This Zimmer hip implant, a porous uncemented metal-and-polymer hip prosthesis, appears in 43 adverse event reports. Most describe the implant loosening or shifting, breaking, or sitting poorly in the socket, leading to pain, leg-length differences, dislocation, and repeat revision surgery; some describe infection around the implant. There were34 injuries and no deaths.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
34
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00889024119840), received between Dec 4, 2019 and May 15, 2026 — classified as 34 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth16
  • Fracture8
  • Migration8
  • Adverse Event Without Identified Device or Use Problem7
  • Activation, Positioning or Separation Problem1
  • Device Dislodged or Dislocated1
  • Loss of or Failure to Bond1
  • Material Deformation1

Patient problems

  • Failure of Implant13
  • No Clinical Signs, Symptoms or Conditions10
  • Pain9
  • Aseptic loosening4
  • Deformity/ Disfigurement4
  • Unequal Limb Length4
  • Foreign Body In Patient3
  • Osteopenia/ Osteoporosis3

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST-HIP PROCEDURE, THE PATIENT PRESENTED WITH SEPTIC SHOCK DUE TO MRSA INFECTION IN THE HIP. THEY HAD FEVERS, DIFFICULTY WALKING, SWELLING, ABSCESS, AND POSITIVE CULTURES FOR MRSA AND SEPTIC SHOCK. THE HEAD WAS EXCHANGED IN AN I&D PROCEDURE BEFORE THE STAGE ONE REVISION ONE MONTH LATER. OP RECORDS NOTE PURULENT FLUID, LARGE FASCIAL DEHISCENCE. ALL COMPONENTS WERE REPLACED WITH A CEMENT SPACER DURING STAGE ONE WITHOUT COMPLICATIONS. THE STAGE TWO REVISION TOOK PLACE FOUR MONTHS LATER. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Abscess; Ambulation or Postural Difficulties; Bacterial Infection; Medical device site infection; Purulent Discharge; Scar Tissue; Sepsis; Septic Shock; Swelling/ Edema; Wound Dehiscence

    MAUDE report 25183724 ↗
  • InjuryMay 15, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST-HIP PROCEDURE, THE PATIENT PRESENTED WITH SEPTIC SHOCK DUE TO MRSA INFECTION IN THE HIP. THEY HAD FEVERS, DIFFICULTY WALKING, SWELLING, ABSCESS, AND POSITIVE CULTURES FOR MRSA AND SEPTIC SHOCK. THE HEAD WAS EXCHANGED IN AN I&D PROCEDURE BEFORE THE STAGE ONE REVISION ONE MONTH LATER. OP RECORDS NOTE PURULENT FLUID, LARGE FASCIAL DEHISCENCE. ALL COMPONENTS WERE REPLACED WITH A CEMENT SPACER DURING STAGE ONE WITHOUT COMPLICATIONS. THE STAGE TWO REVISION TOOK PLACE FOUR MONTHS LATER. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Abscess; Ambulation or Postural Difficulties; Bacterial Infection; Medical device site infection; Purulent Discharge; Scar Tissue; Sepsis; Septic Shock; Swelling/ Edema; Wound Dehiscence

    MAUDE report 25183690 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY FOUR MONTHS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO INSTABILITY, RECURRENT DISLOCATIONS, AND IMPINGEMENT FROM POOR POSITIONING OF IMPLANTS. THE SURGEON ATTRIBUTED THE RECURRENT DISLOCATION TO POOR CUP PLACEMENT. THE TROCHANTERIC CLAW HAD ANTERIOR FIXATION RESULTING IN TISSUE IMPINGEMENT. DURING THE REVISION, A COMPETITOR CUP WAS CEMENTED INTO THE EXISTING CUP; THE HEAD AND LINER WERE EXCHANGED; AND THE STEM AND CLAW WERE RETAINED WITH SOME TISSUE AROUND THE GREATER TROCHANTER DEBRIDED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation; Joint Impingement; Joint Laxity; Pain

    MAUDE report 24455798 ↗
  • InjuryDec 15, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Aseptic loosening

    MAUDE report 23811676 ↗
  • InjuryNov 21, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL RIGHT HIP ARTHROPLASTY AND WAS IMPLANTED WITH ZIMMER BIOMET PRODUCTS. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO LOOSENING. THE CUP AND SCREWS WERE EXPLANTED.

