Skip to content

DJO SURGICAL 509-00-432Recalls & reported problems

Encore Medical, L.P. · RSP HUMERAL SOCKET INSERT, 32MM +4MM, STANDARD HXe-plus · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This DJO Surgical shoulder implant insert has 116 adverse event reports on file. Many describe infection, joint dislocation or instability, and several describe revision surgery, including one after the plastic insert was swapped. The reports include no deaths and 116 injuries. There is 1 open recall, which began in January 2026 over incorrect labels on some knee and shoulder socket implants.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
117
Deaths reported
0
Injuries reported
117
Recalls naming it
1

Summary

The FDA's MAUDE database contains 117 adverse event reports that list this exact device (DI 00888912144575), received between Jan 12, 2016 and Sep 17, 2026 — classified as 117 injuries.

The most frequently reported specific device problem is “Unstable” (19 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 4 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 6 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 4 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 15, 2026Class IIOpen, Classified

    Knee and Humeral socket implants contain incorrect labeling.

    Z-1458-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem54
  • Appropriate Device Problem Term/Code Not Available34
  • Unstable19
  • Device Dislodged or Dislocated18
  • Loose or Intermittent Connection4
  • Contamination1
  • Fracture1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Unspecified Infection43
  • Joint Dislocation22
  • Joint Laxity18
  • No Code Available14
  • Insufficient Information10
  • Failure of Implant5
  • No Known Impact Or Consequence To Patient5
  • Pain4

Recent adverse event reports

The 25 most recent of 117 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Device Dislodged or Dislocated

    Patient problems: Joint Dislocation

    MAUDE report 50812262 ↗
  • InjurySep 13, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50747925 ↗
  • InjurySep 10, 2026Unstable

    Patient problems: Joint Laxity

    MAUDE report 50724619 ↗
  • InjuryJul 6, 2026Patient Device Interaction Problem

    REVISION SURGERY DUE TO UNSTABLE AND INFECTED SHOULDER.

    Patient problems: Joint Laxity; Unspecified Infection

    MAUDE report 25727442 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY DUE TO INFECTED JOINT, WASHOUT AND IMPLANTS CHANGE. SINGLE STAGE REVISION.

    Patient problems: Unspecified Infection

    MAUDE report 25276477 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY DUE TO INFECTED. FULL REMOVAL, THE SURGEON PUT IN A SPACER.

    Patient problems: Unspecified Infection

    MAUDE report 25276472 ↗
  • InjuryMay 19, 2026Device Dislodged or Dislocated

    REVISION SURGERY DUE TO SHOULDER DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 25225294 ↗
  • InjuryMar 20, 2026Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 24656420 ↗
  • InjuryJan 13, 2026Unstable

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 24063623 ↗
  • InjuryJan 6, 2026Unstable

    REVISION SURGERY - DUE TO INSTABILITY AND POSSIBPLE INFECTION.

    Patient problems: Joint Laxity; Unspecified Infection

    MAUDE report 23986275 ↗
  • InjuryDec 8, 2025Unstable; Unstable

    INSTABILITY

    Patient problems: Joint Laxity

    MAUDE report 23737076 ↗
  • InjuryNov 26, 2025Unstable; Unstable

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 23653549 ↗
  • InjuryJul 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 22620386 ↗
  • InjuryJul 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - POLY SWAP.

    Patient problems: Insufficient Information

    MAUDE report 22599495 ↗
  • InjuryJul 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 22465234 ↗
  • InjuryJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 22401330 ↗
  • InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 22303249 ↗
  • InjuryJun 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 22155442 ↗
  • InjuryMay 28, 2025Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 22102173 ↗
  • InjuryMar 18, 2025Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 21638649 ↗
  • InjuryJan 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 21278650 ↗
  • InjuryJan 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 21261106 ↗
  • InjuryJan 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 21261101 ↗
  • InjuryJan 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 21214155 ↗
  • InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATIENT PRESENTED WITH POSSIBLE INFECTION. POLY SWAP. AND GLENOID HEAD SWAP.

    Patient problems: Unspecified Infection

    MAUDE report 21132039 ↗

Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page