Skip to content

Trabecular Metal® 6202-52-20Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of Trabecular Metal (74)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024118867), received between Nov 24, 2021 and Nov 13, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K021891, K093561); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Difficult to Insert1
  • Malposition of Device1
  • Unstable1

Patient problems

  • Failure of Implant2
  • Aseptic loosening1
  • Joint Dislocation1
  • No Clinical Signs, Symptoms or Conditions1
  • Osteopenia/ Osteoporosis1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP REVISION DUE TO RECURRENT DISLOCATION AND CUP LOOSENING. THE G7 OSSEOTI MULTI HOLE WAS REPLACED WITH A DUAL MOBILITY. IT WAS CONFIRMED THAT NO FURTHER INFORMATION COULD BE PROVIDED.

    Patient problems: Aseptic loosening; Joint Dislocation

    MAUDE report 23542027 ↗
  • MalfunctionSep 2, 2025Difficult to Insert; Difficult to Insert

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22937309 ↗
  • InjuryDec 19, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Osteopenia/ Osteoporosis

    MAUDE report 18354080 ↗
  • InjuryNov 24, 2021Unstable; Malposition of Device; Unstable; Malposition of Device

    IT WAS REPORTED THAT PATIENT UNDERWENT A HIP REVISION AFTER AN UNKNOWN AMOUNT OF TIME POST IMPLANTATION DUE TO ABNORMAL WEAR OF THE POLY LINER, INSTABILITY AND MALPOSITIONING OF THE DEVICES. THE LINER, CUP AND COMPETITORS HEAD WERE REMOVED AND REPLACED. ATTEMPTS HAVE BEEN MADE AND THE ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 12870077 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K021891, K093561. One submission often covers many models, so check the product and lot details in each record.

  • Z-1496-2014 · Apr 1, 2014 · Terminated
    TRABECULAR METAL MODULAR ACETABULAR SYSTEM, Shell with Cluster Holes, Porous 60mm. Catalog number 00-6202-060-22; Hip implant component.
  • Z-3279-2011 · Sep 1, 2011 · Terminated
    The Trabecular Metal Modular Acetabular System is a modular acetabular cup system consisting of a shell with a preassembled locking ring, a liner, and optional screws. The Shells
  • Z-0060-2011 · Aug 18, 2010 · Terminated
    Zimmer Trabecular metal Modular Acetabular system 56 mm Porous shell with cluster holes, Catalog 00-6202-056-22, Zimmer Inc., Warsaw, IN. This device is indicated for primary or re

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page