Frigitronics 139Recalls & reported problems
Coopersurgical, Inc. · Retinal, Spherical Probe. Tip Sphere: 3.0mm diameter. Shaft: 28mm length x 2.1mm diameter. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888937004168), received between Aug 13, 2020 and Jun 9, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Excessive Cooling” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Excessive Cooling1
- Infusion or Flow Problem1
- Material Split, Cut or Torn1
- Output Problem1
- Temperature Problem1
Patient problems
- Insufficient Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2021Output Problem
CUSTOMER STATED: "FREEZES WHILE HOLDING PROBE". 1216677-2021-00127-1, 139 N20 SPHERICAL PROBE, E-COMPLAINT- (B)(4).
Patient problems: Insufficient Information
MAUDE report 11964574 ↗MalfunctionMay 10, 2021Infusion or Flow Problem; Temperature Problem
CUSTOMER STATED "EVAL". REPAIR TECH STATED "FOUND - LOW DEFROST FLOW - REPLACED FILTER". (B)(4). N20 SPHERICAL PROBE 139 (B)(4).
Patient problems: Insufficient Information
MAUDE report 11797485 ↗MalfunctionAug 13, 2020Excessive Cooling; Material Split, Cut or Torn
ORDER: 94632. FOUND;SILICON TORN. HIGH FREEZE FLOW.15MM SHOULD BE 10MM. REF E-COMPLAINT-(B)(4). N20 SPHERICAL PROBE 139 E-COMPLAINT-(B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10401885 ↗
FDA clearance and classification
The FDA regulates product code HQA (unit, cryotherapy, ophthalmic) under 21 CFR 886.4170 as a Class II device. 21 CFR 886.4170 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Frigitronics 139
Has the Frigitronics 139 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888937004168), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Frigitronics 139?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 13, 2020 and Jun 9, 2021. In those reports, the most common specific device problems are “Excessive Cooling” (1), “Infusion or Flow Problem” (1) and “Material Split, Cut or Torn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Frigitronics 139?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Frigitronics 139, and is it still sold?
The FDA's device database lists Coopersurgical, Inc. as the company responsible for it (brand Frigitronics, model 139). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Unit, Cryotherapy, Ophthalmic devices
- Frigitronics 124Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic2 reports
- Frigitronics 161Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
- Frigitronics CE-2000Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
- Frigitronincs 126Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
- Frigitronics 147Coopersurgical, Inc. · Unit, Cryotherapy, Ophthalmic1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page