Frigitronics 139Recalls & reported problems
Coopersurgical, Inc. · Retinal, Spherical Probe. Tip Sphere: 3.0mm diameter. Shaft: 28mm length x 2.1mm diameter. · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888937004168), received between Aug 13, 2020 and Jun 9, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Excessive Cooling” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Excessive Cooling1
- Infusion or Flow Problem1
- Material Split, Cut or Torn1
- Output Problem1
- Temperature Problem1
Patient problems
- Insufficient Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2021Output Problem
CUSTOMER STATED: "FREEZES WHILE HOLDING PROBE". 1216677-2021-00127-1, 139 N20 SPHERICAL PROBE, E-COMPLAINT- (B)(4).
Patient problems: Insufficient Information
MAUDE report 11964574 ↗MalfunctionMay 10, 2021Infusion or Flow Problem; Temperature Problem
CUSTOMER STATED "EVAL". REPAIR TECH STATED "FOUND - LOW DEFROST FLOW - REPLACED FILTER". (B)(4). N20 SPHERICAL PROBE 139 (B)(4).
Patient problems: Insufficient Information
MAUDE report 11797485 ↗MalfunctionAug 13, 2020Excessive Cooling; Material Split, Cut or Torn
ORDER: 94632. FOUND;SILICON TORN. HIGH FREEZE FLOW.15MM SHOULD BE 10MM. REF E-COMPLAINT-(B)(4). N20 SPHERICAL PROBE 139 E-COMPLAINT-(B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10401885 ↗
Other Unit, Cryotherapy, Ophthalmic devices
- Frigitronics 124Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic2 reports
- Frigitronics CE-2000Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
- Frigitronics 161Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
- Frigitronics 147Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
- Frigitronincs 126Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.