Skip to content

DJO SURGICAL 385-11-502Recalls & reported problems

Encore Medical, L.P. · INSERT, STD., PRIMARY, SIZE2 11MM CM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888912117821), received between Sep 13, 2017 and Sep 6, 2025 — classified as 7 injuries.

The most frequently reported specific device problem is “Unstable” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Unstable3
  • Loose or Intermittent Connection2
  • Naturally Worn1

Patient problems

  • Joint Laxity3
  • Pain3
  • No Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjurySep 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO WEAR.

    Patient problems: Joint Laxity

    MAUDE report 22987265 ↗
  • InjuryMar 28, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 18991889 ↗
  • InjuryJun 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO WORN POLY.

    Patient problems: Joint Laxity

    MAUDE report 17078099 ↗
  • InjuryNov 30, 2020Loose or Intermittent Connection

    REVISION SURGERY - PATIENT WAS PRESENTED WITH PAIN DUE TO LAXITY 23 YEARS POST OPERATION. SURGEON DETERMINED NEED FOR THICKER POLY.

    Patient problems: Pain

    MAUDE report 10920134 ↗
  • InjuryNov 20, 2020Unstable

    REVISION SURGERY - PATIENT PRESENTED 16 YEARS POST-OPERATION WITH PAIN DUE TO INSTABILITY. SURGEON DETERMINED THICKER INSERT WAS NEEDED.

    Patient problems: Pain

    MAUDE report 10875550 ↗
  • InjurySep 14, 2018Unstable

    REVISION SURGERY - DUE TO THE PATIENT DEVELOPING FLEXION INSTABILITY OF HER KNEE.

    Patient problems: No Information

    MAUDE report 7877662 ↗
  • InjurySep 13, 2017Loose or Intermittent Connection; Naturally Worn

    REVISION SURGERY - THE PATIENT PRESENTED WITH PAIN DUE TO LAXITY NINE YEARS POST OPERATION; THE SURGEON NOTICED MILD POSTERIOR POLY WEAR AND REPLACED THE INSERT.

    Patient problems: Pain

    MAUDE report 6863284 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates