Skip to content

Frigitronics 139Recalls & reported problems

Coopersurgical, Inc. · Retinal, Spherical Probe. Tip Sphere: 3.0mm diameter. Shaft: 28mm length x 2.1mm diameter. · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888937004168), received between Aug 13, 2020 and Jun 9, 2021 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Excessive Cooling” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Excessive Cooling1
  • Infusion or Flow Problem1
  • Material Split, Cut or Torn1
  • Output Problem1
  • Temperature Problem1

Patient problems

  • Insufficient Information2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2021Output Problem

    CUSTOMER STATED: "FREEZES WHILE HOLDING PROBE". 1216677-2021-00127-1, 139 N20 SPHERICAL PROBE, E-COMPLAINT- (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 11964574 ↗
  • MalfunctionMay 10, 2021Infusion or Flow Problem; Temperature Problem

    CUSTOMER STATED "EVAL". REPAIR TECH STATED "FOUND - LOW DEFROST FLOW - REPLACED FILTER". (B)(4). N20 SPHERICAL PROBE 139 (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 11797485 ↗
  • MalfunctionAug 13, 2020Excessive Cooling; Material Split, Cut or Torn

    ORDER: 94632. FOUND;SILICON TORN. HIGH FREEZE FLOW.15MM SHOULD BE 10MM. REF E-COMPLAINT-(B)(4). N20 SPHERICAL PROBE 139 E-COMPLAINT-(B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10401885 ↗