Skip to content

NA 00-7711-009-10Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024131750), received between Jul 1, 2020 and Aug 6, 2024 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Erosion” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K032726, K161830, K192660); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Erosion2
  • Adverse Event Without Identified Device or Use Problem1
  • Corroded1
  • Device Contaminated During Manufacture or Shipping1
  • Difficult to Insert1
  • Insufficient Device Problem Information1
  • Naturally Worn1
  • Separation Problem1

Patient problems

  • Pain3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Failure of Implant1
  • High metal ion levels1
  • Insufficient Information1
  • Joint Impingement1
  • Metal Related Pathology1
  • Muscle/Tendon Damage1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2024Difficult to Insert; Difficult to Insert

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917470 ↗
  • InjuryFeb 27, 2024Material Erosion; Naturally Worn; Insufficient Device Problem Information; Material Erosion; Naturally Worn; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT HAD AN INITIAL RIGHT TOTAL HIP ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO PAIN AND ELEVATED CHROMIUM LEVELS. DURING THE REVISION, NECROTIC AND AVASCULAR TISSUE CONSISTENT WITH TRUNNIONOSIS WAS DEBRIDED FROM THE JOINT SPACE. THE STEM WAS CLEARED OF BLACK METAL DEBRIS, AND A NEW HEAD AND LINER WERE PLACED WHILE THE INITIAL STEM AND SHELL REMAINED INTACT. OVER TIME THE PATIENT¿S SERUM METAL ION LEVELS RETURNED TO NORMAL LEVELS.

    Patient problems: Failure of Implant; High metal ion levels; Insufficient Information; Muscle/Tendon Damage; Necrosis; Osteolysis; Pain

    MAUDE report 18787171 ↗
  • InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT A CORTISONE INJECTION APPROXIMATELY THREE MONTHS POST IMPLANTATION, DUE TO RIGHT HIP IMPINGEMENT, TENDONITIS, AND RIGHT HIP PAIN APPROXIMATELY. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Joint Impingement; Pain

    MAUDE report 17954412 ↗
  • InjuryDec 12, 2022Separation Problem; Separation Problem

    IT WAS REPORTED THAT THE AGENT DISASSEMBLED THE CERAMIC HEAD AND ASSEMBLED IT INTO THE BIPOLAR HEAD IN THE CORRECT WAY. AFTER THE DOCTOR IMPLANTED THE PRODUCT, HE FOUND THAT THE FEMORAL HEAD COULD NOT BE FIXED ON STEM STABLY AND FORMED A TAPER LOCKING. THERE WAS A 70 MINUTE DELAY IN THE SURGERY. PROLONGED ANESTHESIA TIME AND INCREASED BLOOD TRANSFUSION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15961921 ↗
  • InjuryOct 13, 2021Corroded; Material Erosion; Corroded; Material Erosion

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 12629114 ↗
  • MalfunctionJul 1, 2020Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED DURING INSPECTION AT THE (B)(6) WAREHOUSE THAT THE PACKAGE WAS DAMAGED. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Patient Involvement

    MAUDE report 10218227 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032726, K161830, K192660. One submission often covers many models, so check the product and lot details in each record.

  • Z-1191-2018 · Mar 1, 2018 · Terminated
    Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 7.5 Extended Offset Reduced Neck Length; Part Number: 00-7711-0
  • Z-1046-2016 · Feb 11, 2016 · Terminated
    M/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 62937093
  • Z-2033-2014 · May 23, 2014 · Terminated
    Zimmer M/L Taper Hip Stem Standard & Extended Offset, HIP JOINT PROSTHESIS, Sizes: 5.0, 7.5, 9.0, 10.0, 11.0, 12.5, 13.5, 15.0, 11.25, 17.25, 20.0 & 22.5. The femoral stem com