dislocation

    MAUDE report 7362724 ↗
  • InjuryFeb 23, 2018Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO SUPERIOR DISLOCATION, NOT CAUSED BY IMPLANTS. REVISION CUP PLACED SLIGHTLY MORE ANTERIOR.

    Patient problems: No Code Available

    MAUDE report 7293155 ↗
  • InjuryJan 20, 2017Malposition of Device

    REVISION SURGERY - DUE TO THE CUP PUSHING ON THE PATIENT'S BLADDER.

    Patient problems: No Code Available

    MAUDE report 6266685 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →