Skip to content

Mara DDK-16-050Recalls & reported problems

Coopersurgical, Inc. · Mara Water Vapor Probe Kit · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888937026887), received between Oct 11, 2022 and Jun 12, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A MARA ABLATION PROCEDURE ON (B)(6) 2024. FIVE WEEKS POST PROCEDURE, THE PATIENT UNDERWENT A HYSTERECTOMY DUE TO A SIGNIFICANT UTERINE INFECTION. DOCTOR UNSURE IF THE INFECTION IS RELATED TO THE MARA PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE. . DDK-16-050 MARA PROBE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19519438 ↗
  • InjuryNov 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18185673 ↗
  • InjuryNov 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18084781 ↗
  • MalfunctionOct 11, 2022Leak/Splash; Leak/Splash

    ACCOUNT SAID THAT THEY SET UP FOR THE PROCEDURE AND ONCE THEY GOT TO THE PROBE SELF TEST ALERT 176 CAME UP ON THE SCREEN. THEY NEVER GOT TO THE FIRST CHECK MARK ON THE PROBE SELF TEST. THEY MADE SURE THE PROBE WAS SECURE IN THE CARTRIDGE. DR. MAZLIN SAID THAT SALINE WAS GOING INTO THE PROBE COVER AND BEGAN LEAKING OUT. AFTER THEY HIT CLEAR ON THE ALERT IT WOULD BRING THEM BACK TO THE PROBE SELF TEST. SAME THING WOULD HAPPEN. THEY NEVER GOT TO THE FIRST CHECK MARK. THE ACCOUNT SWITCHED OUT THE PROBE AND TRIED AGAIN WITH THE SAME ALERT APPEARING ON THE SCREEN. AFTER SEVERAL MORE ATTEMPTS THEY ABORTED THE PROCEDURE. ONCE THE SYRINGE WAS UNLOCKED FROM THE TOP OF THE MARA CONSOLE THEY NOTICED SALINE AT THE VERY BOTTOM. THE MA DRIED IT UP WITH GAUZE. I WENT TO THE ACCOUNT AND CHECKED TO SEE IF EITHER PROBE WAS LEAKING FROM THE BOTTOM OF THE SYRINGE. BOTH SEEMED TO BE SECURE. I OPENED ONE OF MY DEMO PROBES TO ENSURE THE CONSOLE WAS NOT DAMAGED. I WAS ABLE TO RUN THROUGH THE PROBE SELF TEST WITHOUT ANY ISSUES. I DID RECEIVE AN ALERT 101. I HIT CLEAR AND I WAS ABLE TO PROCEED THROUGH THE INTEGRITY AND PATENCY TESTS WITH OUT ISSUE. MARA CONSOLE SEEMS TO BE WORKING OK AFTER THE UNSUCCESSFUL P

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15575314 ↗

Other Device, Thermal Ablation, Endometrial devices

All Device, Thermal Ablation, Endometrial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page