Skip to content

DJO SURGICAL 509-00-040Recalls & reported problems

Encore Medical, L.P. · RSP STANDARD HUMERAL SOCKET INSERT, 40MM, HXe-plus · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DJO Surgical shoulder prosthesis component is used in reverse total shoulder replacement surgery. The 36 reports mostly describe infection at the surgical site and the shoulder joint dislocating or feeling unstable, with several patients needing revision surgery to remove or swap the implant;36 injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
37
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00888912144551), received between Jan 25, 2017 and Sep 14, 2026 — classified as 37 injuries.

The most frequently reported specific device problem is “Unstable” (5 reports).

5 other recalls cover devices cleared under the same FDA submission (K051075); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem23
  • Appropriate Device Problem Term/Code Not Available5
  • Unstable5
  • Insufficient Device Problem Information2
  • Device Dislodged or Dislocated1
  • Fracture1
  • Inadequacy of Device Shape and/or Size1
  • No Apparent Adverse Event1

Patient problems

  • Unspecified Infection16
  • Joint Dislocation10
  • Joint Laxity6
  • Medical device site infection3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bacterial Infection1
  • Bone Fracture(s)1
  • Fall1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Unstable

    Patient problems: Joint Laxity

    MAUDE report 50765490 ↗
  • InjuryJan 19, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24112140 ↗
  • InjuryDec 10, 2025Unstable; Unstable

    PATIENT UNSTABLE SHOULDER. REMOVED 40 IMPLANTED 55.

    Patient problems: Joint Laxity

    MAUDE report 23774436 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION

    Patient problems: Unspecified Infection

    MAUDE report 23700718 ↗
  • InjuryNov 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 23493121 ↗
  • InjuryOct 27, 2025Inadequacy of Device Shape and/or Size; Unstable; Inadequacy of Device Shape and/or Size; Unstable

    THE PATIENT PRESENTED APPROXIMATELY FOUR (4) YEARS POSTOPERATIVELY FROM A REVERSE TOTAL SHOULDER ARTHROPLASTY WITH COMPLAINTS OF INSTABILITY. THE SURGEON ELECTED TO REVISE THE IMPLANT TO A LARGER GLENOSPHERE. DURING THE PROCEDURE, THE EXISTING GLENOSPHERE AND INSERT WERE REMOVED AND REPLACED.

    Patient problems: Joint Dislocation

    MAUDE report 23399419 ↗
  • InjuryMay 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 22119870 ↗
  • InjuryMay 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY: DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 22049089 ↗
  • InjuryOct 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 20404158 ↗
  • InjuryJul 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 19757645 ↗
  • InjuryJul 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 19672436 ↗
  • InjuryMay 18, 2024No Apparent Adverse Event; No Apparent Adverse Event

    REVISION SURGERY - INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 19347910 ↗
  • InjuryJan 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 18500061 ↗
  • InjuryAug 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 17495069 ↗
  • InjuryJul 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 17401742 ↗
  • InjuryJul 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 17385919 ↗
  • InjuryMay 25, 2023Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 17007850 ↗
  • InjuryFeb 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISCOMFORT AND POSSIBLE INFECTION.

    Patient problems: Implant Pain; Unspecified Infection

    MAUDE report 16382988 ↗
  • InjuryDec 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - THE PATIENT'S JOINT WAS DISLOCATED AND UNSTABLE.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 16015390 ↗
  • InjuryDec 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - THE PATIENT DISLOCATED AND HAS A CHRONIC RISK OF DISLOCATION DUE TO A PRE-EXISTING CONDITION (CHARCOT).

    Patient problems: Joint Dislocation

    MAUDE report 15908432 ↗
  • InjuryNov 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT HAD PAIN.

    Patient problems: Pain

    MAUDE report 15821894 ↗
  • InjuryNov 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO AN INFECTION AND DISLOCATION.

    Patient problems: Joint Dislocation; Medical device site infection

    MAUDE report 15794422 ↗
  • InjuryOct 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - BASEPLATE WAS NOT DJO, BECAME LOOSE SURGEON DECIDED TO CHANGE IT.

    Patient problems: Joint Laxity

    MAUDE report 15693570 ↗
  • InjuryOct 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO AN INFECTION.

    Patient problems: Medical device site infection

    MAUDE report 15663133 ↗
  • InjurySep 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO AN INFECTION AND INSTABILITY.

    Patient problems: Joint Laxity; Medical device site infection

    MAUDE report 15365152 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051075. One submission often covers many models, so check the product and lot details in each record.

  • Z-1460-2026 · Jan 15, 2026 · Open, Classified
    Brand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP STANDARD HUMERAL SOCKET INSERT, 36MM, HXe-plus Model/Catalog Number: 509-00-036
  • Z-0625-2024 · Oct 5, 2023 · Completed
    Reverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797
  • Z-0282-2024 · Sep 25, 2023 · Open, Classified
    RSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.
  • Z-2606-2023 · Aug 4, 2023 · Open, Classified
    DJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder Syst
  • Z-2739-2011 · Apr 18, 2011 · Terminated
    Reverse Shoulder Prosthesis System RSP Humeral Socket Shell DJO 9800 Metric Blvd. Austin, TX 78758 STERILE R Model numbers: 508-00-000 and 508-00-008. The Reverse Shoulder Prost

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page