Trilogy® Longevity® 6310-50-32Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Adverse event reports
15
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00889024124721), received between Jun 16, 2021 and Mar 16, 2026 — classified as 12 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated8
  • Adverse Event Without Identified Device or Use Problem4
  • Difficult to Insert3
  • Unintended Movement1

Patient problems

  • Joint Dislocation8
  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions3
  • Aseptic loosening1
  • Failure of Implant1
  • Metal Related Pathology1
  • Pain1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 16, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT APPROXIMATELY SEVEN WEEKS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO EXPERIENCING SEVERAL DISLOCATIONS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 24606424 ↗
  • InjuryNov 13, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP REVISION DUE TO RECURRENT DISLOCATION AND CUP LOOSENING. THE G7 OSSEOTI MULTI HOLE WAS REPLACED WITH A DUAL MOBILITY. IT WAS CONFIRMED THAT NO FURTHER INFORMATION COULD BE PROVIDED.

    Patient problems: Aseptic loosening; Joint Dislocation

    MAUDE report 23542042 ↗
  • MalfunctionAug 14, 2025Difficult to Insert; Difficult to Insert

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22809625 ↗
  • InjuryAug 11, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Dislocation

    MAUDE report 22779141 ↗
  • InjuryJul 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY 4 AND A HALF YEARS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO INSTABILITY. THE HEAD AND LINER WERE EXCHANGED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22492926 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT WAS REVISED APPROXIMATELY TWO YEARS LATER DUE TO A PSOAS CONFLICT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 22369906 ↗
  • InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY 4 AND A HALF YEARS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO INSTABILITY. THE HEAD AND LINER WERE EXCHANGED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22051934 ↗
  • InjuryAug 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    UPON RECEIPT OF ADDITIONAL INFORMATION, IT WAS DETERMINED THIS PRODUCT SHOULD NOT HAVE BEEN REPORTED UNDER THIS MFR NUMBER. THIS REPORT SHOULD BE VOIDED AND A CORRECTED REPORT WILL BE FILED UNDER MFR NUMBER (5648920).

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 19963304 ↗
  • InjuryDec 29, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED APPROXIMATELY TWELVE YEARS POST-IMPLANTATION DUE TO A DISLOCATION. DURING THE REVISION SURGERY, THE SURGEON NOTICED TRUNION WEAR AND A PSEUDOTUMOR POSSIBLY DUE TO METAL ION RELEASE. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Joint Dislocation; Metal Related Pathology

    MAUDE report 18418145 ↗
  • MalfunctionMay 15, 2023Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE LINER WOULD NOT SEAT. THE SURGEON REAMED FOR A 54MM CUP. THE CUP WAS IMPLANTED WITH ADDITIONAL FIXATION SCREWS. IT WAS ENSURED THAT THE SCREWS WERE NOT RAISED. THERE WAS NO VISIBLE SOFT TISSUE IN THE CUP AND ALIGNED SCALLOPS OF THE CUP WITH THE LINER. THE SURGEON IMPACTED WITH A 32MM IMPACTOR HOWEVER THE LINER STILL WOULD NOT SEAT. THERE WAS A 10-MINUTE DELAY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16930756 ↗
  • InjuryMar 28, 2023Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement

    IT WAS REPORTED A PATIENT UNDERWENT A CLOSED REDUCTION PROCEDURE TO CORRECT A DISLOCATION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Joint Dislocation

    MAUDE report 16627644 ↗
  • InjuryFeb 16, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: Joint Dislocation

    MAUDE report 16387811 ↗
  • InjuryOct 5, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 6 YEARS POST IMPLANTATION DUE TO CHRONIC DISLOCATION. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Dislocation

    MAUDE report 15546630 ↗
  • InjuryJul 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN IN THE UPPER SIDE OF THE HIP AND ALLEGED SENSATION OF THE TENDON MOVING WITHIN THE JOINT APPROXIMATELY 11 YEARS POST IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Pain

    MAUDE report 14939093 ↗
  • MalfunctionJun 16, 2021Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING A HIP PROCEDURE, THE LINER WAS OPENED AND FOUND POLY PROTRUDING AT THE BOTTOM OF THE LINER. THE SURGEON OPENED A NEW DEVICE TO COMPLETE THE SURGERY. THERE WAS APPROXIMATELY 30 MIN OF SURGERY DELAY. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12010271 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.