DONJOY PERFORMANCE DP163CT02Recalls & reported problems

Djo, LLC · DJP COLDFORM SHOULDER WRAP, BLACK, UNIVERSAL · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Sep 25, 2025Open, Classified

    The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.

    Z-0722-2026 · Root cause: Labeling Change Control · FDA record ↗

Other Pack, hot or cold, reusable devices

All Pack, hot or cold, reusable devices →