    Patient problems: Aseptic loosening

    MAUDE report 23616756 ↗
  • InjurySep 8, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT IS BEING CONSIDERED FOR A SECOND REVISION ON AN UNKNOWN DATE DUE TO INSUFFICIENT FIXATION OF THE CUP. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Aseptic loosening

    MAUDE report 22991046 ↗
  • InjurySep 8, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT IS BEING CONSIDERED FOR A SECOND REVISION ON AN UNKNOWN DATE DUE TO INSUFFICIENT FIXATION OF THE CUP. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Aseptic loosening

    MAUDE report 22991037 ↗
  • InjuryAug 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT REPORTED TENDERNESS OVER THE GREATER TROCHANTER CONSISTENT WITH BURSITIS APPROXIMATELY 1 YEAR LATER. THE PATIENT WAS USING CANES TO AMBULATE AND EXPERIENCED A MODERATE LIMP. THE SURGEON PROVIDED A CORTISONE INJECTION AND A TOPICAL ANTI-INFLAMMATORY CREAM, AND THE ISSUES WERE REPORTEDLY RESOLVED. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Bursitis; Pain

    MAUDE report 22884383 ↗
  • MalfunctionJul 30, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE ACETABULAR SCREW HAD METAL SHAVING PRESENT ON HEAD OF SCREW. THERE WAS NO PATIENT IMPACT, NO MEDICAL/SURGICAL INTERVENTIONS AND NO SURGICAL DELAY. THERE WERE NO CONTRIBUTING FACTORS TO THE EVENT. ADDITIONALLY, THERE WAS NO FOREIGN BODY RETAINED IN THE PATIENT. THE SCREW WAS DISCARDED BY THE USER FACILITY. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22668789 ↗
  • InjuryJul 22, 2025Fracture; Fracture

    IT WAS REPORTED THE SCREWS BROKE DURING INSERTION. THE SURGEON COMPLETED THE SURGERY USING SHORT SCREWS WITHOUT FORCING THEM. THE BROKEN TIP OF THE SCREW WAS LEFT INSIDE THE PATIENT¿S BODY AT THE SURGEON¿S DISCRETION.

    Patient problems: Foreign Body In Patient

    MAUDE report 22564620 ↗
  • InjuryJan 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT IS EXPERIENCING PAIN, SWELLING, BURNING SENSATION AND HAS A NICKEL ALLERGY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Burning Sensation; Hypersensitivity/Allergic reaction; Pain; Swelling/ Edema

    MAUDE report 21228370 ↗
  • InjuryDec 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: Pain

    MAUDE report 20820287 ↗
  • InjurySep 11, 2024Migration; Migration

    IT WAS REPORTED THAT APPROXIMATELY 5 MONTHS POST IMPLANTATION OF A LEFT TOTAL HIP ARTHROPLASTY, THE PATIENT IS BEING CONSIDERED FOR A REVISION DUE TO HIP PAIN AND CUP MIGRATION MEDIALLY INTO THE PELVIS. THERE ARE NO KNOWN CONTRIBUTING CONDITIONS EXCEPT FOR POOR BONE QUALITY. NO ADDITIONAL INFORMATION WAS AVAILABLE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 20189498 ↗
  • MalfunctionJun 5, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING THE SURGERY THE SURGEON ATTEMPTED TO INSERT THE SCREW AND IT BENT. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19468954 ↗
  • MalfunctionMar 11, 2024Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18877252 ↗
  • MalfunctionFeb 20, 2024

    BONE SCREW BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18739446 ↗
  • InjuryDec 18, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY TWO YEARS AFTER A SECOND STAGE REVISION DUE TO ACETABULAR LOOSENING. THE HEAD, SHELL, LINER, AND AUGMENT WERE REPLACED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Osteopenia/ Osteoporosis

    MAUDE report 18344633 ↗
  • InjuryDec 18, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Osteopenia/ Osteoporosis

