The RUMI System UMW676Recalls & reported problems
Coopersurgical, Inc. · Disposable Uterine Manipulator Tip; Tip Diameter 6.7mm · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888937001440), received between Apr 23, 2020 and Aug 3, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Device Markings/Labelling Problem1
- Improper or Incorrect Procedure or Method1
- Physical Resistance/Sticking1
Patient problems
- No Known Impact Or Consequence To Patient1
- Rupture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2020Improper or Incorrect Procedure or Method; Device Markings/Labelling Problem; Physical Resistance/Sticking
PER MEDWATCH 5095205- "PATIENT WAS ADMITTED FOR ROBOTIC LAPAROSCOPIC FULGURATION WITH EXCISION OF LESION, SMALL TIP IN THE UTERUS AT SPECIFIC POINT IN THE PROCEDURE, THE SURGEON INSTRUCTED THE SURGICAL TECH TO INSTILL METHYLENE BLUE INTO THE DYE PORT ON THE RUMI. ROBOTIC SURGERIES ARE OFTEN DONE IN THE DARK AND COOPER SURGICAL RUMI PORTS HAVE NO DIFFERENTIATION. THEY LOOK EXACTLY THE SAME WITH UNIVERSAL CONNECTIVITY WHICH CAUSED THE TECH TO CHOOSE WRONG PORT. HE TRIED TO INSTILL THE DYE WHEN SHE MET RESISTANCE, SHE ESCALATED TO THE SURGEON. THE SURGEON INSTRUCTED HET TO TRY AGAIN, SHE MET RESISTANCE THE SECOND TIME AFTER WHICH THE ASSISTANT SURGEON ENTERED INTO THE SURGICAL ARENA, HE ATTEMPTED TO INSTILL THE DYE AND IT WENT INTO THE UTERINE BALLOON CAUSING IT TO RUPTURE THE UTERUS. A 3CM RENT WAS NOTICED AFTER DEFLATING THE RUMI. IT WAS NOTED AT THE TIME AND SUBSEQUENT TO PROCEDURE THAT TWO PORTS DO NOT DIFFERENTIATE IN COLOR OR CONNECTIVITY WHICH LENDS ITSELF TO CONFUSION DURING THE PROCEDURE". UMW676 UTERINE MANIPULATOR TIP W (B)(4).
Patient problems: Rupture
MAUDE report 10357328 ↗MalfunctionApr 23, 2020Burst Container or Vessel
REPORT STATED: "BALLOON EXPLODED WHEN SURGEON INFLATED WITH NORMAL SALINE. DURING LAPAROSCOPIC MYOMECTOMY + OVARIAN CYSTECTOMY + HYSTEROSCOPY. CHECKED FOR THE COMPLETENESS OF THE BALLOON AND SEARCHED - NO FOREIGN BODY REMAINED IN PATIENT BODY". 1216677-2020-00096-1, UMW676 UTERINE MANIPULATOR TIP W E-COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9993097 ↗
Other Cannula, Manipulator/Injector, Uterine devices
- VCARE 60-6085-201AConmed Corporation · Cannula, manipulator/injector, uterine198 reports
- Cook G17147Cook Incorporated · Cannula, manipulator/injector, uterine68 reports
- Mckesson 11-19610Mckesson Medical-Surgical Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
- Mckesson 11-19611Mckesson Medical-Surgical Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
- H/S Elliptosphere™ Catheter with Integrated Stylet 61-4005SCoopersurgical, Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
- H/S Elliptosphere™ Catheter • 5 French 61-4005Coopersurgical, Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
- H/S Catheter with Integrated Stylet • 5 French 61-3005Coopersurgical, Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
- VCARE 60-6085-202AConmed Corporation · Cannula, manipulator/injector, uterine46 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.