The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024007383), received between Jul 13, 2021 and Feb 13, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Deformation3
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2023Fracture; Fracture
IT WAS REPORTED THAT TWO VITALITY CONNECTORS BROKE, THE TIP OF ONE VITALITY T20 FINAL DRIVER BROKE, THE TIP OF A VITALITY T27 DRIVER TWISTED, AND THE TIPS OF TWO VITAL NAVIGATION T27 DRIVERS TWISTED INTRA-OPERATIVELY. OTHER CONNECTORS AND DRIVERS WERE USED TO COMPLETE THE PROCEDURE. THERE WAS A DELAY TO THE PROCEDURE THAT DID NOT LEAD TO PATIENT IMPACTS. THIS IS REPORT ONE OF SIX FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16362866 ↗MalfunctionJul 13, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTIES WITH SET SCREW TIGHTENING INTRA-OPERATIVELY. THE SET SCREWS ON THREE CROSS CONNECTORS CROSS-THREADED, THE DRIVER TIP FRACTURED, AND THE TORQUE HANDLE DID NOT LIMIT THE TORQUE. THE DRIVER TIP WAS RECOVERED, THE CROSS-CONNECTORS WERE REMOVED, AND ALTERNATIVE CROSS-CONNECTORS WERE USED TO COMPLETE THE PROCEDURE WITHOUT REPORTED PATIENT IMPACTS. THIS IS REPORT THREE OF FIVE FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158197 ↗MalfunctionJul 13, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTIES WITH SET SCREW TIGHTENING INTRA-OPERATIVELY. THE SET SCREWS ON THREE CROSS CONNECTORS CROSS-THREADED, THE DRIVER TIP FRACTURED, AND THE TORQUE HANDLE DID NOT LIMIT THE TORQUE. THE DRIVER TIP WAS RECOVERED, THE CROSS-CONNECTORS WERE REMOVED, AND ALTERNATIVE CROSS-CONNECTORS WERE USED TO COMPLETE THE PROCEDURE WITHOUT REPORTED PATIENT IMPACTS. THIS IS REPORT TWO OF FIVE FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158194 ↗MalfunctionJul 13, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTIES WITH SET SCREW TIGHTENING INTRA-OPERATIVELY. THE SET SCREWS ON THREE CROSS CONNECTORS CROSS-THREADED, THE DRIVER TIP FRACTURED, AND THE TORQUE HANDLE DID NOT LIMIT THE TORQUE. THE DRIVER TIP WAS RECOVERED, THE CROSS-CONNECTORS WERE REMOVED, AND ALTERNATIVE CROSS-CONNECTORS WERE USED TO COMPLETE THE PROCEDURE WITHOUT REPORTED PATIENT IMPACTS. THIS IS REPORT ONE OF FIVE FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158188 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.