Kinectiv® 00-7848-013-00Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of Kinectiv (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024137820), received between Sep 9, 2021 and Nov 18, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K063251, K182678); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Device Dislodged or Dislocated1
Patient problems
- Bone Fracture(s)2
- Failure of Implant1
- Hematoma1
- Joint Dislocation1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT WAS REVISED APPROXIMATELY 1 WEEK LATER DUE TO DISLOCATION. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Joint Dislocation
MAUDE report 23584620 ↗InjuryDec 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE. SUBSEQUENTLY, THE PATIENT WAS REVISED APPROXIMATELY 3 MONTHS LATER DUE TO PERIPROSTHETIC FRACTURE FOLLOWING A FALL. DURING THE REVISION 5 CERCLAGE WIRES WERE USED TO REDUCE AND STABILIZE THE FRACTURE. THE STEM AND HEAD WERE REPLACED WITHOUT COMPLICATIONS. IT WAS REPORTED THAT NO FURTHER INFORMATION WAS AVAILABLE
Patient problems: Bone Fracture(s); Failure of Implant; Hematoma
MAUDE report 20835906 ↗InjuryApr 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 1 MONTH POST IMPLANTATION DUE TO A PERIPROSTHETIC FRACTURE FROM A FALL. THE HEAD, NECK, AND STEM WERE EXCHANGED WITH ADDITION OF CERCLAGE CABLES TO STABILIZE THE FRACTURE WITHOUT COMPLICATIONS. NO ADDITIONAL INFORMATION.
Patient problems: Bone Fracture(s)
MAUDE report 14086228 ↗InjurySep 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT UNDERWENT A HIP REVISION AFTER AN UNKNOWN AMOUNT OF TIME POST IMPLANTATION DUE TO PAIN. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Pain
MAUDE report 12447987 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K063251, K182678. One submission often covers many models, so check the product and lot details in each record.
- Z-2364-2018 · May 30, 2018 · TerminatedZimmer M/L Taper Hip Prosthesis With Kinectiv Technology Cementless, Size 6; 00-7713-006-00
- Z-2375-2018 · May 30, 2018 · TerminatedZimmer M/L Taper Hip Prosthesis With Kinective Technology Cementless, Size 9; 00-7713-009-00
- Z-2365-2018 · May 30, 2018 · TerminatedZimmer M/L Taper Hip Prosthesis With Kinective Technology Cementless, Size 10; 00-7713-010-00
- Z-1617-2009 · May 29, 2009 · TerminatedZimmer Kinectivo Technology Modular Neck, BB, 12/14 neck taper, sterile, Zimmer, Warsaw, IN; Catalog Number: 00-7848-022-01. Orthopedic implant used in total hip arthroplasty.
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page