Skip to content

NA 00-7711-005-20Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889024131644), received between Aug 26, 2022 and Aug 26, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Naturally Worn” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K032726, K192660); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Naturally Worn1

Patient problems

  • Failure of Implant1
  • Metal Related Pathology1
  • Osteolysis1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 26, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THE PATIENT UNDERWENT A RIGHT HIP REVISION APPROXIMATELY 6 YEARS POST IMPLANTATION DUE TO HIGH COBALT ION LEVELS, PAIN, WEAR AND OSTEOLYSIS. DURING THE SURGERY, IT WAS DISCOVERED THAT THE FEMORAL HEAD HAD SIGNS OF TRUNIONOSIS. THE STEM WAS RETAINED, BUT THE HEAD AND LINER WERE REPLACED AND DISCARDED. NO ADDITIONAL INFORMATION.

    Patient problems: Failure of Implant; Metal Related Pathology; Osteolysis; Pain

    MAUDE report 15300169 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032726, K192660. One submission often covers many models, so check the product and lot details in each record.

  • Z-1191-2018 · Mar 1, 2018 · Terminated
    Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 7.5 Extended Offset Reduced Neck Length; Part Number: 00-7711-0
  • Z-1046-2016 · Feb 11, 2016 · Terminated
    M/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 62937093
  • Z-2033-2014 · May 23, 2014 · Terminated
    Zimmer M/L Taper Hip Stem Standard & Extended Offset, HIP JOINT PROSTHESIS, Sizes: 5.0, 7.5, 9.0, 10.0, 11.0, 12.5, 13.5, 15.0, 11.25, 17.25, 20.0 & 22.5. The femoral stem com

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page