GemLock® RHD2.5Recalls & reported problems
ZimVie US Corp LLC · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024016705), received between Jul 1, 2022 and May 14, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Detachment of Device or Device Component1
- Mechanical Problem1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2026Separation Failure
DOCTOR REPORTED THAT DEVICE GOT STUCK WHEN INSERTING IMPLANT, THERE IS FRACTURE RISK.PROCEDURE WAS COMPLETED USING ANOTHER A SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25170366 ↗MalfunctionApr 10, 2026Fracture
IT WAS REPORTED THAT THE LATCHLOCK DRIVER FRACTURED AT THE TIP DURING THE IMPLANT PLACEMENT. NO IMPACT ON THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24843268 ↗MalfunctionApr 10, 2026Fracture; Mechanical Problem
IT WAS REPORTED THAT THE LATCHLOCK DRIVER FRACTURED AT THE TIP DURING THE IMPLANT PLACEMENT. NO IMPACT ON THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24843175 ↗MalfunctionSep 5, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE DRIVER BROKE AT THE TIP. THE PATIENT IS FINE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22977347 ↗MalfunctionAug 21, 2025Fracture; Fracture
IT WAS REPORTED THAT DRIVER BROKE DURING PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22865493 ↗MalfunctionJul 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE IMPLANT BECAME DETACHED FROM THE DRIVER. THE DOCTOR CONFIRMS THAT THE DENTAL SURGICAL PROCEDURE WAS COMPLETED WITH ANOTHER IMPLANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14883832 ↗
FDA clearance and classification
The FDA regulates product code NDP (accessories, implant, dental, endosseous) under 21 CFR 872.3980 as a Class I device. 21 CFR 872.3980 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the GemLock® RHD2.5
Has the GemLock® RHD2.5 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00889024016705), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the GemLock® RHD2.5?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jul 1, 2022 and May 14, 2026. In those reports, the most common specific device problems are “Fracture” (4), “Detachment of Device or Device Component” (1) and “Mechanical Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GemLock® RHD2.5?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GemLock® RHD2.5 FDA approved?
Devices of this type (product code NDP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by ZimVie US Corp LLC.
Who makes the GemLock® RHD2.5, and is it still sold?
The FDA's device database lists ZimVie US Corp LLC as the company responsible for it (brand GemLock®, model RHD2.5). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS devices
- N/A 68015189Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous348 reports
- N/A 68015191Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous277 reports
- N/A 68015193Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous258 reports
- N/A 68015187Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous234 reports
- N/A 68015188Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous166 reports
- N/A 68015192Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous144 reports
- N/A 68015190Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous131 reports
- NobelReplace® 37045Nobel Biocare AB · Accessories, implant, dental, endosseous2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page