Skip to content

The RUMI System UMW676Recalls & reported problems

Coopersurgical, Inc. · Disposable Uterine Manipulator Tip; Tip Diameter 6.7mm · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888937001440), received between Apr 23, 2020 and Aug 3, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Device Markings/Labelling Problem1
  • Improper or Incorrect Procedure or Method1
  • Physical Resistance/Sticking1

Patient problems

  • No Known Impact Or Consequence To Patient1
  • Rupture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 3, 2020Improper or Incorrect Procedure or Method; Device Markings/Labelling Problem; Physical Resistance/Sticking

    PER MEDWATCH 5095205- "PATIENT WAS ADMITTED FOR ROBOTIC LAPAROSCOPIC FULGURATION WITH EXCISION OF LESION, SMALL TIP IN THE UTERUS AT SPECIFIC POINT IN THE PROCEDURE, THE SURGEON INSTRUCTED THE SURGICAL TECH TO INSTILL METHYLENE BLUE INTO THE DYE PORT ON THE RUMI. ROBOTIC SURGERIES ARE OFTEN DONE IN THE DARK AND COOPER SURGICAL RUMI PORTS HAVE NO DIFFERENTIATION. THEY LOOK EXACTLY THE SAME WITH UNIVERSAL CONNECTIVITY WHICH CAUSED THE TECH TO CHOOSE WRONG PORT. HE TRIED TO INSTILL THE DYE WHEN SHE MET RESISTANCE, SHE ESCALATED TO THE SURGEON. THE SURGEON INSTRUCTED HET TO TRY AGAIN, SHE MET RESISTANCE THE SECOND TIME AFTER WHICH THE ASSISTANT SURGEON ENTERED INTO THE SURGICAL ARENA, HE ATTEMPTED TO INSTILL THE DYE AND IT WENT INTO THE UTERINE BALLOON CAUSING IT TO RUPTURE THE UTERUS. A 3CM RENT WAS NOTICED AFTER DEFLATING THE RUMI. IT WAS NOTED AT THE TIME AND SUBSEQUENT TO PROCEDURE THAT TWO PORTS DO NOT DIFFERENTIATE IN COLOR OR CONNECTIVITY WHICH LENDS ITSELF TO CONFUSION DURING THE PROCEDURE". UMW676 UTERINE MANIPULATOR TIP W (B)(4).

    Patient problems: Rupture

    MAUDE report 10357328 ↗
  • MalfunctionApr 23, 2020Burst Container or Vessel

    REPORT STATED: "BALLOON EXPLODED WHEN SURGEON INFLATED WITH NORMAL SALINE. DURING LAPAROSCOPIC MYOMECTOMY + OVARIAN CYSTECTOMY + HYSTEROSCOPY. CHECKED FOR THE COMPLETENESS OF THE BALLOON AND SEARCHED - NO FOREIGN BODY REMAINED IN PATIENT BODY". 1216677-2020-00096-1, UMW676 UTERINE MANIPULATOR TIP W E-COMPLAINT (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9993097 ↗

FDA clearance and classification

The FDA regulates product code LKF (cannula, manipulator/injector, uterine) under 21 CFR 884.4530 as a Class II device. 21 CFR 884.4530 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the The RUMI System UMW676

Has the The RUMI System UMW676 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888937001440), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the The RUMI System UMW676?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 23, 2020 and Aug 3, 2020. In those reports, the most common specific device problems are “Burst Container or Vessel” (1), “Device Markings/Labelling Problem” (1) and “Improper or Incorrect Procedure or Method” (1); the most common patient problem is “Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the The RUMI System UMW676?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the The RUMI System UMW676, and is it still sold?

The FDA's device database lists Coopersurgical, Inc. as the company responsible for it (brand The RUMI System, model UMW676). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Cannula, Manipulator/Injector, Uterine devices

All Cannula, Manipulator/Injector, Uterine devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page