    MAUDE report 18344625 ↗
  • MalfunctionDec 14, 2023Fracture; Fracture

    IT WAS REPORTED WHILE SURGEON WAS IMPLANTING THE SCREW THROUGH A G7 SHELL, THE OUTER PORTION OF THE SCREW HEAD SHEARED OFF AND THE SCREW WENT COMPLETELY THROUGH THE ACETABULAR COMPONENT. THE DOCTOR WAS ABLE TO REMOVE THE SHELL, THEN THE SCREW, AND TRY AGAIN WITH A NEW SCREW. THE METAL SHARD FROM THE SCREW WAS RECOVERED AND REMOVED FROM THE PATIENT. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18328884 ↗
  • InjuryDec 1, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT WAS BORN WITH HIP DYSPLASIA. THE PATIENT¿S CUP STARTED TO WEAR AND WAS REVISED. ONE YEAR POST IMPLANTATION, THE PATIENT FELL, AND THE CUP WAS FOUND TO BE LOOSE. THE SURGEON RECOMMENDED A CUSTOM IMPLANT.

    Patient problems: Fall

    MAUDE report 18245616 ↗
  • InjuryNov 29, 2023Migration; Migration

    IT WAS REPORTED A PATIENT HAD A REVISION OF A RIGHT HIP PROSTHESIS. SUBSEQUENTLY, APPROXIMATELY TWO WEEKS LATER, RADIOGRAPHIC IMAGING DISPLAYED DISSOCIATION OF THE CUP FROM THE PELVIS AND ¿HUGE¿ PELVIC DEFECTS. UPON FOLLOW-UP, THE PATIENT REPORTS HAVING SEVERE PAIN AND ASSESSMENT INDICATED THAT THE RIGHT LEG WAS SHORTENED. A PMI PRODUCT IS CURRENTLY BEING PLANNED FOR THE PATIENT¿S NEXT REVISION.

    Patient problems: Deformity/ Disfigurement; Pain; Unequal Limb Length

    MAUDE report 18231790 ↗
  • InjuryNov 29, 2023Migration; Migration

    IT WAS REPORTED A PATIENT HAD A REVISION OF A RIGHT HIP PROSTHESIS. SUBSEQUENTLY, APPROXIMATELY TWO WEEKS LATER, RADIOGRAPHIC IMAGING DISPLAYED DISSOCIATION OF THE CUP FROM THE PELVIS AND ¿HUGE¿ PELVIC DEFECTS. UPON FOLLOW-UP, THE PATIENT REPORTS HAVING SEVERE PAIN AND ASSESSMENT INDICATED THAT THE RIGHT LEG WAS SHORTENED. A PMI PRODUCT IS CURRENTLY BEING PLANNED FOR THE PATIENT¿S NEXT REVISION.

    Patient problems: Deformity/ Disfigurement; Pain; Unequal Limb Length

    MAUDE report 18231783 ↗
  • InjuryNov 29, 2023Migration; Migration

    IT WAS REPORTED A PATIENT HAD A REVISION OF A RIGHT HIP PROSTHESIS. SUBSEQUENTLY, APPROXIMATELY TWO WEEKS LATER, RADIOGRAPHIC IMAGING DISPLAYED DISSOCIATION OF THE CUP FROM THE PELVIS AND ¿HUGE¿ PELVIC DEFECTS. UPON FOLLOW-UP, THE PATIENT REPORTS HAVING SEVERE PAIN AND ASSESSMENT INDICATED THAT THE RIGHT LEG WAS SHORTENED. A PMI PRODUCT IS CURRENTLY BEING PLANNED FOR THE PATIENT¿S NEXT REVISION.

    Patient problems: Deformity/ Disfigurement; Pain; Unequal Limb Length

    MAUDE report 18231754 ↗
  • InjuryNov 29, 2023Migration; Migration

    IT WAS REPORTED A PATIENT HAD A REVISION OF A RIGHT HIP PROSTHESIS. SUBSEQUENTLY, APPROXIMATELY TWO WEEKS LATER, RADIOGRAPHIC IMAGING DISPLAYED DISSOCIATION OF THE CUP FROM THE PELVIS AND ¿HUGE¿ PELVIC DEFECTS. UPON FOLLOW-UP, THE PATIENT REPORTS HAVING SEVERE PAIN AND ASSESSMENT INDICATED THAT THE RIGHT LEG WAS SHORTENED. A PMI PRODUCT IS CURRENTLY BEING PLANNED FOR THE PATIENT¿S NEXT REVISION.

    Patient problems: Deformity/ Disfigurement; Pain; Unequal Limb Length

    MAUDE report 18231731 ↗
  • InjurySep 22, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 17799149 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →