Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024149168), received between Nov 19, 2020 and Jan 23, 2024 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K091508, K123019, K151448); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
- Manufacturing, Packaging or Shipping Problem1
- Naturally Worn1
- Packaging Problem1
- Unstable1
Patient problems
- Pain3
- Failure of Implant2
- Joint Dislocation2
- Metal Related Pathology2
- Muscle/Tendon Damage2
- Necrosis2
- Ambulation or Postural Difficulties1
- Joint effusion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 23, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 5 YEARS POST IMPLANTATION DUE TO THE DEVELOPMENT OF PAIN, LIMB LENGTH DISCREPANCY, INSTABILITY WITH RECURRENT DISLOCATIONS, AND METALLOSIS. DURING THE PROCEDURE LARGE GRAY-BLACK SYNOVIAL EFFUSION, EXTENSIVE OSTEOLYSIS AND NECROTIC LESION OF R HIP SOFT TISSUE WITH COMPLETE LOSS OF HIP ABDUCTOR ATTACHMENTS ON PROXIMAL FEMUR NOTED. THERE WAS EXTENSIVE DEBRIDEMENT OF R HIP PSEUDOTUMOR. GROSS BLACK DEBRIS AROUND THE FEMORAL TRUNNION, THE PATIENT UNDERWENT REVISION, THE HEAD AND LINER COMPONENTS WERE EXCHANGED WITHOUT COMPLICATIONS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Ambulation or Postural Difficulties; Joint Dislocation; Joint effusion; Metal Related Pathology; Muscle/Tendon Damage; Necrosis; Osteolysis; Pain; Unequal Limb Length
MAUDE report 18562111 ↗MalfunctionFeb 28, 2023Manufacturing, Packaging or Shipping Problem; Packaging Problem; Manufacturing, Packaging or Shipping Problem; Packaging Problem
IT WAS REPORTED THAT THE CIRCULATING NURSE OPENED THE IMPLANT, AND THE SCRUB NURSE NOTICED A HAIR INSIDE OF THE STERILE PACKAGING LAYING ON THE INSIDE OF THE LINER. THERE WAS NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16453904 ↗InjuryOct 13, 2021Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT HIP ARTHROPLASTY REVISION TO ADDRESS PAIN, INSTABILITY, SUBLUXATION AND DISLOCATION APPROXIMATELY NINE (9) YEARS POST-OPERATIVELY. DURING THE REVISION THE SURGEON WAS UNABLE TO REMOVE NECK FROM TAPER, AND ELECTED TO RETAIN STEM AS IT WAS COLD-WELDED. ALL OTHER COMPONENTS WERE REVISED. INITIAL OPERATIVE NOTES NOTED NO INTRAOPERATIVE COMPLICATIONS. REVISION OPERATIVE NOTES IDENTIFIED BROWNISH COLORED FLUID IN THE JOINT AND THE HIP WAS FOUND TO BE DISLOCATED WHEN THE CAPSULE WAS OPENED. THE HEAD, SHELL AND LINER WERE REMOVED AND SOME SIGNIFICANT ANTERIOR WALL BONE LOSS WAS NOTED. THE NECK COULD NOT BE DISASSOCIATED FROM THE STEM AND NEITHER WERE REVISED. NO INTRAOPERATIVE COMPLICATIONS RELEVANT TO THE REPORTED EVENT WERE IDENTIFIED.
Patient problems: Failure of Implant; Joint Dislocation; Metal Related Pathology; Muscle/Tendon Damage; Necrosis; Osteopenia/ Osteoporosis; Pain; Scar Tissue; Subluxation
MAUDE report 12627835 ↗InjuryNov 19, 2020Naturally Worn; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY AND WAS SUBSEQUENTLY REVISED 10 YEARS LATER DUE TO PAIN, INSTABILITY, CUP MIGRATION, LINER WEAR, AND METAL RELATED PATHOLOGY. OPERATIVE NOTE REPORTED INSTABILITY DUE TO POLY LINER DAMAGE WITH VERTICALLY ORIENTATED ACETABULAR COMPONENT. METALLOSIS 2ND TRUNNION WEAR MODULAR NECK AND STEM INTERFACE. TISSUE STAINED WITH GROSS GREYISH BLACK METALLOSIS TYPE TISSUE ON BOTH ACETABULAR AND FEMORAL COMPONENTS. CUP. LINER, HEAD AND STEM REVISED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Patient problems: Failure of Implant; No Information; Pain
MAUDE report 10869501 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K091508, K123019, K151448. One submission often covers many models, so check the product and lot details in each record.
- Z-0094-2015 · Oct 1, 2014 · TerminatedContinuum Acetabular System Trabecular Metal Shell with Multi Holes, Porous 68mm, Prosthesis, Hip, Semi-Constrained, Metal/Polymer
- Z-0071-2011 · Mar 19, 2010 · TerminatedContinuum Acetabular Systems Shell Inserter Adapter with Rotational Control, 00-8790-003-00, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
- Z-0072-2011 · Mar 19, 2010 · TerminatedContinuum Acetabular Systems Shell Inserter Adapter without Rotational Control, 00-8790-003-10, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page