NA 00-7711-009-10Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024131750), received between Jul 1, 2020 and Aug 6, 2024 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Erosion” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K032726, K161830, K192660); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Erosion2
- Adverse Event Without Identified Device or Use Problem1
- Corroded1
- Device Contaminated During Manufacture or Shipping1
- Difficult to Insert1
- Insufficient Device Problem Information1
- Naturally Worn1
- Separation Problem1
Patient problems
- Pain3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Failure of Implant1
- High metal ion levels1
- Insufficient Information1
- Joint Impingement1
- Metal Related Pathology1
- Muscle/Tendon Damage1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2024Difficult to Insert; Difficult to Insert
NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19917470 ↗InjuryFeb 27, 2024Material Erosion; Naturally Worn; Insufficient Device Problem Information; Material Erosion; Naturally Worn; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT HAD AN INITIAL RIGHT TOTAL HIP ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO PAIN AND ELEVATED CHROMIUM LEVELS. DURING THE REVISION, NECROTIC AND AVASCULAR TISSUE CONSISTENT WITH TRUNNIONOSIS WAS DEBRIDED FROM THE JOINT SPACE. THE STEM WAS CLEARED OF BLACK METAL DEBRIS, AND A NEW HEAD AND LINER WERE PLACED WHILE THE INITIAL STEM AND SHELL REMAINED INTACT. OVER TIME THE PATIENT¿S SERUM METAL ION LEVELS RETURNED TO NORMAL LEVELS.
Patient problems: Failure of Implant; High metal ion levels; Insufficient Information; Muscle/Tendon Damage; Necrosis; Osteolysis; Pain
MAUDE report 18787171 ↗InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT A CORTISONE INJECTION APPROXIMATELY THREE MONTHS POST IMPLANTATION, DUE TO RIGHT HIP IMPINGEMENT, TENDONITIS, AND RIGHT HIP PAIN APPROXIMATELY. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Joint Impingement; Pain
MAUDE report 17954412 ↗InjuryDec 12, 2022Separation Problem; Separation Problem
IT WAS REPORTED THAT THE AGENT DISASSEMBLED THE CERAMIC HEAD AND ASSEMBLED IT INTO THE BIPOLAR HEAD IN THE CORRECT WAY. AFTER THE DOCTOR IMPLANTED THE PRODUCT, HE FOUND THAT THE FEMORAL HEAD COULD NOT BE FIXED ON STEM STABLY AND FORMED A TAPER LOCKING. THERE WAS A 70 MINUTE DELAY IN THE SURGERY. PROLONGED ANESTHESIA TIME AND INCREASED BLOOD TRANSFUSION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15961921 ↗InjuryOct 13, 2021Corroded; Material Erosion; Corroded; Material Erosion
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Metal Related Pathology; Pain
MAUDE report 12629114 ↗MalfunctionJul 1, 2020Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED DURING INSPECTION AT THE (B)(6) WAREHOUSE THAT THE PACKAGE WAS DAMAGED. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Patient Involvement
MAUDE report 10218227 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032726, K161830, K192660. One submission often covers many models, so check the product and lot details in each record.
- Z-1191-2018 · Mar 1, 2018 · TerminatedZimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 7.5 Extended Offset Reduced Neck Length; Part Number: 00-7711-0
- Z-1046-2016 · Feb 11, 2016 · TerminatedM/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 62937093
- Z-2033-2014 · May 23, 2014 · TerminatedZimmer M/L Taper Hip Stem Standard & Extended Offset, HIP JOINT PROSTHESIS, Sizes: 5.0, 7.5, 9.0, 10.0, 11.0, 12.5, 13.5, 15.0, 11.25, 17.25, 20.0 & 22.5. The femoral stem com
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page