Skip to content

DJO SURGICAL 130-03-738Recalls & reported problems

Encore Medical, L.P. · DOMED TRI-PEG PATELLA, 38x9mm, E-PLUS · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
21
Deaths reported
0
Injuries reported
21
Recalls naming it
1

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00888912108560), received between Nov 1, 2018 and Jul 28, 2025 — classified as 21 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 5, 2023Class IICompleted

    Mislabeling

    Z-0626-2024 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem12
  • Appropriate Device Problem Term/Code Not Available4
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Break1
  • Loose or Intermittent Connection1
  • Therapeutic or Diagnostic Output Failure1
  • Unstable1

Patient problems

  • Joint Laxity6
  • Failure of Implant4
  • Unspecified Infection4
  • Fall2
  • Implant Pain2
  • Insufficient Information2
  • Pain2
  • Discomfort1

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 22638411 ↗
  • InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO BAD PATELLA PAIN.

    Patient problems: Implant Pain

    MAUDE report 22128611 ↗
  • InjuryFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATELLA FRACTURE.

    Patient problems: Failure of Implant

    MAUDE report 21360807 ↗
  • InjuryOct 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO IMPLANT BROKE, PEGS ON PATELLA BROKE AND WERE FLOATING.

    Patient problems: Failure of Implant

    MAUDE report 20518619 ↗
  • InjuryMar 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 18852883 ↗
  • InjuryJan 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - THE PATELLA SHOWED UP ON THE BONE SCAN. 74 YR OLD MALE.

    Patient problems: Insufficient Information

    MAUDE report 18522864 ↗
  • InjurySep 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO FRACTURE AND LOOSENING.

    Patient problems: Joint Laxity; Limb Fracture

    MAUDE report 17687363 ↗
  • InjuryAug 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE INFECTION AND FRACTURE.

    Patient problems: Muscle/Tendon Damage; Unspecified Infection

    MAUDE report 17592672 ↗
  • InjuryDec 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSE PATELLA.

    Patient problems: Joint Laxity

    MAUDE report 15959176 ↗
  • InjuryNov 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT COMPLAINED ABOUT PAIN, SO HIS PATELLA WAS REMOVED, AND THIS IMPLANT WAS ADDED.

    Patient problems: Pain

    MAUDE report 15801364 ↗
  • InjurySep 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT PRESENTED TO SURGEON WITH RIGHT KNEE PAIN. ON X-RAY IT APPEARED THAT THE PATELLA IMPLANT WAS LOOSE AND HAD DISASSOCIATED FROM THE BONE. IN THE REVISION SURGERY, IT WAS CONFIRMED THAT THE PATELLA HAD COME LOOSE, AND APPEARED THAT THE PEGS ON THE PATELLA IMPLANT HAD SHEARED OFF. PRIOR BONE CEMENT WAS ON THE IMPLANT BUT DID NOT APPEAR ON THE NATIVE PATELLA. SURGEON SPECULATED THAT POSSIBLY IN THE PRIMARY SURGERY BONE CEMENT HAD SETUP QUICKLY AND PERHAPS DID NOT GET STRONG ADHERENCE TO THE BONE. PRE-REVISION X-RAYS AND INTRA-OPERATIVE PICTURES ATTACHED.

    Patient problems: Implant Pain; Joint Laxity

    MAUDE report 15341572 ↗
  • InjuryAug 3, 2022Therapeutic or Diagnostic Output Failure

    REVISION SURGERY - PATIENT NEEDED PATELLA.

    Patient problems: Failure of Implant

    MAUDE report 15157184 ↗
  • InjuryMar 24, 2022Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - LOOSE KNEE AND PAIN, PATELLA WAS LOOSE. NO KNOWN TRAUMA.

    Patient problems: Joint Laxity; Pain

    MAUDE report 13876460 ↗
  • InjuryNov 1, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO CHRONICALLY INFECTED KNEE. PRIMARY TOTAL KNEE ARTHROPLASTY IMPLANTS WERE REMOVED AND REPLACED WITH AN OSTEOREMEDIES ANTIBIOTIC SPACER WAS USED IN THE FIRST STAGE OF THE REVISION. THE REPRESENTATIVE WAS NOT NOTIFIED OR PRESENT FOR THE REMOVAL OF THESE IMPLANTS. LATER THE REPRESENTATIVE WAS PRESENT FOR THE 2ND STAGE OF THE REVISION (REMOVAL OF ANTIBIOTIC SPACER AND IMPLANTATION OF REVISION KNEE SYSTEM), USING EXPRT KNEE AND FOUND OUT THAT THIS PATIENT PREVIOUSLY HAD A KNEE THAT WAS REMOVED.

    Patient problems: Unspecified Infection

    MAUDE report 12733915 ↗
  • InjuryOct 20, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 12666826 ↗
  • InjuryMay 17, 2021Appropriate Device Problem Term/Code Not Available; Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - THE PATIENT HAD A LOOSE PATELLA BUTTON.

    Patient problems: Insufficient Information

    MAUDE report 11832659 ↗
  • InjuryMay 2, 2021Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY - DUE TO PATIENT WAS INFECTED. ALL COMPONENTS WERE REMOVED.

    Patient problems: Unspecified Infection

    MAUDE report 11758779 ↗
  • InjuryMay 26, 2020Break

    REVISION SURGERY - THE PATIENT HAD AN AGGRESSIVE FALL AND IT SHEERED THE PATELLA.

    Patient problems: Failure of Implant; Fall

    MAUDE report 10087673 ↗
  • InjuryDec 5, 2018Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO THE PATIENT HAVING DISCOMFORT.

    Patient problems: Discomfort

    MAUDE report 8134877 ↗
  • InjuryDec 4, 2018Unstable

    REVISION SURGERY - DUE TO INSTABILITY AND QUADRICEPS REPAIR. THE SURGEON REMOVED AND EXCHANGED OUT THE INSERT. TOWARDS THE END, THE INSERT WAS NOT BIG ENOUGH, SO THE NEW INSERT WAS REMOVED ALONG WITH ALL ORIGINAL DJO IMPLANTS AND REPLACED WITH COMPETITOR BRAND IMPLANTS.

    Patient problems: No Code Available

    MAUDE report 8129205 ↗
  • InjuryNov 1, 2018Loose or Intermittent Connection

    REVISION SURGERY - DUE TO THE PATIENT FALLING AND HAVING A LOOSE PATELLA.

    Patient problems: Fall

    MAUDE report 8029280 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page